Skip to content

Conformity stem, #4 1110-3004Recalls & reported problems

United Orthopedic Corp. · Conformity stem, #4 · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04711605583213), received between Sep 23, 2026 and Sep 23, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Arthralgia1
  • Aseptic loosening1
  • Cartilage damage1
  • Metal Related Pathology1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 23, 2026Loosening of Implant Not Related to Bone-Ingrowth

    A complaint case regarding the Conformity Stem was received by the French branch on (b)(6) 2026. An 82-year-old female patient underwent THA on (b)(6) 2024, using a Femoral Head and Conformity Stem manufactured by United Orthopedic Corporation, along with a Serf Novae? Dual Mobility Cup from another manufacturer. Approximately 1 year and 9 months after implantation, the patient developed hip pain. A CT scan revealed loosening of the femoral stem without an associated periprosthetic fracture. Further testing was performed to rule out infection as a potential cause of the loosening. A revision surgery was performed on (b)(6) 2026. Additionally, significant black deposits were observed around the loosened femoral stem intraoperatively. Histopathological analysis of the blackish soft tissue confirmed the diagnosis of metallosis.

    Patient problems: Arthralgia; Aseptic loosening; Cartilage damage; Metal Related Pathology

    MAUDE report 50865561 ↗

Other Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate devices

All Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page