Skip to content

iSert Intraocular Lens 251 (yellow), +24.00DRecalls & reported problems

Hoya Medical Singapore Pte. Ltd. · Class 3 · Implant

All models of iSert Intraocular Lens (69)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04547480450377), received between Jul 12, 2023 and Feb 18, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Crack” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack1
  • Material Split, Cut or Torn1

Patient problems

  • Failure of Implant2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 18, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    EVENT OCCURRED IN (B)(6). DAMAGED HAPTIC AFTER IMPLANTATION. TRAILING; DAMAGED HAPTIC; TRAILING HAPTIC DOES NOT FOLD AS PER IFU; HAPTIC STUCK TO OPTIC TEMPORARILY~ FINALLY HAPTIC IS SEPARATED. TRAILING; TUCKING FAILURE; STICKING HAPTIC. PRODUCT REPLACED WITH ANOTHER LENS IMMEDIATELY DURING SURGERY; PATIENT HAS RECOVERED AND IS FINE. THE IOL WAS EXPLANTED ON (B)(6) 2024. PROBLEM CODE: A0414 - MATERIAL SPLIT, CUT OR TORN.

    Patient problems: Failure of Implant

    MAUDE report 21406685 ↗
  • InjuryJul 12, 2023Crack; Crack

    EVENT OCCURRED IN VIETNAM. ORIGINALLY ASSESSED AS NOT REPORTABLE TO LOCAL AUTHORITIES. OVD REPORTED AS "UNKNOWN." DAMAGED OPTIC AFTER IMPLANTATION. PROBLEM CODE: A0404 CRACK. IMPLANT DATE: ON (B)(6) 2020. EXPLANT DATE: ON (B)(6) 2020. PATIENT IMPACT: DEVICE EXPLANTATION.

    Patient problems: Failure of Implant

    MAUDE report 17305500 ↗

Other Intraocular Lens devices

All Intraocular Lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page