ULTRASONIC GASTROVIDEOSCOPE OLYMPUS GF TYPE UC140P-AL5 (UDI 04953170062803)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of ULTRASONIC GASTROVIDEOSCOPE (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04953170062803), received between Aug 6, 2020 and Jan 16, 2026 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Peeled/Delaminated” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Peeled/Delaminated3
- Biofilm coating in Device1
- Break1
- Detachment of Device or Device Component1
- Failure to Clean Adequately1
- Fluid/Blood Leak1
- Mechanical Problem1
- Microbial Contamination of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient2
- No Patient Involvement1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 16, 2026Detachment of Device or Device Component
IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE PROBE ON THE ULTRASONIC GASTROVIDEOSCOPE WAS REMOVED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24092685 ↗MalfunctionJun 4, 2024Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE ULTRASONIC GASTROVIDEOSCOPE FORCEPS ELEVATOR WAS NOT WORKING AND THE WIRE FOR THE ELEVATOR WAS LOOSE. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19459786 ↗MalfunctionNov 22, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED TO OLYMPUS, THE OLYMPUS ULTRASONIC GASTROVIDEOSCOPE HAD AIR/WATER LEAKAGES FROM THE DISTAL END. THE DEVICE WAS RETURNED FOR EVALUATION. DURING THE DEVICE EVALUATION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE FOUND: THERE WAS A GAP OF ADHESIVE ON THE DISTAL END DUE TO THE ADHESIVE PEELING. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18192694 ↗MalfunctionMay 14, 2022Failure to Clean Adequately; Failure to Clean Adequately
A USER FACILITY RETURNED THE OLYMPUS, MODEL GF-UC140P-AL5, ULTRASONIC GASTROVIDEOSCOPE TO OLYMPUS FOR REPAIR DUE TO A REPORT OF "DISTAL END LEAK." UPON INSPECTION AND TESTING OF THE RETURNED DEVICE, IT WAS DETERMINED THAT THE SCOPE WAS INSUFFICIENTLY OR INCORRECTLY REPROCESSED AND THE SCOPE WAS USED FOR NEXT PROCEDURE. THIS REPORT IS SUBMITTED DUE TO THE SCOPE WAS INSUFFICIENTLY OR INCORRECTLY REPROCESSED AND THE SCOPE WAS USED FOR NEXT PROCEDURE. THERE WAS NO PATIENT INJURY, ASSOCIATED WITH THE PROBLEM, REPORTED TO OLYMPUS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14403028 ↗MalfunctionAug 11, 2021Microbial Contamination of Device; Microbial Contamination of Device
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF ROUTINE MICROBIOLOGICAL TESTING BY THE USER FACILITY, FOLLOWING MICROBES WERE DETECTED FROM THE SAMPLE COLLECTED FROM THE SUBJECT DEVICE. COAGULASE-NEGATIVE STAPHYLOCOCCI (>100 CFU/ENDOSCOPE). THE DEVICE HAD BEEN REPROCESSED WITH A NON-OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR, WASSENBURG WD440 ADAPTASCOPE, USING PERACETIC ACID. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12306277 ↗MalfunctionOct 7, 2020Poor Quality Image; Peeled/Delaminated; Poor Quality Image; Peeled/Delaminated
AS REPORTED FOR THIS EVENT, DURING AN UNKNOWN PROCEDURE THERE WAS BAD ULTRASOUND IMAGE. THERE IS NO REPORTED ADVERSE IMPACT OR HARM TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10646282 ↗MalfunctionOct 7, 2020Break; Fluid/Blood Leak; Peeled/Delaminated; Break; Fluid/Blood Leak; Peeled/Delaminated
AS REPORTED FOR THIS EVENT, DURING REPROCESSING THE BLACK TIP OF THE INSERTION TUBE OF THE DEVICE IS READY TO FALL OFF. THERE IS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 10645995 ↗MalfunctionAug 6, 2020Biofilm coating in Device; Biofilm coating in Device
THE CUSTOMER REPORTED TO OLYMPUS A SCOPE THAT HAD A POSITIVE CULTURE FOUND DURING MAINTENANCE. THERE WAS NO PATIENT INJURY, PATIENT INFECTION OR DEATH REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10376057 ↗
Other Transducer, ultrasonic, diagnostic devices
- Site~Rite 8 Ultrasound System 9770603Bard Access Systems, Inc. · Transducer, ultrasonic, diagnostic620 reports
- ULTRASONIC PROBE UM-S20-17SOlympus Medical Systems Corp. · Transducer, ultrasonic, diagnostic2 deaths344 reports
- Transducer X7-2t 989605414122Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic6 recalls21 reports
- EPIQ Diagnostic Ultrasound System 989605488741Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic3 recalls165 reports
- bk3000, bk3500, bk5000, bkActiv 2300Bk Medical Aps · Transducer, Ultrasonic, Diagnostic6 recalls1 report
- Transducer L12-3 989605412072Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls1 report
- Transducer L15-7io 989605387852Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls0 reports
- Transducer eL18-4 EMT 989605431531Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page