PROBEAT PROBEAT-FRRecalls & reported problems

Hitachi, Ltd. · The PROBEAT-FR is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient treatment site. The PROBEAT-FR is a modification to the cleared PROBEAT-CR system, for installation at a different clinical site. · Class 2

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04560333350204), received between Dec 16, 2025 and Jan 20, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Computer Software Problem” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

2 other recalls cover devices cleared under the same FDA submission (K232032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 6, 2026Open, Classified

    Software anomaly in the patient positioning system may result in positional discrepancy.

    Z-1429-2026 · Root cause: Software design · FDA record ↗

  2. May 27, 2024Open, Classified

    During positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are insufficient in number.

    Z-3125-2024 · Root cause: Software design · FDA record ↗

Most reported problems

Device problems

  • Computer Software Problem1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 20, 2026Computer Software Problem

    DURING PATIENT SETUP FOR PROTON THERAPY, AFTER CBCT IMAGING, THE HMI DISPLAYED COORDINATES MATCHED THE PLANNED VALUES; HOWEVER, A POSITIONAL DEVIATION OF APPROXIMATELY 3 CM IN THE X-AXIS WAS CONFIRMED USING A LASER MARKER. THIS DISCREPANCY WAS RESOLVED AFTER RE-PERFORMING THE SETUP AND VERIFYING WITH PIAS 2D-2D MATCHING MODE. NO PATIENT INJURY OCCURRED, AND TREATMENT PROCEEDED NORMALLY AFTER CORRECTION. A SOFTWARE ANOMALY IN THE PATIENT POSITIONING SYSTEM (PPS) MAY CAUSE A MISMATCH BETWEEN THE COMMANDED ERROR CORRECTION PARAMETER FLAG IN THE COMMAND DATA AND THE COMMANDED ERROR CORRECTION PARAMETER FLAG IN THE COMMAND DATA OF THE COUCH CONTROLLER, RESULTING IN UNINTENDED POSITIONAL DEVIATION WHILE THE HMI CONTINUES TO DISPLAY THE PLANNED COORDINATES. THE INVESTIGATION SUGGESTS THAT THE ERROR-CORRECTION PARAMETER MAY FAIL TO BE APPLIED AS INTENDED AT A RATE OF (B)(4) OCCURRENCES OUT OF APPROXIMATELY (B)(4) OPERATIONS (REPRODUCTION PROBABILITY OF ABOUT (B)(4)). IN CASE WHERE TREATMENT BEGINS WITHOUT PATIENT POSITION VERIFICATION AFTER A SOFTWARE ANOMALY OCCURS IN THE PPS, THERE IS A RISK OF DELIVERING RADIATION TO NON-TARGET TISSUE, POTENTIALLY CAUSING SERIOUS INJURY. THE INCIDENT WA

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24121790 ↗
  • MalfunctionDec 16, 2025Use of Device Problem; Use of Device Problem

    "DURING POSITIONING IN THE PIAS'S 3D3D MATCHING MODE, WHILE TRANSFERRING CBCT IMAGES FROM CIRS TO PIAS, WHEN THE FLOAT IMAGE SELECTION BUTTON ON THE PIAS SCREEN WAS PRESSED, IT DISPLAYED CBCT IMAGES THAT WERE INSUFFICIENT IN NUMBER. THIS REVEALED A CLEAR DISPLACEMENT IN THE Y-DIRECTION (BODY AXIS DIRECTION) COMPARED TO THE PLANNING CT. SUBSEQUENTLY, UPON CHECKING OIS AT THE CUSTOMER'S SITE ON A LATER DATE, IT WAS FOUND THAT INSTEAD OF THE EXPECTED 160 IMAGES, THERE WERE ONLY 135 IMAGES. IT WAS REPORTED THAT THE INCIDENT OCCURRED ON (B)(6) 2024, AND THE CUSTOMER CONTACTED US ON (B)(6) 2024. FOLLOWING THE ABOVE CUSTOMER CONTACT, WE CONDUCTED A REVIEW OF THE LOGS FOR ALL TREATMENTS. IT WAS FOUND THAT ON (B)(6) 2024, THERE WAS A POSSIBILITY OF COMPLETING 3D3D PATIENT POSITIONING APPROXIMATELY 5MM OFF IN THE BODY AXIS DIRECTION (WITH 5 IMAGES MISSING) AND PROCEEDING WITH TREATMENT. THERE HAVE BEEN NO REPORTS OF HEALTH INJURY. [TREATMENT ON (B)(6) 2024] AFTER PERFORMING POSITION CONFIRMATION IN 2D2D MODE FOLLOWING 3D3D POSITIONING, TREATMENT WAS CARRIED OUT AT THE CORRECT POSITION AFTER REPOSITIONING, SO THERE ARE NO HEALTH INJURY. [TREATMENT ON (B)(6) 2024] THERE IS A POSSIBILITY THAT T

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23819356 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K232032. One submission often covers many models, so check the product and lot details in each record.

  • Z-3127-2024 · May 27, 2024 · Open, Classified
    Brand Name: PROBEAT Product Name: PROBEAT-V Model/Catalog Number: N/A Software Version: N/A Product Description: The PROBEAT-V is a proton beam irradiation system, which provides a
  • Z-3126-2024 · May 27, 2024 · Open, Classified
    Brand Name: PROBEAT Product Name: PROBEAT-CR Model/Catalog Number: N/A Software Version: N/A Product Description: The PROBEAT-CR is a proton beam irradiation system, which provides

Other System, Radiation Therapy, Charged-Particle, Medical devices

All System, Radiation Therapy, Charged-Particle, Medical devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.