iSert Intraocular Lens 251 (yellow), +21.00DRecalls & reported problems
Hoya Medical Singapore Pte. Ltd. · Class 3 · Implant
All models of iSert Intraocular Lens (69)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04547480450315), received between Oct 19, 2018 and Jul 18, 2023 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break3
- Material Split, Cut or Torn2
Patient problems
- No Known Impact Or Consequence To Patient4
- Failure of Implant1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryJul 18, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
EVENT OCCURRED IN VIETNAM. ORIGINALLY ASSESSED AS NOT REPORTABLE TO LOCAL AUTHORITIES. BROKEN HAPTIC DURING USE PROBLEM CODE: A0414 MATERIAL SPLIT CUT OR TORN IMPLANT DATE: (B)(6)2020. EXPLANT DATE: (B)(6)2020. PATIENT IMPACT: DEVICE EXPLANTATION.
Patient problems: Failure of Implant
MAUDE report 17347738 ↗MalfunctionDec 5, 2019Break; Break
PEELING HAPTIC DURING USE. PATIENT IMPACT: NO IMPACT. IT OCCURRED IN CHINA, THERE WAS NO IMPACT TO PATIENT, BUT IT WAS REPORTED AS AE TO LOCAL AUTHORITY BY HOSPITAL.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9419427 ↗MalfunctionJul 23, 2019Break; Break
BOTH LENSES WERE PREPARED CORRECTLY USING ADEQUATE AND CORRECT AMOUNT OF HELAON. AFTER INJECTION AND LENS UNFOLDING, CHIPS WERE NOTICED MISSING FROM THE OPTICS. THE LENSES WERE NOT EXPLANTED AS RISK TO THE PATIENT WAS CONSIDERED HIGHER IF EXPLANTING RATHER THAN LEAVING IN SITU. NO PATIENT IMPACT YET, BUT SHARP EDGES OF CHIPS IS CONCERNING AS COULD CAUSE PC RUPTURE AS CAPSULAR BAG SHRINKS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8817558 ↗MalfunctionJan 25, 2019Material Split, Cut or Torn; Material Split, Cut or Torn
LEADING HAPTIC DID NOT FOLD PROPERLY AND WAS TRAPPED BETWEEN PLUNGER AN THE CARTRIDGE TIP, SO THE HAPTIC TORN OFF. LENS HAD TO BE EXPLANTED (NO PATIENT IMPACT). NEW LENS WAS IMPLANTED. SECOND IMPLANTATION WENT WELL WITHOUT HARMING THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8278763 ↗MalfunctionOct 19, 2018Break; Break
REALIZED THERE WAS A CHIP IN THE LENS AND CUT IT OUT OF PATIENTS EYE AND PUT ANOTHER IN. NEW LENS HAD TO BE IMPLANTED BUT WITHOUT HARMING THE PATIENT. PATIENT IMPACT: ONLY LENGTHENING OF PROCEDURE. PATIENT OUTCOME: NO IMPACT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7985670 ↗
Other Intraocular Lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page