Skip to content

Life Scope PT BSM-1733Recalls & reported problems

Nihon Kohden Corporation · Transport Monitor, Masimo SpO2 · Class 2

All models of Life Scope PT (3)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Nihon Kohden Life Scope PT transport monitor tracks vital signs such as blood oxygen. Of its 90 reports, 1 involves a death and 1 involves an injury. Reports mostly describe the monitor losing power or going blank, missing or inaccurate readings, and failing to start back up. One unit reportedly began smoking while in standby. Most reports describe no clinical signs or impact on the patient. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
93
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 93 adverse event reports that list this exact device (DI 04931921111833), received between Jan 18, 2017 and Sep 23, 2026 — classified as 1 death, 1 injury, 91 malfunctions.

The most frequently reported specific device problem is “Output Problem” (46 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 4 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 3 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 2 reportsOct 2025: 4 reportsNov 2025: 4 reportsDec 2025: 1 reportsJan 2026: 2 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 2 reportsJun 2026: 3 reportsJul 2026: 2 reportsAug 2026: 6 reportsSep 2026: 2 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output Problem46
  • Incorrect, Inadequate or Imprecise Result or Readings16
  • Unable to Obtain Readings14
  • Display or Visual Feedback Problem8
  • Application Program Problem6
  • No Device Output6
  • Power Problem6
  • Mechanical Problem5

Patient problems

  • No Clinical Signs, Symptoms or Conditions75
  • No Known Impact Or Consequence To Patient12
  • No Consequences Or Impact To Patient9
  • Cardiopulmonary Arrest1
  • No Information1
  • Sepsis1

Recent adverse event reports

The 25 most recent of 93 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026Unable to Obtain Readings; Pumping Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50873523 ↗
  • MalfunctionSep 23, 2026Incorrect Measurement

    The Biomedical Engineer (BME) reported the Bedside Monitor (BSM) is intermittently giving inaccurate non-invasive blood pressure (NIBP) readings, and the blood pressure pump sounds very weak. Customer indicated that multiple function tests were performed and the NIBP pump is failing when compared to a known good unit. No patient harm was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50867930 ↗
  • MalfunctionAug 31, 2026Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50519183 ↗
  • MalfunctionAug 27, 2026Electrical /Electronic Property Problem; Failure to Run on Battery

    The Biomedical Engineer (BME) reported that the Bedside Monitor (BSM) would power off when it was undocked from a Host/Main Bedside Monitor. No patient harm was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50477084 ↗
  • MalfunctionAug 19, 2026Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    The Biomedical Engineer (Biomed) reported that during setup the NIBP pump and SpO2 readings on the BSM-1733A were intermittently failing. No patient harm was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50326284 ↗
  • MalfunctionAug 18, 2026Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    The Biomedical Engineer (Biomed) reported that during setup the NIBP pump and SpO2 readings on the BSM-1733A were intermittently failing. No patient harm was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50309174 ↗
  • MalfunctionAug 12, 2026No Display/Image; Display or Visual Feedback Problem

    The Biomedical Engineer (BME) reported that the Bedside Monitor's screen was not displaying anything, even though the power light was green. No reports of patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50226201 ↗
  • MalfunctionAug 4, 2026Failure to Pump; Unable to Obtain Readings; Output Problem; Pumping Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50133289 ↗
  • MalfunctionJul 27, 2026Unable to Obtain Readings; Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50006140 ↗
  • MalfunctionJul 16, 2026Unable to Obtain Readings; Output Problem; Power Problem; Complete Loss of Power

    The Biomedical Engineer (Biomed) reported that while in patient use, the BSM-1733A was working but then for some reason, it stopped reading vitals. The BSM would then not power back on. No patient harm was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25840484 ↗
  • MalfunctionJun 11, 2026Smoking; Temperature Problem

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE BEDSIDE MONITOR (BSM) STARTED TO SMOKE. THE UNIT WAS IN STANDBY MODE AND NOT IN USE WITH A PATIENT. NO REPORTS OF HARM OR INJURY WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25478105 ↗
  • MalfunctionJun 10, 2026Unable to Obtain Readings; Output Problem

    THE BIOMEDICAL ENGINEER (BIOMED) REPORTED THAT THE NIBP PUMP WAS INTERMITTENTLY FAILING ON THE TRANSPORT MONITOR (BSM-1733A) WHILE MONITORING A PATIENT. BUT THE ISSUE PERSISTED. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25459426 ↗
  • MalfunctionJun 3, 2026Incorrect, Inadequate or Imprecise Result or Readings

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE BEDSIDE MONITOR (BSM) WAS GIVING INCONSISTENT AND INACCURATE NON-INVASIVE BLOOD PRESSURE (NIBP) READINGS. THERE WERE NO ERROR MESSAGES. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25385560 ↗
  • MalfunctionMay 19, 2026Electrical /Electronic Property Problem

    THE CUSTOMER REPORTED THAT THE BEDSIDE MONITOR (BSM) QI PACK WAS NOT ILLUMINATING THE WLAN OR LINK LIGHTS, ONLY THE POWER LIGHT WAS ON. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25215860 ↗
  • MalfunctionMay 8, 2026Power Problem; Complete Loss of Power

    THE BIOMEDICAL ENGINEER (BIOMED) REPORTED THAT THE TRANSPORT MONITOR (BSM-1733A) DID NOT REMAIN POWERED ON WHEN UNDOCKED FROM THE HOST MONITOR. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25124995 ↗
  • MalfunctionApr 7, 2026Output Problem

    THE BIOMEDICAL ENGINEER (BME) REPORTED THE BEDSIDE MONITOR (BSM) IS READING INCONSISTENTLY AND GIVING INCORRECT READINGS, VALIDATED INCORRECT WITH A PATIENT SIMULATOR. CUSTOMER NOTES THAT THE PLASTIC AROUND THE CONNECTOR IS DAMAGED CAUSING INTERMITTENT CONTACT ON ONE OF THE ECG PINS IN THE CONNECTOR INTEGRATED INTO THE DEVICE. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24811008 ↗
  • MalfunctionFeb 23, 2026Output Problem

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE BEDSIDE MONITOR WAS NOT TAKING FULL NIBP READINGS. THE NIBP WAS ONLY PUMPING TO 30-35 MMHG AND NO ERROR MESSAGES WERE DISPLAYED. TODAY, HOWEVER, THE DEVICE IS ABLE TO TAKE PROPER READINGS WITHOUT ANY ISSUES. CURRENTLY, THE DEVICE IS FUNCTIONING PROPERLY. HE WAS WONDERING WHAT MAY HAVE CAUSED THE ISSUE. SINCE THEY ARE UNABLE TO DUPLICATE THE PROBLEM, WE CANNOT DETERMINE THE ROOT CAUSE. TECH SUPPORT (TS) ADVISED HIM THAT IT IS POSSIBLE THERE WAS A FAULTY HOSE OR CUFF, OR THAT THE CONNECTIONS WERE NOT SECURED PROPERLY AT THE TIME. AT THIS TIME, HE IS UNABLE TO DUPLICATE THE ISSUE. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24423640 ↗
  • MalfunctionJan 12, 2026Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE BSM-1733A WOULD NOT DETECT ECG WHILE IN PATIENT USE. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24048356 ↗
  • MalfunctionJan 9, 2026Power Problem; Complete Loss of Power

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE BSM-1733A RANDOMLY SHUT DOWN DURING TRANSPORT. HE CHANGED THE BATTERIES AND AFTER 10 HOURS, THE DEVICE GOT VERY HOT. AT THE TIME OF THE CALL, THE BSM WOULD NOT POWER ON. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24026121 ↗
  • MalfunctionDec 16, 2025No Device Output; Output Problem; No Device Output; Output Problem

    THE CUSTOMER REPORTED THAT THE NIBP PUMP WILL NOT INITIATE WHEN SELECTING THE NIBP FUNCTION ON THEIR BSM-1733A BEDSIDE MONITOR (BSM). THE ISSUE WAS IDENTIFIED DURING A CASE. FORTUNATELY, THE PATIENT WAS ALSO BEING MONITORED WITH AN INVASIVE BLOOD PRESSURE LINE, ALLOWING AN ALTERNATIVE READING TO BE OBTAINED. THE ISSUE WAS BROUGHT TO THE BIOMED'S ATTENTION VIA A WORK ORDER. NO PATIENT HARM WAS REPORTED. THE BIOMED TESTED THE UNIT AS A STANDALONE DEVICE AND WAS ABLE TO DUPLICATE THE ISSUE IN HIS SHOP. HE STATED HE CANNOT HEAR THE TYPICAL CLICK OR PUMP ACTIVATION WHEN INITIATING NIBP. HE SWAPPED OUT THE CUFF AND HOSE, AND THE ISSUE PERSISTED. WHEN ASKED ABOUT CUFF TYPE, HE CONFIRMED THEY ARE USING WELCH ALLYN CUFFS BUT EMPHASIZED THE CUFF IS NOT THE ISSUE, AS THE SAME CUFF WORKED PROPERLY ON ANOTHER UNIT. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23824092 ↗
  • MalfunctionNov 26, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE BEDSIDE MONITOR (BSM) HAD ERRATIC NON-INVASIVE BLOOD PRESSURE (NIBP) READINGS. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23658094 ↗
  • MalfunctionNov 26, 2025Power Problem; Power Problem

    THE BIOMEDICAL ENGINEER (BME) REPORTED RECEIVING THE ERROR MESSAGE "CHECK INPUT UNIT BATTERY" ON A MAIN UNIT WHEN THEY DOCK THIS BEDSIDE MONITOR (BSM-1733). THE BSM-1733 WOULD NOT STAY POWERED WHEN IT WAS UNDOCKED FROM THE MAIN UNIT DESPITE USING A NEW CHARGED BATTERY. NOT IN PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23657595 ↗
  • MalfunctionNov 14, 2025Application Program Problem; Application Program Freezes or Fails to Launch; Application Program Problem; Application Program Freezes or Fails to Launch

    THE CUSTOMER REPORTED THAT THE TOUCHSCREEN WAS UNRESPONSIVE. IT WILL NOT CALIBRATE, AND TECH SUPPORT (TS) TRIED WALKING HER THROUGH THE PROCESS, BUT THE TOUCHSCREEN DOES NOT RESPOND AT ALL. NOT IN PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23562884 ↗
  • MalfunctionNov 12, 2025Output Problem; Output Problem

    THE CUSTOMER REPORTED THAT EXCESSIVE ARTIFACT / NOISE WAS DETECTED ON THE BSM-1733A DURING SETUP BY A NURSE. THERE WAS NO PATIENT HARM, AND THE PATIENT WAS TRANSFERRED TO ANOTHER BSM-1733A. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23536267 ↗
  • MalfunctionOct 28, 2025Display or Visual Feedback Problem; Image Display Error/Artifact; Output Problem; Display or Visual Feedback Problem; Image Display Error/Artifact; Output Problem

    THE NURSE REPORTED THAT THE INPUT UNIT SHOWED ECG ARTIFACTS. THEY WERE USING THIS UNIT WITH A G9 MONITOR AND A DOCKING STATION. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23403247 ↗

FDA clearance and classification

The FDA regulates product code MHX (monitor, physiological, patient(with arrhythmia detection or alarms)) under 21 CFR 870.1025 as a Class II device. 21 CFR 870.1025 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Life Scope PT BSM-1733

Has the Life Scope PT BSM-1733 been recalled?

No FDA recall record in our data names this exact device identifier (DI 04931921111833), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Life Scope PT BSM-1733?

The FDA's MAUDE database lists 93 adverse event reports for this device, received between Jan 18, 2017 and Sep 23, 2026. In those reports, the most common specific device problems are “Output Problem” (46), “Incorrect, Inadequate or Imprecise Result or Readings” (16) and “Unable to Obtain Readings” (14); the most common patient problems are “Cardiopulmonary Arrest” (1) and “Sepsis” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Life Scope PT BSM-1733?

1 report is classified as a death, 1 as an injury and 91 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Life Scope PT BSM-1733 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the Life Scope PT BSM-1733, and is it still sold?

The FDA's device database lists Nihon Kohden Corporation as the company responsible for it (brand Life Scope PT, model BSM-1733). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices

All MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page