Skip to content

TPS 5100004000Recalls & reported problems

Stryker Corporation · Cord · Class 1

All models of TPS (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker TPS cord is a fiber-optic light cord used in surgery. Most of the 50 reports describe the cord or its light source overheating, sometimes powering up on its own, staying activated, or detaching; a few mention burns on the patient's skin or neck. There are no deaths and no injuries reported. No recalls name this device, and reporting has recently slowed, with 1 report in the last 12 months compared with 2 in the prior 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
50
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 04546540970275), received between Jul 24, 2018 and Apr 22, 2026 — classified as 50 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (29 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 3 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 3 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device29
  • Unintended Power Up13
  • Device Remains Activated6
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions32
  • No Patient Involvement14
  • Burn(s)6
  • Insufficient Information3
  • No Known Impact Or Consequence To Patient2
  • Burn, Thermal1
  • Full thickness (Third Degree) Burn1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 22, 2026Overheating of Device

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1 ¿ MARCH 31, 2026. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: OVERHEATING OF DEVICE. 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24954322 ↗
  • MalfunctionOct 28, 2025Overheating of Device; Overheating of Device

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: OVERHEATING OF DEVICE. 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23399942 ↗
  • MalfunctionJan 21, 2025Overheating of Device; Overheating of Device

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 2 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21201158 ↗
  • MalfunctionOct 21, 2024Overheating of Device; Overheating of Device

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20500482 ↗
  • MalfunctionJul 24, 2024Overheating of Device; Overheating of Device

    THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 4 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19823121 ↗
  • MalfunctionJul 24, 2024Unintended Power Up; Unintended Power Up

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE RAN WITHOUT USER ACTIVATION. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19820055 ↗
  • MalfunctionJul 12, 2024

    AFTER UNDRAPING, BLACK SPOTS WERE NOTICED UNDER CHIN AND ONE ON NECK. APPEARS TO BE A BURN. SURGEONS AND TECH ARE UNSURE HOW THIS COULD'VE HAPPENED BUT SPECULATE POSSIBLE BURN FROM LIGHTED RETRACTOR OR HEATED SALINE/SOOT FROM IRRIGATION RUNNING OFF OF TPS. LATER LEARNED THAT THE BLACK WAS FROM SOOT GENERATED BY HEATED LUBRICANT.

    Patient problems: Full thickness (Third Degree) Burn

    MAUDE report 19728734 ↗
  • MalfunctionApr 16, 2024Unintended Power Up; Unintended Power Up

    THIS REPORT SUMMARIZES (B)(4) MALFUNCTION EVENT IN WHICH THE DEVICE RAN WITHOUT USER ACTIVATION. - (B)(4) EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19113388 ↗
  • MalfunctionApr 15, 2024Overheating of Device; Overheating of Device

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19113337 ↗
  • MalfunctionApr 2, 2024Overheating of Device; Overheating of Device

    THE DISTRIBUTER REPORTED THAT THE DEVICE OVERHEATED DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO INFORMATION WAS AVAILABLE REGARDING ADVERSE CONSEQUENCES, SURGICAL DELAY OR MEDICAL INTERVENTION.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19028790 ↗
  • MalfunctionJan 22, 2024Overheating of Device; Overheating of Device

    THIS REPORT SUMMARIZES (B)(4) MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - (B)(4) EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. - (B)(4) EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT. - (B)(4) EVENTS HAD INSUFFICIENT INFORMATION RECEIVED.

    Patient problems: Burn(s); No Clinical Signs, Symptoms or Conditions

    MAUDE report 18557622 ↗
  • MalfunctionJan 22, 2024Unintended Power Up; Unintended Power Up

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE RAN WITHOUT USER ACTIVATION. - 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18556702 ↗
  • MalfunctionOct 23, 2023Overheating of Device; Overheating of Device

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17993052 ↗
  • MalfunctionJul 26, 2023Overheating of Device; Overheating of Device

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 2 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 1 EVENT WHICH HAD A MILD INJURY, ILLNESS OR IMPAIRMENT WHICH CAN BE TREATED WITH MINIMAL OR NO INTERVENTION.

    Patient problems: Burn(s); No Clinical Signs, Symptoms or Conditions

    MAUDE report 17401602 ↗
  • MalfunctionApr 25, 2023Overheating of Device; Overheating of Device

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16807092 ↗
  • MalfunctionApr 24, 2023Unintended Power Up; Unintended Power Up

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE RAN WITHOUT USER ACTIVATION. - 2 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16802854 ↗
  • MalfunctionJan 29, 2023Overheating of Device; Overheating of Device

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 1 EVENT HAD INSUFFICIENT INFORMATION RECEIVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16261910 ↗
  • MalfunctionJul 26, 2022Overheating of Device; Overheating of Device

    THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 3 EVENTS HAD INSUFFICIENT INFORMATION RECEIVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15102159 ↗
  • MalfunctionJul 26, 2022Device Remains Activated; Device Remains Activated

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD RUN-ON. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15101737 ↗
  • MalfunctionApr 25, 2022Overheating of Device; Overheating of Device

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14193156 ↗
  • MalfunctionJan 19, 2022Overheating of Device; Overheating of Device

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 3 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13291650 ↗
  • MalfunctionOct 22, 2021Overheating of Device; Overheating of Device

    THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 5 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12680259 ↗
  • MalfunctionJul 21, 2021Unintended Power Up; Unintended Power Up

    THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS IN WHICH THE DEVICE RAN WITHOUT USER ACTIVATION. 5 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12204162 ↗
  • MalfunctionApr 27, 2021Unintended Power Up; Unintended Power Up

    THE USER FACILITY REPORTED THAT THE DEVICE HAD UNINTENDED ACTIVATION DURING TESTING PRIOR TO A PROCEDURE.  THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11731477 ↗
  • MalfunctionApr 26, 2021Unintended Power Up; Unintended Power Up

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE RAN WITHOUT USER ACTIVATION. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11728518 ↗

Other UNIT, OPERATIVE DENTAL devices

All UNIT, OPERATIVE DENTAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page