Skip to content

FUJIFILM EG-580UTRecalls & reported problems

Fujifilm Corporation · FUJIFILM Endoscope Model EG-580UT is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of submucosal and peripheral organs of the esophagus, stomach, and duodenum. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04547410349061), received between Jul 6, 2023 and Oct 8, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Microbial Contamination of Device” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Microbial Contamination of Device3
  • Contamination /Decontamination Problem2
  • Residue After Decontamination2
  • Contamination1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 8, 2024Microbial Contamination of Device; Microbial Contamination of Device

    IT WAS REPORTED THAT THERE WAS A LEAK AT THE BIFURCATION OF VIDEO/US CONNECTOR. ALSO REPORTED A FAILED CULTURE PROCESS WITH POSITIVE FINDING AT THE ELEVATOR AND A SECOND FINDING (ALSO AT ELEVATOR) FOLLOWING THE SUBSEQUENT REPROCESSING. CUSTOMER REQUESTED SERVICE TO PERFORM A FULL PRODUCT MAINTENANCE WITH THIS IN MIND. THERE WAS NO PATIENT INVOLVEMENT. THE FOLLOWING INFORMATION WAS PROVIDED REGARDING THE RESULTS OF THE CULTURE TESTS. ON (B)(6) 2024, THE SCOPE WAS CULTURED. RESULTS FROM THE CULTURE GREW CANDIDA ALBICANS. ON (B)(6) 2024 THE SCOPE WAS RE-CULTURED. POSITIVE CULTURE GREW KLEBSIELLA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20394591 ↗
  • MalfunctionDec 5, 2023Contamination; Contamination

    ON (B)(6) 2023, FUJIFILM CORPORATION WAS INFORMED OF AN EVENT INVOLVING EG-580UT. IT WAS REPORTED THAT THERE IS AN ISSUE OCCURRED WITH THE ELEVATOR MECHANISM. IN ADDITIONAL, THE SCOPE FAILED BOTH ATP AND CULTURING. ON (B)(6) 2023, FUJIFILM CORPORATION WAS INFORMED OF AN EVENT INVOLVING THE SAME EG-580UT. IT WAS REPORTED THAT THE SCOPE FAILED CULTURING MULTIPLE TIMES. THERE IS NO PATIENT INVOLVEMENT, SERIOUS INJURY, OR DEATH ASSOCIATED WITH THE EVENT. THIS REPORT IS BEING SUBMITTED DUE TO POTENTIAL OF CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18271829 ↗
  • MalfunctionDec 4, 2023Microbial Contamination of Device; Microbial Contamination of Device

    ON NOVEMBER 8TH, 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EG-580UT. IT WAS REPORTED THAT THE SCOPE FAILED CULTURING MULTIPLE TIMES. THERE IS NO PATIENT INVOLVEMENT, SERIOUS INJURY, OR DEATH ASSOCIATED WITH THE EVENT. AS SUCH, THIS REPORT IS BEING SUBMITTED DUE TO POTENTIAL OF CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18256858 ↗
  • MalfunctionNov 13, 2023Microbial Contamination of Device; Microbial Contamination of Device

    ON OCTOBER 17TH, 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EG-580UT. IT WAS REPORTED THAT THERE IS AN ISSUE OCCURRED WITH THE ELEVATOR MECHANISM. IN ADDITIONAL, THE SCOPE FAILED BOTH ATP AND CULTURING. THERE IS NO PATIENT INVOLVEMENT, SERIOUS INJURY, OR DEATH ASSOCIATED WITH THE EVENT. AS SUCH, THIS REPORT IS BEING SUBMITTED DUE TO POTENTIAL OF CONTAMINATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18120769 ↗
  • MalfunctionJul 13, 2023Residue After Decontamination; Contamination /Decontamination Problem; Residue After Decontamination; Contamination /Decontamination Problem

    ON JUNE 13, 2023, FUJIFILM HEALTHCARE AMERICAS CORPORATION WAS INFORMED OF AN EVENT INVOLVING EG-580UT. IT WAS REPORTED THAT DURING PREPARATION A RED LIQUID WAS OBSERVED ON THE BALLOON ATTACHED TO THE DISTAL END OF THE SCOPE. THE SCOPE WAS WASHED TWICE WITH AER, BUT THE RED LIQUID FLOWED OUT OF THE NOZZLE, AND USE OF THE SCOPE WAS DISCONTINUED. THERE IS NO PATIENT INVOLVEMENT, SERIOUS INJURY, OR DEATH ASSOCIATED WITH THE EVENT. AS SUCH, THIS REPORT IS BEING SUBMITTED IN AN ABUNDANCE OF CAUTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17316100 ↗
  • MalfunctionJul 6, 2023Residue After Decontamination; Contamination /Decontamination Problem; Residue After Decontamination; Contamination /Decontamination Problem

    ON JUNE 8, 2023, FUJIFILM CORPORATION WAS INFORMED OF AN EVENT INVOLVING EG-580UT. IT WAS REPORTED THAT DURING PREPARATION A RED LIQUID WAS OBSERVED ON THE BALLOON ATTACHED TO THE DISTAL END OF THE SCOPE. THE SCOPE WAS WASHED TWICE WITH AER, BUT THE RED LIQUID FLOWED OUT OF THE NOZZLE, AND USE OF THE SCOPE WAS DISCONTINUED. THERE IS NO PATIENT INVOLVEMENT, SERIOUS INJURY, OR DEATH ASSOCIATED WITH THE EVENT. AS SUCH, THIS REPORT IS BEING SUBMITTED IN AN ABUNDANCE OF CAUTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17274415 ↗

Other Transducer, Ultrasonic, Diagnostic devices

All Transducer, Ultrasonic, Diagnostic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page