Skip to content

"UNITED" USTAR II Knee System 2115-1330Recalls & reported problems

United Orthopedic Corp. · Femoral component, #3, left · Class 2 · Implant

All models of "UNITED" USTAR II Knee System (22)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04719872179481), received between Mar 5, 2024 and Mar 5, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

5 other recalls cover devices cleared under the same FDA submission (K190100); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1

Patient problems

  • Joint Dislocation1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 5, 2024Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    THE PATIENT REPORTED THAT THEY GOT OUT OF A CHAIR AND EXPERIENCED PAIN. THEY INDICATED NO FALL OR TRAUMA. PREVIOUS USTAR II PATIENT PRESENTED WITH PAIN AND INSTABILITY. X-RAY IMAGING REVEALED DISLOCATION OF THE FEMORAL COMPONENT AND TIBIAL INSERT YOKE AND A REVISION WAS SCHEDULED. DURING SURGERY, SURGEON NOTED THE INSERT SCREW CONNECTING THE FEMORAL COMPONENT INSERT AND YOKE HAD BENT. ORIGINAL SURGERY WAS ON (B)(6) 2023 AND REVISION WAS CONDUCTED ON (B)(6) 2024. REVISION SURGERY WAS SUCCESSFULLY COMPLETED.

    Patient problems: Joint Dislocation

    MAUDE report 18834187 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K190100. One submission often covers many models, so check the product and lot details in each record.

  • Z-2884-2026 · May 20, 2026 · Open, Classified
    Brand Name: USTAR II Knee System Product Name: Hinge Box Trial, L, #1-6 and Hinge Box Trial, R, #1-6 Model/Catalog Number: 9325-5711 thru 9325-5716 and 9325-5721 thru 9325-5726 Sof
  • Z-1215-2024 · Jan 3, 2024 · Open, Classified
    USTAR II Hip System Press-Fit Curved Stem, RHS Ti Plasma Spray for the following reference numbers: REF: PE11151211 - 11x150mm, PE11151213 - 13x150mm, PE11151215 - 15x150mm, PE1
  • Z-1217-2024 · Jan 3, 2024 · Open, Classified
    USTAR II Knee System : Cemented Curved Stem, RHS Ti Plasma Spray, Reference Numbers: REF: PE25153109 - PE25151109 - 9x125mm, PE25151111 - 11X125, PE25151113 - 13X125mm, PE2515111
  • Z-1216-2024 · Jan 3, 2024 · Open, Classified
    USTAR II Knee System Cemented curved stem, RHS, Non Coated for the following reference numbers: REF: PE25153109 - 9x125mm, PE25153209 - 9x150mm, PE25153113 - 13x125mm, PE25153111-
  • Z-0110-2022 · Aug 27, 2021 · Terminated
    United Orthopedics USTAR II Knee System USTAR II Knee System Segment Part, RHS, 25 mm Length, Catalog # 2915-1025

Other Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer devices

All Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer devices →

Sources and updates