Skip to content

NEOVEIL NV-ET-M60A-2Recalls & reported problems

Gunze Limited · As packaged, NEOVEIL is a suture- and staple-reinforcement product composed of 100% bioabsorbable polyglycolic acid (PGA). This nonwoven product is dyed with D&C Green No.6 in order to make it readily visible to the surgeon. Two forms of NEOVEIL are provided. The Tube type model constitutes of pre-formed porous bio-absorbable nonwoven sheets and is intended for staple-line reinforcement. The Tube type model is provided in the form of sleeves, one for the cartridge and one for the anvil on a corresponding stapler. After deployment of the tube type reinforcement material, the non-degradable elastic knits, comprised of polyurethane and nylon, are removed and discarded along with the PGA tacking sutures. The Sheet type model is simply a porous fibrous bio-absorbable sheet which is intended for suture-line reinforcement. The thickness of the bioabsorbable NEOVEIL staple line reinforcement ranges from 0.1 mm to 0.85 mm. Mesh weave characteristics and pore size are not applicable since NEOVEIL is nonwoven material. Average basis weight of the NEOVEIL model ranges from 35 to 225 (g/m2). · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04547301462534), received between May 31, 2026 and May 31, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Material Integrity Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Integrity Problem1

Patient problems

  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 31, 2026Material Integrity Problem

    A CUSTOMERS USES NEOVEIL TUBES IN BARIATRIC SURGERY. HE RECENTLY TESTED REFERENCE NV-ET-M60A-2 WITH GOLD PANTHER STAPLERS CADF-60S (GOLD). DURING STAPLING, A WIRE BROKE, AND ONE OF THE WHITE PARTS REMAINED SEWN TO THE BLUE PART. THE SURGEON CUT THE WHITE PART TO REMOVE IT, BUT A SMALL FRAGMENT WAS LEFT IN THE PATIENT. BASED ON OUR OBSERVATIONS, THIS ISSUE MAY BE DUE TO A CERTAIN FRAGILITY OF THE NV TUBE WIRES.

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 25344872 ↗

Other Mesh, Surgical, Absorbable, Staple Line Reinforcement devices

All Mesh, Surgical, Absorbable, Staple Line Reinforcement devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page