NA MU-110ARecalls & reported problems

Nihon Kohden Corporation · Main Unit for the PSG-1100 · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04931921108437), received between Aug 16, 2023 and Dec 9, 2024 — classified as 4 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Temperature Problem” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5
  • Temperature Problem5
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Blister5

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 9, 2024Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE CUSTOMER REPORTED THAT THEY ARE EXPERIENCING INCORRECT CPAP PRESSURE READINGS DURING STUDIES. FOR EXAMPLE, WHEN SETTING CPAP PRESSURE TO 12 CM H2O, THE STUDY DISPLAYS A VALUE OF 2.2. AFTER TROUBLESHOOTING WITH THE CUSTOMER, IT WAS DETERMINED THAT THE ISSUE WAS WITH THE MU-110AK. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20887881 ↗
  • MalfunctionDec 21, 2023Adverse Event Without Identified Device or Use Problem; Temperature Problem; Adverse Event Without Identified Device or Use Problem; Temperature Problem

    THE CUSTOMER REPORTED THAT ON 11/06/2023 THEY HAD ANOTHER OCCURRENCE OF THE PULSE OXIMETER (SPO2 FINGER PROBE) LEAVING SKIN BLISTERS ON THE PATIENT FINGER. OTHER RELATED TICKETS: 300337771, 300337766, 300340025, 300340026 - DIFFERENT THIRD PARTY PROBE BEING USED.

    Patient problems: Blister; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18378753 ↗
  • InjuryAug 17, 2023Adverse Event Without Identified Device or Use Problem; Temperature Problem; Adverse Event Without Identified Device or Use Problem; Temperature Problem

    THE CUSTOMER REPORTED THAT THEY HAD 4 SEPARATE OCCURRENCES OF THE PULSE OXIMETER (SPO2 FINGER PROBE) LEAVING SKIN BLISTERS ON PATIENT FINGER. THIS IS FOR THE (B)(6) 2023 OCCURRENCE. THE ISSUE IS NOT FOR THE PSG-1100A SYSTEM ITSELF, AND IS FOR ACCESSORY SPO2 PROBE BEING USED WITH THE SYSTEM. OTHER RELATED COMPLAINTS WILL BE REPORTED: (B)(4) - DIFFERENT PROBE MODEL BEING USED; (B)(4) - DIFFERENT PROBE MODEL BEING USED; (B)(4) - DIFFERENT PROBE MODEL BEING USED.

    Patient problems: Blister; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17561848 ↗
  • InjuryAug 16, 2023Adverse Event Without Identified Device or Use Problem; Temperature Problem; Adverse Event Without Identified Device or Use Problem; Temperature Problem

    THE CUSTOMER REPORTED THAT THEY HAD 4 SEPARATE OCCURRENCES OF THE PULSE OXIMETER (SPO2 FINGER PROBE) LEAVING SKIN BLISTERS ON PATIENT FINGER. THIS IS FOR THE (B)(6) 2023 OCCURRENCE. THE ISSUE IS NOT FOR THE PSG-1100A SYSTEM ITSELF, AND IS FOR ACCESSORY SPO2 PROBE BEING USED WITH THE SYSTEM. OTHER RELATED COMPLAINTS WILL BE REPORTED: (B)(4) DIFFERENT PROBE MODEL BEING USED.

    Patient problems: Blister; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17561736 ↗
  • InjuryAug 16, 2023Adverse Event Without Identified Device or Use Problem; Temperature Problem; Adverse Event Without Identified Device or Use Problem; Temperature Problem

    THE CUSTOMER REPORTED THAT THEY HAD 4 SEPARATE OCCURRENCES OF THE PULSE OXIMETER (SPO2 FINGER PROBE) LEAVING SKIN BLISTERS ON PATIENT FINGER. THIS IS FOR THE (B)(6) 2023 OCCURRENCE; AND AS THEY DID NOT PROVIDE THE DAY IT OCCURRED, WE WILL NOTE IT AS (B)(6) 2023. THE ISSUE IS NOT FOR THE PSG-1100A SYSTEM ITSELF, AND IS FOR ACCESSORY SPO2 PROBE BEING USED WITH THE SYSTEM. OTHER RELATED COMPLAINTS WILL BE REPORTED: (B)(4) - DIFFERENT PROBE MODEL BEING USED.

    Patient problems: Blister; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17561712 ↗
  • InjuryAug 16, 2023Adverse Event Without Identified Device or Use Problem; Temperature Problem; Adverse Event Without Identified Device or Use Problem; Temperature Problem

    THE CUSTOMER REPORTED THAT THAT THEY HAD 4 SEPARATE OCCURRENCES OF THE PULSE OXIMETER (SPO2 FINGER PROBE) LEAVING SKIN BLISTERS ON PATIENT FINGER. THIS IS FOR THE (B)(6) 2023 OCCURRENCE. THE ISSUE IS NOT FOR THE PSG-1100A SYSTEM ITSELF, AND IS FOR ACCESSORY SPO2 PROBE BEING USED WITH THE SYSTEM. OTHER RELATED COMPLAINTS WILL BE REPORTED: (B)(4). DIFFERENT PROBE MODEL BEING USED.

    Patient problems: Blister; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17561697 ↗

Other Full-montage Standard Electroencephalograph devices