BALLOON CONTROL UNIT OBCU (UDI 04953170239533)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 04953170239533), received between Apr 5, 2022 and May 17, 2026 — classified as 3 injuries, 12 malfunctions.
The most frequently reported specific device problem is “Pressure Problem” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Pressure Problem6
- Adverse Event Without Identified Device or Use Problem3
- Improper Flow or Infusion2
- Break1
- Decrease in Pressure1
- Deflation Problem1
- Inflation Problem1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- Unspecified Tissue Injury2
- Erosion1
- Fever1
- Hemorrhage/Blood Loss/Bleeding1
- Perforation1
- Peritonitis1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 17, 2026Improper Flow or Infusion
IT WAS REPORTED THAT THE BALLOON CONTROL UNIT HAD AIR PRESSURE UNSTABLE. THE ISSUE WAS FOUND DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25196851 ↗MalfunctionApr 28, 2026Pressure Problem
IT WAS REPORTED THAT THE BALLON CONTROL UNIT HAD AIR PRESSURE WAS NOT HOLDING. THE ISSUE OCCURRED DURING SERVICE AND MAINTENANCE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25016273 ↗MalfunctionOct 28, 2025Improper Flow or Infusion; Improper Flow or Infusion
IT WAS REPORTED THE SUBJECT DEVICE HAD NO PRESSURE DISPLAY DURING THERAPEUTIC ENTEROSCOPY PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23407638 ↗MalfunctionSep 15, 2025Pressure Problem; Pressure Problem
IT WAS REPORTED THAT THE SUBJECT DEVICE UNABLE TO BUILD UP PRESSURE. PRESSURE UNSTABLE. THE EVENT OCCURRED DURING SET UP BEFORE AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23052095 ↗MalfunctionSep 6, 2025Decrease in Pressure; Decrease in Pressure
IT WAS REPORTED THE SUBJECT DEVICE'S NEGATIVE PRESSURE IS INSUFFICIENT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22988968 ↗MalfunctionAug 22, 2024Pressure Problem; Pressure Problem
IT WAS REPORTED THAT THE ALARM WAS TRIGGERED WHEN THE MAXIMUM PRESSURE WAS EXCEEDED AND PRESSING THE BUTTON ON THE BALLOON CONTROL UNIT NEITHER STOPPED THE ALARM NOR DEFLATED THE BALLOON. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE, AND THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20028345 ↗MalfunctionMar 19, 2024Pressure Problem; Pressure Problem
IT WAS REPORTED, THE BALLOON CONTROL UNIT HAD BALLOON/PRESSURE SENSOR ISSUES, SPONTANEOUSLY DROPPED TO ZERO PRESSURE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18932480 ↗MalfunctionNov 1, 2023Pressure Problem; Pressure Problem
THE CUSTOMER REPORTED TO OLYMPUS THAT AT THE END OF AN ENTEROSOCOPY FOR ANEMIA, USING THIS BALLOON CONTROL UNIT, INFLATION PROBLEM, ABNORMAL PRESSURE WAS OBSERVED. IT PROVIDED A NEGATIVE 2 AND A DEFLATED BALLOON BUT WHEN THE HOSE WAS REMOVED, THE BALLOON WAS STILL INFLATED RESULTING IN MINIMAL MUCOSAL EROSION. THERE WAS NO DELAY IN THE PROCEDURE. NO CLINICAL AND THERAPEUTIC FINDINGS, THE PATIENT WAS DISCHARGED THE NEXT DAY. THE DEVICE WAS CHECKED BEFORE USE. ADDITIONAL INFORMATION HAS BEEN REQUESTED, BUT NO INFORMATION HAS BEEN RECEIVED AT THIS TIME.
Patient problems: Erosion
MAUDE report 18048654 ↗MalfunctionOct 12, 2023Pressure Problem; Pressure Problem
THE OLYMPUS REPRESENTATIVE REPORTED (ON BEHALF OF THE CUSTOMER) THAT THE BALLOON CONTROL UNIT OLYMPUS ASSET AIR PRESSURE VALUE ABNORMAL. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17924533 ↗MalfunctionSep 21, 2023No Audible Alarm; Inflation Problem; No Audible Alarm; Inflation Problem
NO CLEANING, DISINFECTION, AND STERILIZATION CONDITIONS BECAUSE IT IS A BOX ITEM. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE AND THE FREQUENCY OF USE OR NUMBER OF TIMES OF USE IS UNKNOWN. THE PURPOSE OF THE PROCEDURE WAS INSPECTION. INFORMATION NOT AVAILABLE ON WHAT CONGESTION TO THE MUCOUS MEMBRANE IS. NO ADDITIONAL INTERVENTION OR ADDITIONAL TREATMENT. NO HEALTH DAMAGE TO THE PATIENT DUE TO THE EVENT. ADDITIONALLY, INFORMATION RECEIVED UNSURE IF "BALLOON AIR CANNOT BE STOPPED" BUT IT DOES NOT STOP EVEN IF YOU PRESS THE PUMP PAUSE SWITCH OR IF "THE ALARM DOES NOT SOUND" THIS ONLY OCCURS IF INFLATED FOR 20 SECONDS OR MORE. UNABLE TO OBTAIN THE EXACT TIME FOR THE POWER TO BE TURNED OFF. WHEN THE USER CONNECTED THE BALLOON, IT SWELLED, AND THERE WERE NO SCRATCHES AND THERE SEEMED TO BE NO AIR LEAKAGE. THE USER DOESN¿T REMEMBER IF THERE WERE WARNING INDICATOR LAMPS ON THE MAIN UNIT AND/OR OBCU REMOTE CONTROLLER TIPPED OVER WHEN THE PHENOMENON OCCURRED.
Patient problems: Unspecified Tissue Injury
MAUDE report 17789900 ↗MalfunctionSep 8, 2023Deflation Problem; Deflation Problem
THE CUSTOMER REPORTED TO OLYMPUS, THE BALLOON CONTROL UNIT AUTOMATICALLY DEFLATES ON ITS OWN. THERE WERE NO REPORTS OF PATIENT HARM. NO FURTHER INFORMATION WAS PROVIDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17712877 ↗MalfunctionDec 29, 2022Break; Mechanical Problem; Output above Specifications; Break; Mechanical Problem; Output above Specifications
THE CUSTOMER REPORTED AN ABNORMAL SOUND (LOANER FROM OLYMPUS). NO HARM WAS REPORTED. NO PATIENT HARM, NO USER INJURY REPORTED AS THE RESULT OF THE REPORTED ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16070322 ↗InjuryJun 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT IS REPORTED IN THE LITERATURE TITLED: "SINGLE BALLOON ENTEROSCOPY: IS IT FEASIBLE? PATIENTS EXPERIENCED ADVERSE EVENTS AFTER PROCEDURES USING OLYMPUS DEVICES. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS OBCU. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS ST-SB1. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS SIF-Q260. BACKGROUND/AIMS. CURRENTLY, TWO ENDOSCOPIC MODALITIES ARE USED TO INVESTIGATE THE SMALL INTESTINE. CAPSULE ENDOSCOPY (CE) HAS BEEN USED MAINLY AS A SCREENING TOOL FOR EVALUATING OBSCURE GASTROINTESTINAL BLEEDING. HOWEVER, CE ALSO HAS SEVERAL DISADVANTAGES, SUCH AS INADEQUATE MANIPULATION IN THE LUMEN, THE INABILITY TO TAKE TISSUE SAMPLES, AND THE INABILITY TO TREAT INTESTINAL STRICTURES. DOUBLE-BALLOON ENTEROSCOPY (DBE) AND ITS INSERTION TECHNIQUE WERE FIRST PUBLISHED BY YAMAMOTO AND COWORKERS IN 2001. ALTHOUGH DBE IS MORE INVASIVE THAN CE, DBE IS CLINICALLY A VERY USEFUL TOOL NOT ONLY FOR DIAGNOSIS, BUT ALSO FOR ENDOSCOPIC THERAPY IN THE SMALL INTESTINE.5-15 ALTHOUGH THE METHOD FOR DBE INSERTION HAS ALREADY BEEN ESTABLISHED, THERE ARE A FEW PROBLEMS IN TERMS OF PREPARATION AND INSERTION FOR DBE AS COMPARED WITH CONVENTIONAL ENDOSCOPIES. FOR EXAMPLE, DURING PREPARATION,
Patient problems: Fever; Peritonitis; Unspecified Tissue Injury
MAUDE report 14848615 ↗InjuryJun 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT IS REPORTED IN THE LITERATURE TITLED ¿IMPACT OF SCOPE EXCHANGE FROM A LONG SINGLE BALLOON ENTEROSCOPE TO A GASTROSCOPE DURING ENTEROSCOPY-ASSISTED ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY IN PATIENTS WITH SURGICALLY ALTERED ANATOMY,¿ PATIENTS EXPERIENCED ADVERSE EFFECTS DURING OR AFTER PROCEDURES USING OLYMPUS DEVICES. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS GIF-CQ260. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS SIF-Q260. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS ST-SB1. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS OBCU. CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS D-201-10704. BACKGROUND/AIMS: ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) IN PATIENTS WITH SURGICALLY ALTERED ANATOMY (SAA) IS CHALLENGING TO GASTROINTESTINAL ENDOSCOPISTS. THE AIM OF THIS STUDY WAS TO EVALUATE THE IMPACT OF SCOPE EXCHANGE FROM A LONG SINGLE BALLOON ENTEROSCOPE (SBE) TO A GASTROSCOPE DURING SBE-ASSISTED ERCP (SBE-ERCP) IN PATIENTS WITH SAA. METHODS: PATIENTS WHO UNDERWENT SBE-ERCP BETWEEN FEBRUARY 2019 AND OCTOBER 2020 WERE RETROSPECTIVELY IDENTIFIED. INTUBATION SUCCESS, SCOPE EXCHANGE SUCCESS, CANNULATION SUCCESS, AND THERAPEUTIC SUCCESS WERE ANALYZED ALONG WITH COMPLICATIONS. RES
Patient problems: Perforation
MAUDE report 14760302 ↗InjuryApr 5, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS REVIEWED THE FOLLOWING LITERATURE ARTICLE: "WATER EXCHANGE-ASSISTED VS. CARBON DIOXIDE-INSUFFLATED SINGLE-BALLOON ENTEROSCOPY: RANDOMIZED CONTROLLED TRIAL" BY SHUHUI LIANG, ET AL. ABSTRACT: BACKGROUND SINGLE-BALLOON ENTEROSCOPY (SBE) IS A VALUABLE BUT DIFFICULT MODALITY FOR THE DIAGNOSIS AND TREATMENT OF SMALL-BOWEL DISEASE. THE WATER EXCHANGE METHOD HAS THE ADVANTAGE OF FACILITATING INTUBATION DURING COLONOSCOPY. HERE, WE EVALUATED THE EFFECTS OF WATER EXCHANGE ON PROCEDURE-RELATED VARIABLES RELATED TO SBE. METHODS THIS RANDOMIZED CONTROLLED TRIAL WAS CONDUCTED IN A TERTIARY-CARE REFERRAL CENTER IN (B)(6). PATIENTS DUE FOR ATTEMPTED TOTAL ENTEROSCOPY WERE RANDOMLY ALLOCATED TO UNDERGO WATER EXCHANGE-ASSISTED (WATER EXCHANGE GROUP) OR CARBON DIOXIDE-INSUFFLATED ENTEROSCOPY (CO2 GROUP). ALL PATIENTS WERE PLANNED TO UNDERGO BOTH ANTEROGRADE AND RETROGRADE PROCEDURES. THE PRIMARY OUTCOME WAS THE TOTAL ENTEROSCOPY RATE. SECONDARY OUTCOMES INCLUDED THE MAXIMAL INSERTION DEPTH, POSITIVE FINDINGS, PROCEDURAL TIME, AND ADVERSE EVENTS. RESULTS: IN TOTAL, 110 PATIENTS WERE ENROLLED, WITH 55 IN EACH GROUP. BASELINE CHARACTERISTICS BETWEEN THE TWO GROUPS WERE COMPARABLE. TOTAL ENTERO
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 14014189 ↗
FDA clearance and classification
The FDA regulates product code FDA (enteroscope and accessories) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the BALLOON CONTROL UNIT OBCU
Has the BALLOON CONTROL UNIT OBCU been recalled?
No FDA recall record in our data names this exact device identifier (DI 04953170239533), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the BALLOON CONTROL UNIT OBCU?
The FDA's MAUDE database lists 15 adverse event reports for this device, received between Apr 5, 2022 and May 17, 2026. In those reports, the most common specific device problems are “Pressure Problem” (6), “Improper Flow or Infusion” (2) and “Break” (1); the most common patient problems are “Unspecified Tissue Injury” (2), “Erosion” (1) and “Fever” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the BALLOON CONTROL UNIT OBCU?
None of the reports are classified as deaths, 3 as injuries and 12 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the BALLOON CONTROL UNIT OBCU, and is it still sold?
The FDA's device database lists Olympus Medical Systems Corp. as the company responsible for it (brand BALLOON CONTROL UNIT, model OBCU). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Enteroscope and accessories devices
- EVIS EXERA II SMALL INTESTINAL VIDEOSCOPE OLYMPUS SIF TYPE Q180Olympus Medical Systems Corp. · Enteroscope and accessories1 death87 reports
- EVIS EXERA II SMALL INTESTINAL VIDEOSCOPE OLYMPUS SIF TYPE Q180Olympus Medical Systems Corp. · Enteroscope and accessories48 reports
- ENDOSCOPE POSITION DETECTING UNIT UPD-3Olympus Medical Systems Corp. · Enteroscope and accessories4 reports
- BALLOON CONTROL UNIT OBCUOlympus Medical Systems Corp. · Enteroscope and accessories2 reports
- WATER-RESISTANT CAP MAJ-962Olympus Medical Systems Corp. · Enteroscope and accessories1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page