ASPIRE Cristalle DR 3500 W 24X30 USA + ERecalls & reported problems
Fujifilm Corporation · The Fujifilm Digital Mammography Systems, ASPIRE Cristalle (FDR MS-3500) generates full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04547410297546), received between Nov 7, 2018 and Mar 26, 2024 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (K133972, K173132, K212873); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jun 11, 2025Class IIOpen, Classified
Devices had an unapproved slabbing software function enabled for use.
Z-2219-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method2
- No Apparent Adverse Event2
- Patient Data Problem1
Patient problems
- Bruise/Contusion2
- No Known Impact Or Consequence To Patient2
- Insufficient Information1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 26, 2024Patient Data Problem; Patient Data Problem
ON 4TH MARCH 2024, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING DR 3500 W 24X30 USA E. IT WAS REPORTED THAT THE DEVICE IS UNABLE TO SEPARATE CASES IN THE MAMMO UNIT, IT KEEPS MIXING UP THE PATIENT'S INFORMATION THAT DO NOT MATCH THE CORRECT PATIENT. THERE IS NO PATIENT HARM OR INJURY REPORTED. THERE IS NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: Insufficient Information
MAUDE report 18983478 ↗MalfunctionFeb 28, 2019No Apparent Adverse Event; No Apparent Adverse Event
THE CUSTOMER REPORTS THAT SCREENINGS CONDUCTED WITH THE ASPIRE CRISTALLE DIGITAL BREAST TOMOSYNTHESIS (DBT) OPTION MAY BE GENERATING LOWER PATIENT RECALL RATES AT THEIR HEALTHCARE FACILITY THAN EXPECTED. WHILE THE ASPIRE CRISTALLE PRODUCT IS FUNCTIONING TO SPECIFICATIONS, FURTHER REVIEW IS BEING CONDUCTED. THERE IS NO KNOWN DEATH OR SERIOUS INJURY ASSOCIATED WITH THIS EVENT. THIS EVENT IS BEING REPORTED IN AN ABUNDANCE OF CAUTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8378403 ↗MalfunctionFeb 28, 2019No Apparent Adverse Event; No Apparent Adverse Event
THE CUSTOMER REPORTS THAT SCREENINGS CONDUCTED WITH THE ASPIRE CRISTALLE DIGITAL BREAST TOMOSYNTHESIS (DBT) OPTION MAY BE GENERATING LOWER PATIENT RECALL RATES AT THEIR HEALTHCARE FACILITY THAN EXPECTED. WHILE THE ASPIRE CRISTALLE PRODUCT IS FUNCTIONING TO SPECIFICATIONS, FURTHER REVIEW IS BEING CONDUCTED. THERE IS NO KNOWN DEATH OR SERIOUS INJURY ASSOCIATED WITH THIS EVENT. THIS EVENT IS BEING REPORTED IN AN ABUNDANCE OF CAUTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8378399 ↗InjuryNov 7, 2018Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT AN INJURY OCCURRED DURING AN EXAM BEING PERFORMED WITH AN ASPIRE CRISTALLE. THE PATIENT WITH MULTIPLE SCLEROSIS WHO WAS AN INVALID AND COULD NOT USE THEIR HANDS WAS SEATED IN A BIOPSY CHAIR. THE PRIMARY MAMMOGRAPHER PRESSED THE "ONE TOUCH" BUTTON TO MOVE THE GANTRY TO THE HOME POSITION, AT THIS TIME THE ASSISTING MAMMOGRAPHER HAD THEIR HAND BETWEEN THE CHAIR AND THE MOVING TUBE ARM. THE MOVING TUBE ARM HIT THE CHAIR AND THE ASSISTING MAMMOGRAPHER'S HAND IN BETWEEN; THE TUBE ARM BENDED WHEN IT HIT THE CHAIR AND ASSISTING MAMMOGRAPHER. THE ASSISTING MAMMOGRAPHER CONTINUED WORKING UNTIL THE NEXT DAY, HOWEVER THEN THE MAMMOGRAPHER (ASSISTING) STARTED FEELING PAIN AND WENT TO URGENT CARE; THE HAND WAS BLACK AND BLUE. AFTER FURTHER FOLLOW UP WITH THE ASSISTING MAMMOGRAPHER, IT WAS FOUND THAT AN X-RAY WAS PERFORMED FOR THE INJURED HAND AND THERE WERE NO BONE BREAKS; SEVERAL DAYS LATER THE BRUISING WENT DOWN AND THE HAND WAS BACK TO NORMAL. THERE WAS NO MALFUNCTION OF THE DEVICE. THERE WAS NO ADVERSE EVENT OR INJURY TO THE PRIMARY MAMMOGRAPHER OR THE PATIENT. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Bruise/Contusion
MAUDE report 8046224 ↗InjuryNov 7, 2018Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT AN INJURY OCCURRED DURING AN EXAM BEING PERFORMED WITH AN ASPIRE CRISTALLE. THE PATIENT WITH MULTIPLE SCLEROSIS WHO WAS AN INVALID AND COULD NOT USE THEIR HANDS WAS SEATED IN A BIOPSY CHAIR. THE PRIMARY MAMMOGRAPHER PRESSED THE "ONE TOUCH" BUTTON TO MOVE THE GANTRY TO THE HOME POSITION, AT THIS TIME THE ASSISTING MAMMOGRAPHER HAD THEIR HAND BETWEEN THE CHAIR AND THE MOVING TUBE ARM. THE MOVING TUBE ARM HIT THE CHAIR AND THE ASSISTING MAMMOGRAPHER'S HAND IN BETWEEN; THE TUBE ARM BENDED WHEN IT HIT THE CHAIR AND ASSISTING MAMMOGRAPHER. THE ASSISTING MAMMOGRAPHER CONTINUED WORKING UNTIL THE NEXT DAY; HOWEVER, THEN THE MAMMOGRAPHER (ASSISTING) STARTED FEELING PAIN AND WENT TO URGENT CARE. THE HAND WAS BLACK AND BLUE. AFTER FURTHER FOLLOW UP WITH THE ASSISTING MAMMOGRAPHER, IT WAS FOUND THAT AN X-RAY WAS PERFORMED FOR THE INJURED HAND AND THERE WERE NO BONE BREAKS; SEVERAL DAYS LATER THE BRUISING WENT DOWN AND THE HAND WAS BACK TO NORMAL. THERE WAS NO MALFUNCTION OF THE DEVICE. THERE WAS NO ADVERSE EVENT OR INJURY TO THE PRIMARY MAMMOGRAPHER OR THE PATIENT. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: Bruise/Contusion
MAUDE report 8046195 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K133972, K173132, K212873. One submission often covers many models, so check the product and lot details in each record.
- Z-0661-2018 · Aug 4, 2017 · TerminatedASPIRE Cristalle (for FDR-3000AWS) The Fujifilm Digital Mammography System, ASPIRE Cristalle (FDR MS-3500) generates full field digital mammography images that can, as other full
- Z-0386-2018 · Jul 14, 2017 · TerminatedASPIRE Cristalle (FDR MS-3500) Mammography machine The Fujifilm Digital Mammography Systems, Aspire HD Plus (FDR MS-2500) and Aspire HD-s (FDR MS-2000), generate full-field digit
Other Full field digital,system,x-ray,mammographic devices
- Selenia Dimensions ASY-04160Hologic, Inc. · Full field digital,system,x-ray,mammographic2 recalls47 reports
- Selenia Dimensions ASY-01966Hologic, Inc. · Full field digital,system,x-ray,mammographic131 reports
- Mammomat Inspiration 10140000Siemens Healthcare GmbH · Full field digital,system,x-ray,mammographic2 recalls8 reports
- Senographe Pristina EKB3Ge Medical Systems · Full field digital, system, x-ray, mammographic2 recalls0 reports
- Senographe Pristina NAY3Ge Medical Systems · Full field digital, system, x-ray, mammographic2 recalls0 reports
- Senographe Pristina EKB1Ge Medical Systems · Full field digital,system,x-ray,mammographic2 recalls0 reports
- Senographe Pristina NAY2Ge Medical Systems · Full field digital, system, x-ray, mammographic2 recalls0 reports
- Senographe Pristina EKB2Ge Medical Systems · Full field digital, system, x-ray, mammographic2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page