iSert Intraocular Lens 251 (yellow), +20.50DRecalls & reported problems
Hoya Medical Singapore Pte. Ltd. · Class 3 · Implant
All models of iSert Intraocular Lens (69)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 04547480450308), received between Jun 3, 2016 and Apr 15, 2026 — classified as 10 injuries, 1 malfunction.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Split, Cut or Torn4
- Device Appears to Trigger Rejection3
- Adverse Event Without Identified Device or Use Problem1
- Entrapment of Device1
- Material Opacification1
- Scratched Material1
- Sharp Edges1
Patient problems
- Failure of Implant4
- Capsular Bag Tear2
- Uveitis2
- No Known Impact Or Consequence To Patient1
- Toxic Anterior Segment Syndrome (TASS)1
- Visual Impairment1
- Vitrectomy1
- Vitreous Loss1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryApr 15, 2026Sharp Edges
EVENT OCCURRED IN CHINA. AFTER IMPLANTATION OF THE INTRAOCULAR LENS INTO THE CAPSULAR BAG, A POSTERIOR CAPSULE RUPTURE WAS NOTED. REPEATED REVIEW OF THE SURGICAL VIDEO SUGGESTED A SUSPICION THAT THE POSTERIOR CAPSULE WAS TORN BY THE IOL HAPTIC. CURRENTLY, THE INTRAOCULAR LENS REMAINS WELL POSITIONED, INTRAOCULAR PRESSURE IS NORMAL, AND VISUAL ACUITY IS 0.4. CLOSE FOLLOW-UP WILL BE CONDUCTED. CAPSULAR BAG TEAR / RUPTURE. PRODUCT REMAINED IN EYE AFTER CLINICAL ASSESSMENT. PATIENT HEALTH NOT IMPACTED PROBLEM CODE: A051102 - SHARP EDGES.
Patient problems: Capsular Bag Tear
MAUDE report 24891884 ↗InjuryJul 11, 2023Scratched Material; Scratched Material
EVENT OCCURRED IN VIETNAM. ORIGINALLY ASSESSED AS NOT REPORTABLE TO LOCAL AUTHORITIES. OVD USED WAS REPORTED AS "UNKNOWN." DAMAGED OPTIC AFTER IMPLANTATION. PROBLEM CODE: A0415 SCRATCHED MATERIAL. IMPLANT DATE: (B)(6) 2020. EXPLANT DATE: (B)(6) 2020. PATIENT IMPACT: DEVICE EXPLANTATION.
Patient problems: Failure of Implant
MAUDE report 17297457 ↗InjuryJul 11, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
EVENT OCCURRED IN VIETNAM. ORIGINALLY ASSESSED AS NOT REPORTABLE TO LOCAL AUTHORITIES. OVD USED WAS REPORTED AS "UNKNOWN." DAMAGED HAPTIC AFTER IMPLANTATION PROBLEM CODE: (B)(4). MATERIAL SPLIT CUT OR TORN IMPLANT DATE: (B)(6) 2020. EXPLANT DATE: (B)(6)2020. PATIENT IMPACT: DEVICE EXPLANTATION.
Patient problems: Failure of Implant
MAUDE report 17297450 ↗InjuryJun 28, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
EVENT OCCURRED IN VIETNAM. ORIGINALLY ASSESSED AS NOT REPORTABLE TO LOCAL AUTHORITIES. OVD USED WAS REPORTED AS "UNKNOWN." DAMAGED HAPTIC AFTER IMPLANTATION PROBLEM CODE: A0414 MATERIAL SPLIT, CUT OR TORN IMPLANT DATE: (B)(6)2020. EXPLANT DATE: (B)(6)2020. PATIENT IMPACT: DEVICE EXPLANTATION.
Patient problems: Failure of Implant
MAUDE report 17218182 ↗InjuryJun 28, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
EVENT OCCURRED IN VIETNAM. ORIGINALLY ASSESSED AS NOT REPORTABLE TO LOCAL AUTHORITIES. OVD USED WAS REPORTED AS "UNKNOWN." DAMAGED HAPTIC AFTER IMPLANTATION PROBLEM CODE: A0414 MATERIAL SPLIT, CUT OR TORN IMPLANT DATE: (B)(6) 2020. EXPLANT DATE: (B)(6) 2020. PATIENT IMPACT: DEVICE EXPLANTATION.
Patient problems: Failure of Implant
MAUDE report 17218171 ↗InjuryDec 2, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
SURGERY DATE: (B)(6)2022. DATE OF EVENT: (B)(6)2022. POST-OPERATIVE COMPLICATIONS INFLAMMATORY RESPONSE AFTER IMPLANTATION TOXIC ANTERIOR SEGMENT SYNDROME (TASS).
Patient problems: Toxic Anterior Segment Syndrome (TASS)
MAUDE report 15911712 ↗InjuryJul 12, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
REPORTED CASE OF INFLAMMATION (B)(6). DATE OF SURGERY: (B)(6) 2022. DATE OF EVENT - NOT PROVIDED. (B)(6) 2022: MANUFACTURER'S DATE OF AWARENESS. (B)(6) 2022: AUTHORIZED COUNTRY REPRESENTATIVE ASSESSED AS A REPORTABLE EVENT TO COFEPRIS BY END OF YEAR.
Patient problems: Uveitis
MAUDE report 14989218 ↗InjuryJul 12, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
REPORTED CASE OF INFLAMMATION / UVEITIS IN MEXICO. DATE OF SURGERY: (B)(6) 2022. DATE OF EVENT - NOT PROVIDED. 05-MAY-2022: MANUFACTURER'S DATE OF AWARENESS. 15-JUN-2022: AUTHORIZED COUNTRY REPRESENTATIVE ASSESSED AS A REPORTABLE EVENT TO COFEPRIS BY END OF YEAR.
Patient problems: Uveitis
MAUDE report 14989168 ↗InjuryFeb 18, 2021Material Opacification; Material Opacification
IMPLANT DATE: (B)(6) 2020. DATE OF EVENT: (B)(6) 2021. PATIENT HAS DEVELOPED POSTERIOR CAPSULAR OPACIFICATION AFTER IMPLANTATION OF IOL. THIS RESULTED IN DROP IN VISUAL ACUITY. YAG NG TREATMENT DONE AND THE CONDITION OF THE PATIENT IS BETTER THAN PREVIOUS.
Patient problems: Visual Impairment
MAUDE report 11346076 ↗InjuryMay 24, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INITIAL COMPLAINT INFORMATION RECEIVED FROM DISTRIBUTOR STATED THAT: "THE TECH STATED THE CAPSULE TORE AFTER THE IOL WAS IMPLANTED. THE IOL WAS THEN EXPLANTED AND REPLACED WITH A DIFFERENT IOL." ADDITIONAL INFORMATION IN THE DATA COLLECTION FORM RECEIVED FROM THE DISTRIBUTOR INDICATED THAT: "CAPSULE TORE WHEN IOL INSERTED. SURGEON THEN REMOVED THE IOL, PERFORMED A VITRECTOMY, AND INSERTED A DIFFERENT IOL." "IOL WAS REMOVED/EXPLANTED - DATE: (B)(6) 2018"
Patient problems: Capsular Bag Tear; Vitrectomy; Vitreous Loss
MAUDE report 7542008 ↗MalfunctionJun 3, 2016Entrapment of Device; Material Split, Cut or Torn
TORN HAPTIC WAS REPORTED DURING IOL INSERTION. THE TRAILING HAPTIC WAS STUCK IN THE INTRODUCER. THE IOL WAS STILL USED. PATIENT OUTCOME IS UNKNOWN AT THIS TIME. THE LENS WAS NOT EXPLANTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5698955 ↗
Other Intraocular Lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page