Skip to content

DISPOSABLE ELECTROSURGICAL SNARE SD-240U-25 (REF N1083430)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of DISPOSABLE ELECTROSURGICAL SNARE (13)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04953170165986), received between Apr 17, 2019 and Sep 8, 2023 — classified as 4 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Difficult to Remove2
  • Failure to Cut2
  • Adverse Event Without Identified Device or Use Problem1
  • Therapeutic or Diagnostic Output Failure1
  • Use of Device Problem1

Patient problems

  • Bowel Perforation3
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1
  • Perforation1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 8, 2023Difficult to Remove; Failure to Cut; Difficult to Remove; Failure to Cut

    AN OLYMPUS EMPLOYEE REPORTED ON BEHALF OF A CUSTOMER, DURING AN UNSPECIFIED PROCEDURE, THE DISPOSABLE ELECTROSURGICAL SNARE COULD NOT CUT THROUGH A LARGE POLYP (25 MILLIMETER). ALSO, THE SNARE LOOP COULD NOT BE REMOVED FROM THE TARGET TISSUE. THE HANDLE WAS CUT OFF SNARE AND THE DOCTOR WAS ABLE TO PUSH THE SCOPE FORWARD TO RELEASE SNARE AND REMOVE SCOPE. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17712703 ↗
  • InjuryAug 21, 2019Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY THE USER FACILITY THAT DURING A POLYP REMOVAL PROCEDURE USING A SD-240U-25 DISPOSABLE ELECTROSURGICAL SNARE, THE POLYP COULD NOT BE REMOVED. THE CAUTERY UNIT AND THE SNARE WERE CHANGED TWICE, WITHOUT SUCCESS. THE PATIENT WAS TRANSFERRED TO THE HOSPITAL AND UNDERWENT A COLON RESECTION WITH A ILEOSTOMY SURGERY TO REPAIR THE PERFORATION. THE PATIENT DISCHARGED TWO WEEKS LATER. NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME.

    Patient problems: Bowel Perforation

    MAUDE report 8917762 ↗
  • InjuryAug 19, 2019Use of Device Problem; Use of Device Problem

    IT WAS REPORTED BY THE USER FACILITY THAT DURING A POLYP REMOVAL PROCEDURE USING A SD-240U-25 DISPOSABLE ELECTROSURGICAL SNARE, THE POLYP COULD NOT BE REMOVED. THE CAUTERY UNIT AND THE SNARE WERE CHANGED TWICE, WITHOUT SUCCESS. THE PATIENT WAS TRANSFERRED TO THE HOSPITAL AND UNDERWENT A COLON RESECTION WITH A ILEOSTOMY SURGERY TO REPAIR THE PERFORATION. THE PATIENT DISCHARGED TWO WEEKS LATER. NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME.

    Patient problems: Perforation

    MAUDE report 8906772 ↗
  • InjuryAug 19, 2019Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT DURING A POLYP REMOVAL USING A SD-240U-25, THE POLYP COULD NOT BE REMOVED. THE PHYSICIAN CHANGED THE CAUTERY UNIT AND SNARE DEVICE TWICE WITHOUT SUCCESS. THE PATIENT WAS TRANSFERRED TO THE HOSPITAL FOR ADDITIONAL MEDICAL INTERVENTION DUE TO BOWEL PERFORATION. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Bowel Perforation

    MAUDE report 8906308 ↗
  • InjuryAug 19, 2019Failure to Cut; Failure to Cut

    THE USER FACILITY REPORTED THAT THE PHYSICIAN WAS USING A SD-240U-25 FOR A POLYPECTOMY, BUT THE POLYP COULD NOT BE REMOVED. THE SNARE AND THE CAUTERY UNIT WERE CHANGED TWICE WITHOUT SUCCESS. THE PATIENT HAD TO BE TRANSFERRED TO THE HOSPITAL DUE TO A BOWEL PERFORATION, AND REMAINED IN THE HOSPITAL 2 WEEKS FOR RECOVERY. THE PATIENT WAS DISCHARGED FROM THE HOSPITAL TO HOME.

    Patient problems: Bowel Perforation

    MAUDE report 8906277 ↗
  • MalfunctionApr 17, 2019Difficult to Remove; Difficult to Remove

    OLYMPUS WAS INFORMED THAT DURING A POLYPECTOMY PROCEDURE, THE SNARE BECAME STUCK THE PATIENT¿S POLYP. THE POLYP WAS CONSIDERED TO BE MODERATE IN SIZE AND FIRM. THERE WAS NO BLEEDING OBSERVED. THE PHYSICIAN USED PLIERS AND APPLIED FORCE TO REMOVE THE SNARE LOOP. THE POLYP WAS CLOSE TO THE PATIENT¿S RECTUM, SO THE PHYSICIAN WAS ABLE TO REMOVE IT. THE INTENDED PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THE PATIENT'S CURRENT CONDITION IS STABLE. ADDITIONALLY, THE SNARE WAS TESTED WITH CURRENT PRIOR TO THE PROCEDURE WITH NO ANOMALIES NOTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8525484 ↗

Other SNARE, FLEXIBLE devices

All SNARE, FLEXIBLE devices →

Sources and updates