ULTRASONIC PROBE UM-2R (UDI 04953170063206)Recalls & reported problems
Olympus Medical Systems Corp. · "6890130 OLA UM-2R-3 ULTRASONIC PROBE 12M" · Class 2
All models of ULTRASONIC PROBE (17)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04953170063206), received between Nov 17, 2022 and Apr 22, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Twisted/Bent2
- Device Reprocessing Problem1
- Material Deformation1
- No Display/Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 22, 2026Material Twisted/Bent
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC PROBE INSERTION SHEATH WAS TWISTED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24955080 ↗MalfunctionMar 17, 2026Material Deformation; Material Twisted/Bent
IT WAS REPORTED THE ULTRASONIC PROBE WAS KINKED. THE ISSUE OCCURRED DURING MAINTENANCE OF THE DEVICE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24617194 ↗MalfunctionAug 17, 2023Device Reprocessing Problem; Device Reprocessing Problem
THE OLYMPUS MEDICAL SERVICE DEPARTMENT DETERMINED THAT THE CUSTOMER WAS MISSING A REQUIRED COMPONENT DURING REPROCESSING OF THE ULTRASONIC PROBE WHILE USING AN OLYMPUS ENDOSCOPE REPROCESSOR (OER-3). THE ISSUE WAS IDENTIFIED DURING REPROCESSING OF THE DEVICE. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM HAS BEEN REPORTED. THE OER-3 OLYMPUS ENDOSCOPE REPROCESSOR WAS REPORTED IN THE MEDWATCH WITH PATIENT IDENTIFIER (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17565504 ↗MalfunctionNov 17, 2022No Display/Image; No Display/Image
WHILE AN OLYMPUS FIELD SERVICE ENGINEER (FSE) WAS AT THE USER¿S FACILITY, HE REPORTED THAT THE CUSTOMER¿S ULTRASONIC PROBE WAS NOT PRODUCING AN IMAGE DURING A PROCEDURE. ACCORDING TO THE FSE, THE CUSTOMER FINISHED THE PROCEDURE USING ANOTHER DEVICE WITH MINIMAL DELAYS AND NO PATIENT HARM. DURING THE DEVICE EVALUATION, IT WAS DISCOVERED THAT THERE WAS A BREAKAGE IN THE INSERTION TUBE. THIS REPORT IS BEING SUBMITTED TO CAPTURE THE DAMAGE FOUND IN THE INSERTION TUBE DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15814332 ↗
Other Transducer, ultrasonic, diagnostic devices
- Site~Rite 8 Ultrasound System 9770603Bard Access Systems, Inc. · Transducer, ultrasonic, diagnostic620 reports
- ULTRASONIC PROBE UM-S20-17SOlympus Medical Systems Corp. · Transducer, ultrasonic, diagnostic2 deaths344 reports
- Transducer X7-2t 989605414122Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic6 recalls21 reports
- EPIQ Diagnostic Ultrasound System 989605488741Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic3 recalls165 reports
- bk3000, bk3500, bk5000, bkActiv 2300Bk Medical Aps · Transducer, Ultrasonic, Diagnostic6 recalls1 report
- Transducer L12-3 989605412072Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls1 report
- Transducer eL18-4 EMT 989605431531Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls0 reports
- Transducer L12-3 Ergo 989605492481Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page