Skip to content

iSert Intraocular Lens 250 (clear), +24.00DRecalls & reported problems

Hoya Medical Singapore Pte. Ltd. · Class 3 · Implant

All models of iSert Intraocular Lens (69)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04547480480862), received between Mar 22, 2017 and Dec 28, 2023 — classified as 4 injuries.

The most frequently reported specific device problem is “Crack” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Crack1
  • Entrapment of Device1
  • Inaccurate Information1

Patient problems

  • Capsular Bag Tear1
  • Injury1
  • No Clinical Signs, Symptoms or Conditions1
  • Visual Impairment1
  • Vitreous Loss1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 28, 2023Crack; Crack

    DAMAGED OPTIC AFTER IMPLANTATION/POST-OPERATIVE COMPLICATIONS. THE IMPLANTED IOL HAD A SCRATCH AND THE PATIENT COMPLAINED UNUSUAL FOR THEIR VISION AFTER THE OPERATION. THE DOCTOR DID NOT NOICE THE SCRATCH OF THE IOL DURING THE OPERATION, BUT AFTER THE SURGERY THE DOCTOR CONFIRMED THE SCRATCH ON THE IOL. THE IOL WAS EXPLANTED FROM THE PATIENT EYE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18412544 ↗
  • InjurySep 12, 2019Inaccurate Information; Inaccurate Information

    INCORRECT INFORMATION ON LABEL~THIS IS JUST ONE POSSIBILITY. AFTER IMPLANTING THE 24.00 D. IOL THE PATIENT ENDED UP IN 4.00 D. HYPERCORRECTION. BIOMETRIES BEFORE AND AFTER SURGERY ARE CONSISTENT. THE LENS WAS EXPLANTED. SECONDARY SURGERY WAS DONE ON (B)(6) 2019. A NEW LENS WAS INSERTED WITHOUT ANY PATIENT HARM. AFTER SECOND SURGERY PATIENT FEELS GOOD.

    Patient problems: Visual Impairment

    MAUDE report 9014663 ↗
  • InjuryJul 28, 2017Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED A VITREOUS LEAK AND A SPILT IN THE POSTERIOR CAPSULE WHEN THE LENS WAS INSERTED. THE LENS WAS CUT OUT OF THE BAG AND ANOTHER LENS MODEL WAS PLACED IN THE SULCUS. IN THE SURGEON'S OPINION, THE LIKELY CAUSE OF THE EVENT WAS "SPILT IN CAPSULE BAG."

    Patient problems: Capsular Bag Tear; Vitreous Loss

    MAUDE report 6750665 ↗
  • InjuryMar 22, 2017Entrapment of Device

    IT WAS REPORTED WHILE DOCTOR WAS ADVANCING THE LENS, IT STUCK IN THE INJECTOR. THE INJECTOR/LENS WAS REMOVED AND THE INCISION WAS ENLARGED. A BACK-UP LENS WAS IMPLANTED WITH NO ISSUE. PATIENT'S OUTCOME IS GOOD.

    Patient problems: Injury

    MAUDE report 6424840 ↗

Other Intraocular Lens devices

All Intraocular Lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page