FUJIFILM EG-580UTRecalls & reported problems
Fujifilm Corporation · FUJIFILM Endoscope Model EG-580UT is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of submucosal and peripheral organs of the esophagus, stomach, and duodenum. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04547410349061), received between Jul 6, 2023 and Oct 8, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Microbial Contamination of Device” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Microbial Contamination of Device3
- Contamination /Decontamination Problem2
- Residue After Decontamination2
- Contamination1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 8, 2024Microbial Contamination of Device; Microbial Contamination of Device
IT WAS REPORTED THAT THERE WAS A LEAK AT THE BIFURCATION OF VIDEO/US CONNECTOR. ALSO REPORTED A FAILED CULTURE PROCESS WITH POSITIVE FINDING AT THE ELEVATOR AND A SECOND FINDING (ALSO AT ELEVATOR) FOLLOWING THE SUBSEQUENT REPROCESSING. CUSTOMER REQUESTED SERVICE TO PERFORM A FULL PRODUCT MAINTENANCE WITH THIS IN MIND. THERE WAS NO PATIENT INVOLVEMENT. THE FOLLOWING INFORMATION WAS PROVIDED REGARDING THE RESULTS OF THE CULTURE TESTS. ON (B)(6) 2024, THE SCOPE WAS CULTURED. RESULTS FROM THE CULTURE GREW CANDIDA ALBICANS. ON (B)(6) 2024 THE SCOPE WAS RE-CULTURED. POSITIVE CULTURE GREW KLEBSIELLA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20394591 ↗MalfunctionDec 5, 2023Contamination; Contamination
ON (B)(6) 2023, FUJIFILM CORPORATION WAS INFORMED OF AN EVENT INVOLVING EG-580UT. IT WAS REPORTED THAT THERE IS AN ISSUE OCCURRED WITH THE ELEVATOR MECHANISM. IN ADDITIONAL, THE SCOPE FAILED BOTH ATP AND CULTURING. ON (B)(6) 2023, FUJIFILM CORPORATION WAS INFORMED OF AN EVENT INVOLVING THE SAME EG-580UT. IT WAS REPORTED THAT THE SCOPE FAILED CULTURING MULTIPLE TIMES. THERE IS NO PATIENT INVOLVEMENT, SERIOUS INJURY, OR DEATH ASSOCIATED WITH THE EVENT. THIS REPORT IS BEING SUBMITTED DUE TO POTENTIAL OF CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18271829 ↗MalfunctionDec 4, 2023Microbial Contamination of Device; Microbial Contamination of Device
ON NOVEMBER 8TH, 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EG-580UT. IT WAS REPORTED THAT THE SCOPE FAILED CULTURING MULTIPLE TIMES. THERE IS NO PATIENT INVOLVEMENT, SERIOUS INJURY, OR DEATH ASSOCIATED WITH THE EVENT. AS SUCH, THIS REPORT IS BEING SUBMITTED DUE TO POTENTIAL OF CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18256858 ↗MalfunctionNov 13, 2023Microbial Contamination of Device; Microbial Contamination of Device
ON OCTOBER 17TH, 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EG-580UT. IT WAS REPORTED THAT THERE IS AN ISSUE OCCURRED WITH THE ELEVATOR MECHANISM. IN ADDITIONAL, THE SCOPE FAILED BOTH ATP AND CULTURING. THERE IS NO PATIENT INVOLVEMENT, SERIOUS INJURY, OR DEATH ASSOCIATED WITH THE EVENT. AS SUCH, THIS REPORT IS BEING SUBMITTED DUE TO POTENTIAL OF CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18120769 ↗MalfunctionJul 13, 2023Residue After Decontamination; Contamination /Decontamination Problem; Residue After Decontamination; Contamination /Decontamination Problem
ON JUNE 13, 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EG-580UT. IT WAS REPORTED THAT DURING PREPARATION A RED LIQUID WAS OBSERVED ON THE BALLOON ATTACHED TO THE DISTAL END OF THE SCOPE. THE SCOPE WAS WASHED TWICE WITH AER, BUT THE RED LIQUID FLOWED OUT OF THE NOZZLE, AND USE OF THE SCOPE WAS DISCONTINUED. THERE IS NO PATIENT INVOLVEMENT, SERIOUS INJURY, OR DEATH ASSOCIATED WITH THE EVENT. AS SUCH, THIS REPORT IS BEING SUBMITTED IN AN ABUNDANCE OF CAUTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17316100 ↗MalfunctionJul 6, 2023Residue After Decontamination; Contamination /Decontamination Problem; Residue After Decontamination; Contamination /Decontamination Problem
ON JUNE 8, 2023, FUJIFILM CORPORATION WAS INFORMED OF AN EVENT INVOLVING EG-580UT. IT WAS REPORTED THAT DURING PREPARATION A RED LIQUID WAS OBSERVED ON THE BALLOON ATTACHED TO THE DISTAL END OF THE SCOPE. THE SCOPE WAS WASHED TWICE WITH AER, BUT THE RED LIQUID FLOWED OUT OF THE NOZZLE, AND USE OF THE SCOPE WAS DISCONTINUED. THERE IS NO PATIENT INVOLVEMENT, SERIOUS INJURY, OR DEATH ASSOCIATED WITH THE EVENT. AS SUCH, THIS REPORT IS BEING SUBMITTED IN AN ABUNDANCE OF CAUTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17274415 ↗
Other Transducer, Ultrasonic, Diagnostic devices
- Site~Rite 8 Ultrasound System 9770603Bard Access Systems, Inc. · Transducer, ultrasonic, diagnostic620 reports
- ULTRASONIC PROBE UM-S20-17SOlympus Medical Systems Corp. · Transducer, ultrasonic, diagnostic2 deaths344 reports
- Transducer X7-2t 989605414122Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic6 recalls21 reports
- EPIQ Diagnostic Ultrasound System 989605488741Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic3 recalls165 reports
- bk3000, bk3500, bk5000, bkActiv 2300Bk Medical Aps · Transducer, Ultrasonic, Diagnostic6 recalls1 report
- Transducer L12-3 989605412072Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls1 report
- Transducer eL18-4 EMT 989605431531Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls0 reports
- Transducer L12-3 Ergo 989605492481Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page