NA PE-210AKRecalls & reported problems
Nihon Kohden Corporation · Switch Box for the EEG-1200A · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04931921103937), received between Sep 24, 2025 and Sep 24, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Electrical /Electronic Property Problem1
- Inappropriate/Inadequate Shock/Stimulation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2025Electrical /Electronic Property Problem; Inappropriate/Inadequate Shock/Stimulation; Electrical /Electronic Property Problem; Inappropriate/Inadequate Shock/Stimulation
THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE SWITCH BOX (PE-210AK, SN: 73) GAVE A "THE ELECTRIC STIMULATOR IS OVERLOADED" ERROR MESSAGE WHILE PERFORMING A CORTICAL MAPPING ON A PATIENT. THEY WERE STIMULATING LB1-LB2 (LEFT ANTERIOR HIPPOCAMPUS) WITH BOTH LOW FREQUENCY AND HIGH FREQUENCY STIMULATION (3 MA FOR BOTH). WITH THE LOW FREQUENCY STIMULATION, THEY HEARD A CRACKLING NOISE COMING FROM THE STIMULATION BOX IN ADDITION TO SEEING ARTIFACTS. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23136975 ↗
Other ELECTRODE, CORTICAL devices
- Subdural Electrode TS06R-SP10X-000Ad-Tech Medical Instrument Corporation · ELECTRODE, CORTICAL2 reports
- Subdural Electrode TS04R-SP10X-000Ad-Tech Medical Instrument Corporation · ELECTRODE, CORTICAL1 report
- Subdural Electrode TS08R-SP10X-000Ad-Tech Medical Instrument Corporation · ELECTRODE, CORTICAL1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page