Life Scope VS BSM-3552Recalls & reported problems

Nihon Kohden Corporation · Bedside Monitor, Low Acuity, Mellcor SpO2 · Class 2

All models of Life Scope VS (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
21
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 21 adverse event reports that list this exact device (DI 04931921113677), received between Jun 15, 2018 and May 19, 2026 — classified as 21 malfunctions.

The most frequently reported specific device problem is “Output Problem” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 3 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output Problem6
  • Communication or Transmission Problem4
  • Incorrect, Inadequate or Imprecise Result or Readings3
  • Power Problem3
  • Application Program Problem2
  • Complete Loss of Power2
  • Display or Visual Feedback Problem2
  • Protective Measures Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions19
  • No Consequences Or Impact To Patient2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 21 most recent of 21 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 19, 2026Electrical /Electronic Property Problem; Unexpected Shutdown

    THE CUSTOMER REPORTED THAT 3 MONITORS SHUTDOWN WHILE IN PATIENT USE AND WHEN THEY BOOTED BACK UP, THEY DISPLAYED AN ERROR. ROOMS 1 (SN (B)(6)), 6 (SN (B)(6)), AND 7 (SN (B)(6)) IN THE (B)(6) DEPARTMENT. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25216930 ↗
  • MalfunctionMay 18, 2026Power Problem; Complete Loss of Power

    THE BIOMEDICAL ENGINEER (BIOMED) REPORTED THAT THIS WAS ONE OF 3 LIFE SCOPE MONITORS THAT SHUT DOWN WHILE IN PATIENT USE AND THEN GAVE A "DEVICE ERROR" MESSAGE WHEN THEY BOOTED BACK UP. ALL 3 DEVICES DROPPED OUT AT THE SAME TIME. THE OTHER 2 DEVICES WERE REPORTED IN TICKETS 300433550 AND 300433765. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25206265 ↗
  • MalfunctionMay 18, 2026Power Problem; Complete Loss of Power

    THE BIOMEDICAL ENGINEER (BIOMED) REPORTED THAT THIS WAS ONE OF 3 LIFE SCOPE MONITORS THAT SHUT DOWN WHILE IN PATIENT USE AND THEN GAVE A "DEVICE ERROR" MESSAGE WHEN THEY BOOTED BACK UP. ALL 3 DEVICES DROPPED OUT AT THE SAME TIME. THE OTHER 2 DEVICES WERE REPORTED IN TICKETS (B)(4) AND (B)(4). NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25205279 ↗
  • MalfunctionApr 27, 2026Application Program Problem; Unintended Application Program Shut Down

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE BEDSIDE MONITOR (BSM) SPONTANEOUSLY REBOOTED WHILE IT WAS IN PATIENT USE. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24999811 ↗
  • MalfunctionDec 17, 2025Signal Artifact/Noise; Signal Artifact/Noise

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THEIR CLINICAL STAFF WITNESSED THIS BEDSIDE MONITOR (BSM) HAD ARTIFACTS AND SHOWING ECG READINGS THAT WERE UNSTABLE, RESULTING IN A TACHYCARDIA ALARM. NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23828485 ↗
  • MalfunctionNov 24, 2025Application Program Problem; Unintended Application Program Shut Down; Application Program Problem; Unintended Application Program Shut Down

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE LIFE SCOPE MONITOR WAS IDLING ON A STANDBY SCREEN WHEN IT STARTED ALARMING AND REBOOTING, SHOWING A BLUE BLOCK MEMORY ERROR SCREEN. TECHNICAL SUPPORT (TS) TRIED TROUBLESHOOTING WITH THE BME, BUT WHEN TRYING TO LOAD THE SOFTWARE, IT GAVE A "WATCHDOG ERROR" MESSAGE BEFORE TRYING TO REBOOT. IT WOULD THEN OCCASIONALLY LOAD THE MONITORING SCREEN AND ALARM AND THEN DISPLAY THE BLUE BLOCK MEMORY ERRORS AGAIN. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23639740 ↗
  • MalfunctionJul 9, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    THE CUSTOMER REPORTED THAT THE NON-INVASIVE BLOOD PRESSURE (NIBP) PORT WILL NOT HOLD THE HOSE ANYMORE. THE UNIT WAS NOT IN PATIENT USE AT THE TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22460152 ↗
  • MalfunctionJun 9, 2025Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE CUSTOMER REPORTED THAT THE BEDSIDE MONITOR WAS NOT PRODUCING ACCURATE SPO2 PULSE RATE READINGS AS COMPARED TO THE ECG HEART RATE READINGS. THEY DO NOT HAVE A SIMULATOR TO TEST THE UNIT. THERE WAS NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22171421 ↗
  • MalfunctionNov 4, 2024Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE CUSTOMER REPORTED THAT THIS BEDSIDE MONITOR (BSM-3500) IS TAKING INACCURATE ECG READINGS. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20604774 ↗
  • MalfunctionMay 5, 2022Power Problem; Power Problem

    THE CUSTOMER REPORTED THAT THIS BEDSIDE MONITOR (BSM) WAS MAKING A LOUD SCREECHING NOISE AFTER WHICH THE UNIT SHUT DOWN AND REBOOT ITSELF. AT THE TIME, THE BEDSIDE WAS DISPLAYING A "PRESS F4 TO START" ERROR. NO HARM OR INJURY OCCURRED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14289137 ↗
  • MalfunctionFeb 24, 2022Output Problem; Output Problem

    THE CUSTOMER REPORTED THAT THEIR BEDSIDE MONITOR (BSM) IS PRODUCING INCORRECT READINGS. THEY ALSO STATED THAT THE READINGS ARE GOING UP AND DOWN, BUT NO ALARMS ARE BEING TRIGGERED. NO HARM OR INJURY OCCURRED. THE ISSUE WAS RESOLVED WHEN THE CUSTOMER SWAPPED PROBES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13595925 ↗
  • MalfunctionJan 7, 2022Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE CUSTOMER REPORTED THAT THEIR BEDSIDE MONITOR (BSM) IS SHOWING INCORRECT ECG READINGS. THEY CLARIFIED THAT THE BEDSIDE WAS FALSELY ALARMED FOR TACHYCARDIA, BUT WHEN THEY MOVED THE PATIENT TO ANOTHER UNIT, THEY REALIZED THAT THE PREVIOUS READING WAS INCORRECT. NO HARM OR INJURY OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13197667 ↗
  • MalfunctionJan 7, 2022Output Problem; Output Problem

    THE CUSTOMER REPORTED THAT THEIR BEDSIDE MONITOR (BSM) IS SHOWING INCORRECT ECG READINGS. THEY CLARIFIED THAT THE BEDSIDE WAS FALSELY ALARMED FOR TACHYCARDIA, BUT WHEN THEY MOVED THE PATIENT TO ANOTHER UNIT, THEY REALIZED THAT THE PREVIOUS READING WAS INCORRECT. NO HARM OR INJURY OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13197665 ↗
  • MalfunctionSep 20, 2021Protective Measures Problem; Protective Measures Problem

    THE CUSTOMER REPORTED THAT THE LOW SPO2 ALARM LIMIT FOR THE BEDSIDE MONITOR (BSM) WAS CONFIGURED TO ALARM WHEN IT DIPPED TO 88, BUT IT DID NOT ALARM UNTIL 80. NO HARM OR INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12497441 ↗
  • MalfunctionSep 20, 2021Protective Measures Problem; Protective Measures Problem

    THE CUSTOMER REPORTED THAT THEIR BEDSIDE MONITOR (BSM) ALARM WAS CONFIGURED TO ALARM AT 88, BUT DIDN'T ALARM UNTIL 80. NO HARM OR INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12497422 ↗
  • MalfunctionJun 9, 2021Communication or Transmission Problem; Communication or Transmission Problem

    THE CUSTOMER REPORTED THAT THE BEDSIDE MONITORS (BSMS) RECENTLY STOPPED BEING DISPLAYED AT THE CENTRAL NURSE'S STATION (CNS). ACCORDING TO THE CUSTOMER, BOTH UNITS SHOW COMM LOSS ON THE CNS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11971379 ↗
  • MalfunctionJun 9, 2021Communication or Transmission Problem; Communication or Transmission Problem

    THE CUSTOMER REPORTED THAT THE BEDSIDE MONITORS (BSMS) RECENTLY STOPPED BEING DISPLAYED AT THE CENTRAL NURSE'S STATION (CNS). ACCORDING TO THE CUSTOMER, BOTH UNITS SHOW COMM LOSS ON THE CNS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11971373 ↗
  • MalfunctionDec 2, 2020Erratic or Intermittent Display; Display or Visual Feedback Problem; Output Problem; Erratic or Intermittent Display; Display or Visual Feedback Problem; Output Problem

    THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THE BEDSIDE MONITOR (BSM) WOULD DISPLAY SPO2; AND THEN SUDDENLY STOP DISPLAYING IT, APPEARING AS IF NO CABLE WAS PLUGGED IN. THERE WERE NO ERROR MESSAGES ON THE BSM WHEN THIS OCCURED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10936092 ↗
  • MalfunctionJul 31, 2019Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THAT THE BEDSIDE MONITOR (BSM) DROPPED OUT OF NETWORK DURING PATIENT MONITORING ACTIVITIES. NO CONSEQUENCE OR IMPACT TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 8848957 ↗
  • MalfunctionJul 31, 2019Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THAT THE BEDSIDE MONITOR (BSM) DROPPED OUT OF NETWORK DURING PATIENT MONITORING ACTIVITIES. NO CONSEQUENCE OR IMPACT TO THE PATIENT WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8848908 ↗
  • MalfunctionJun 15, 2018Failure to Analyze Signal; Device Issue; Device Operational Issue; Device Sensing Problem; Failure to Analyze Signal; Device Issue; Device Operational Issue; Device Sensing Problem

    THE NURSE CALLED TO ASK QUESTIONS ABOUT THE BEDSIDE MONITOR (BSM) SPO2 SETTINGS. SHE REPORTED THAT ANOTHER NURSE WAS COMPLAINING THAT THE SPO2 DID NOT ALARM WHEN THE VALUE DROPPED BELOW THE PRESET.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7606224 ↗

Other MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices

All MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) devices →