ULTRASONIC PROBE UM-3R (UDI 04953170063213)Recalls & reported problems
Olympus Medical Systems Corp. · "6890230 OLA UM-3R-3 ULTRASONIC PROBE 20M" · Class 2
All models of ULTRASONIC PROBE (17)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04953170063213), received between Dec 22, 2020 and Mar 27, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Fluid/Blood Leak2
- Contamination /Decontamination Problem1
- Crack1
- Material Puncture/Hole1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 27, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT AFTER THE ULTRASONIC PROBE WAS CLEANED AND DISINFECTED DURING REPROCESSING, YELLOW LIQUID ADHERED WHEN THE PROBE DEVICE WAS WIPED WITH GAUZE. THE YELLOW LIQUID ADHERED EVEN AFTER WIPING IT OFF SEVERAL TIMES. THE FACILITY'S CLEANING METHOD IS AS FOLLOWS: CLEANING USING ENDOPURE, IMMERSION IN DISOPA, FLUSHING WITH WATER, AND WIPING WITH DISINFECTANT ETHANOL SOLUTION B IP ON GAUZE. YELLOW LIQUID ADHERED AT THE LAST STEP IN THE CLEANING PROCESS. THERE WAS NO PATIENT INVOLVEMENT. THERE WAS NO REPORT THE DEVICE WAS USED IN A PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21712275 ↗MalfunctionDec 5, 2024Crack; Crack
IT WAS REPORTED THAT ULTRASONIC PROBE HAD A CRACK ON ITS EXTERIOR, LOCATED MIDWAY ALONG THE PROBE. THE ISSUE OCCURRED DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20855232 ↗MalfunctionMar 21, 2023Break; Break
THE CUSTOMER REPORTED TO OLYMPUS, DURING MAINTENANCE, THE ULTRASONIC PROBE HAD DEFECTIVE IMAGE. UPON INSPECTION OF THE CUSTOMER¿S RETURNED DEVICE IT WAS OBSERVED, BREAKAGE OF THE INSERTION TUBE WAS NOTED. THIS REPORT IS BEING SUBMITTED FOR THE MALFUNCTION FOUND DURING EVALUATION. THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16584813 ↗MalfunctionDec 7, 2022Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
AN OLYMPUS EMPLOYEE REPORTED ON BEHALF OF THE CUSTOMER, THE ULTRASONIC PROBE ULTRASOUND IMAGES ARE UNCLEAR AND DIFFICULT TO SEE DURING SURGERY. THE FREQUENCY IS NOT HIGH. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THIS EVENT. DURING INCOMING INSPECTION, IT WAS DETERMINED THERE WAS LEAKAGE IN THE ULTRASONIC MEDIUM DUE TO PERFORATION IN THE TIP SHEATH. THIS MEDWATCH IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15933607 ↗MalfunctionApr 5, 2022Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THAT WHEN PREPARING THE DEVICE FOR USE DURING A DIAGNOSTIC ULTRASONOGRAPHY PROCEDURE, THE SUBJECT DEVICE HAD A LEAKAGE OF ULTRASONIC MEDIUM FROM THE INSERTION TUBE. THE ISSUE WAS IDENTIFIED DURING THE FIRST USE AFTER DELIVERY TO THE FACILITY. THERE WAS NO PATIENT OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14018409 ↗MalfunctionDec 22, 2020Break; Break
THE CUSTOMER FOUND FLUID COMING OUT OF THE DISTAL TIP OF THE DEVICE, WHICH WAS AT THE STORAGE AREA. THIS IS A SIGN OF DEVICE BREAKAGE. OTHER DETAILED INFORMATION WAS NOT PROVIDED. THERE WAS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11061596 ↗
Other Transducer, ultrasonic, diagnostic devices
- Site~Rite 8 Ultrasound System 9770603Bard Access Systems, Inc. · Transducer, ultrasonic, diagnostic620 reports
- ULTRASONIC PROBE UM-S20-17SOlympus Medical Systems Corp. · Transducer, ultrasonic, diagnostic2 deaths344 reports
- Transducer X7-2t 989605414122Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic6 recalls21 reports
- EPIQ Diagnostic Ultrasound System 989605488741Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic3 recalls165 reports
- bk3000, bk3500, bk5000, bkActiv 2300Bk Medical Aps · Transducer, Ultrasonic, Diagnostic6 recalls1 report
- Transducer L12-3 989605412072Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls1 report
- Transducer eL18-4 EMT 989605431531Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls0 reports
- Transducer L12-3 Ergo 989605492481Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page