Life Scope PT BSM-1773Recalls & reported problems

Nihon Kohden Corporation · Transport Monitor, NK SpO2 · Class 2

All models of Life Scope PT (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
17
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 04931921111956), received between May 30, 2018 and Apr 11, 2026 — classified as 8 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Compatibility Problem” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Compatibility Problem8
  • Device Operational Issue8
  • Thermal Decomposition of Device8
  • Communication or Transmission Problem2
  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Output Problem2
  • Power Problem2
  • Battery Problem1

Patient problems

  • Injury8
  • No Clinical Signs, Symptoms or Conditions8
  • Patient Problem/Medical Problem8
  • Pressure Sore/Ulcer8
  • Abrasion7
  • No Consequences Or Impact To Patient2
  • Bruise/Contusion1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 11, 2026Electrical /Electronic Property Problem; Battery Problem

    THE CUSTOMER REPORTED THAT THIS BEDSIDE MONITOR (BSM) DOES NOT RECOGNIZE THE BATTERY AND THE UNIT DISPLAYS A BATTERY ALARM. IT IS UNKNOWNN WHETHER THE UNIT WAS IN PATIENT USE AT THE TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24846030 ↗
  • MalfunctionSep 9, 2025Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings; Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE BEDSIDE MONITOR (BSM) WAS NOT TAKING ECG READINGS. NOT IN PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23003832 ↗
  • MalfunctionNov 10, 2023No Device Output; Output Problem; No Device Output; Output Problem

    THE NURSE REPORTED THAT THEY ARE HAVING INTERMITTENT NIBP ISSUES, AND ON (B)(6) 2023 THEY REPORTED THEY WERE UNABLE TO OBTAIN NIBP. THE PATIENT ALSO EXPIRED. QA HAS ASKED FOR THE SERIAL NUMBER OF THE BSM-1773A AND ADDITIONAL DEVICE INFORMATION, BUT THE CUSTOMER HAS BEEN UNRESPONSIVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18108830 ↗
  • MalfunctionJun 16, 2022Incorrect, Inadequate or Imprecise Result or Readings; Patient Data Problem; Incorrect, Inadequate or Imprecise Result or Readings; Patient Data Problem

    THE NURSE REPORTED THAT DURING A SURGERY THERE WAS A DISCREPANCY BETWEEN THE BEDSIDE MONITOR (BSM) AND THEIR ELECTRONIC MEDICAL RECORD (EMR). THE NURSE REPORTED THAT THE DISCREPANCY WAS BETWEEN THE SATURATION OF PERIPHERAL OXYGEN (SPO2) AND ELECTROCARDIOGRAM (ECG) HEART RATE WHICH IS OFF BY ABOUT 50. NO PATIENT HARM REPORTED. NIHON KOHDEN CLINICAL SUPPORT (NK CAS) INFORMED THE NURSE THAT THERE IS NO SETTINGS THAT WOULD CAUSE THAT DISCREPANCY AND RECOMMEDED THAT IT COULD POSSIBLY BE THE WAY THEIR VITALS ARE BEING MAPPED TO THEIR ELECTRONIC MEDICAL RECORDS (EMR) SYSTEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14715248 ↗
  • MalfunctionJan 23, 2021Communication or Transmission Problem; Communication or Transmission Problem

    THE CUSTOMER REPORTED THAT THEY HAVE INSTALLED FOUR NEW WIRELESS BEDSIDE MONITORS (BSM) AND THEY WOULD RANDOMLY GO INTO COMMUNICATION LOSS. THEY WOULD WORK FOR MOST OF THE TIME AND THEN RANDOMLY GET COMMUNICATION LOSS. THE CUSTOMER STATED THAT WHEN THEY SEE COMMUNICATION LOSS, THE MOMENT THEY TAKE THE UNITS OUT OF THE ROOM, ITS COMMUNICATION WOULD BE RESTORED. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11220350 ↗
  • MalfunctionJan 23, 2021Communication or Transmission Problem; Communication or Transmission Problem

    THE CUSTOMER REPORTED THAT THEY HAVE INSTALLED FOUR NEW WIRELESS BEDSIDE MONITORS ( BSM) AND THEY WOULD RANDOMLY GO INTO COMMUNICATION LOSS. THEY WOULD WORK FOR MOST OF THE TIME AND THEN RANDOMLY GET COMMUNICATION LOSS. THE CUSTOMER STATED THAT WHEN THEY SEE COMMUNICATION LOSS, THE MOMENT THEY TAKE THE UNITS OUT OF THE ROOM, ITS COMMUNICATION WOULD BE RESTORED. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11220088 ↗
  • MalfunctionFeb 13, 2020Output Problem; Power Problem; Complete Loss of Power; Output Problem; Power Problem; Complete Loss of Power

    THE CUSTOMER REPORTED THAT THEIR BEDSIDE MONITOR (BSM) DISPLAYED "INPUT UNIT FAILURE" DURING A PROCEDURE. THE CUSTOMER ALSO REPORTED THAT AFTER THEY REPLACED THE INPUT UNIT THEIR BSM AUDIBLY ALARMED AND DISPLAYED A BLUE SCREEN WITH ERROR CODES, AFTER WHICH IT SHUT DOWN ALL TOGETHER.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 9709096 ↗
  • MalfunctionJan 23, 2020Power Problem; Power Problem

    THE CUSTOMER REPORTED THAT THEIR BEDSIDE MONITOR (BSM) REBOOTED ON ITS OWN DURING A PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 9624714 ↗
  • MalfunctionJan 13, 2020Pressure Problem; Pumping Problem; Pressure Problem; Pumping Problem

    THE CUSTOMER REPORTED THAT THE BEDSIDE MONITOR'S (BSM) NON-INVASIVE BLOOD PRESSURE (NIBP) PUMP WAS SOMETIMES GOING TO "40" AND KEPT PUMPING (THE CUFF WOULD NOT DEFLATE). THE CUSTOMER ALSO REPORTED THAT THE PUMP SOMETIMES GENERATED A READING, AND OTHER TIMES GENERATED AN "AIR LEAK ERROR" MESSAGE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9583808 ↗
  • InjuryMay 30, 2018Thermal Decomposition of Device; Device Operational Issue; Compatibility Problem; Thermal Decomposition of Device; Device Operational Issue; Compatibility Problem

    THE NURSING MANAGER REPORTED THAT AN SPO2 PROBE CAUSED A BRUISE ON A NEONATE PATIENT.

    Patient problems: Bruise/Contusion; Injury; Patient Problem/Medical Problem; Pressure Sore/Ulcer

    MAUDE report 7555516 ↗
  • InjuryMay 30, 2018Thermal Decomposition of Device; Device Operational Issue; Compatibility Problem; Thermal Decomposition of Device; Device Operational Issue; Compatibility Problem

    THE NURSING MANAGER REPORTED THAT AN SPO2 PROBE CAUSED A REDDENED SKIN TEAR ON THE RIGHT WRIST OF A NEONATE PATIENT.

    Patient problems: Abrasion; Injury; Patient Problem/Medical Problem; Pressure Sore/Ulcer

    MAUDE report 7555510 ↗
  • InjuryMay 30, 2018Thermal Decomposition of Device; Device Operational Issue; Compatibility Problem; Thermal Decomposition of Device; Device Operational Issue; Compatibility Problem

    THE NURSING MANAGER REPORTED THAT AN SPO2 PROBE CAUSED A REDDENED PRESSURE AREA ON THE TOP OF THE RIGHT FOOT OF A NEONATE PATIENT.

    Patient problems: Abrasion; Injury; Patient Problem/Medical Problem; Pressure Sore/Ulcer

    MAUDE report 7555509 ↗
  • InjuryMay 30, 2018Thermal Decomposition of Device; Device Operational Issue; Compatibility Problem; Thermal Decomposition of Device; Device Operational Issue; Compatibility Problem

    THE NURSING MANAGER REPORTED THAT AN SPO2 PROBE CAUSED AN OPEN ABRASION ON THE LEFT FOOT OF A NEONATE PATIENT.

    Patient problems: Abrasion; Injury; Patient Problem/Medical Problem; Pressure Sore/Ulcer

    MAUDE report 7555507 ↗
  • InjuryMay 30, 2018Thermal Decomposition of Device; Device Operational Issue; Compatibility Problem; Thermal Decomposition of Device; Device Operational Issue; Compatibility Problem

    THE NURSING MANAGER REPORTED THAT AN SPO2 PROBE CAUSED A REDDENED PRESSURE AREA ON THE RIGHT FOOT OF A NEONATE PATIENT.

    Patient problems: Abrasion; Injury; Patient Problem/Medical Problem; Pressure Sore/Ulcer

    MAUDE report 7555503 ↗
  • InjuryMay 30, 2018Thermal Decomposition of Device; Device Operational Issue; Compatibility Problem; Thermal Decomposition of Device; Device Operational Issue; Compatibility Problem

    THE NURSING MANAGER REPORTED THAT AN SPO2 PROBE CAUSED A BLISTER ON THE RIGHT HAND OF A NEONATE PATIENT.

    Patient problems: Abrasion; Injury; Patient Problem/Medical Problem; Pressure Sore/Ulcer

    MAUDE report 7555501 ↗
  • InjuryMay 30, 2018Thermal Decomposition of Device; Device Operational Issue; Compatibility Problem; Thermal Decomposition of Device; Device Operational Issue; Compatibility Problem

    THE NURSING MANAGER REPORTED THAT AN SPO2 PROBE CAUSED A REDDENED PRESSURE AREA ON THE TOP OF THE LEFT FOOT OF A NEONATE PATIENT.

    Patient problems: Abrasion; Injury; Patient Problem/Medical Problem; Pressure Sore/Ulcer

    MAUDE report 7555500 ↗
  • InjuryMay 30, 2018Thermal Decomposition of Device; Device Operational Issue; Compatibility Problem; Thermal Decomposition of Device; Device Operational Issue; Compatibility Problem

    THE NURSING MANAGER REPORTED THAT AN SPO2 PROBE CAUSED AN OPEN ABRASION ON THE RIGHT FOOT OF A NEONATE PATIENT.

    Patient problems: Abrasion; Injury; Patient Problem/Medical Problem; Pressure Sore/Ulcer

    MAUDE report 7555499 ↗

Other MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices

All MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices →