VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Recalls & reported problems
Olympus Medical Systems Corp. · CLV-S190 VISERA ELITE XENON LIGHT SRCE · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Olympus Visera Elite xenon light source powers Olympus endoscopes. Its 1,148 reports are mostly malfunctions, with 5 injuries and no deaths. The reports describe the unit not powering on, a broken or stuck power button, lamp usage readings that look wrong, and no light output. A few burns appear in the patient problem list. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,155
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,155 adverse event reports that list this exact device (DI 04953170301995), received between Jun 1, 2020 and Sep 29, 2026 — classified as 5 injuries, 1,150 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (189 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Electrical /Electronic Property Problem189
- Output Problem169
- Communication or Transmission Problem152
- No Visual Prompts/Feedback114
- No Device Output106
- Failure to Power Up86
- Display or Visual Feedback Problem60
- Optical Problem33
Patient problems
- No Clinical Signs, Symptoms or Conditions1,110
- No Consequences Or Impact To Patient28
- No Patient Involvement14
- Superficial (First Degree) Burn2
- Burn(s)1
- Full thickness (Third Degree) Burn1
Recent adverse event reports
The 25 most recent of 1,155 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Complete Blockage; Incomplete or Inadequate Connection
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935650 ↗MalfunctionSep 24, 2026No Display/Image; Poor Quality Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50876571 ↗MalfunctionSep 17, 2026Communication or Transmission Problem; Power Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50810255 ↗MalfunctionSep 17, 2026Activation Failure
The customer reported changed the Lamp and he said it came on but in that moment he was not sure if it was on and did not had the scope attached and still had the side panel off of the light source during maintenance involving an Olympus Xenon Light Source. There was no patient involvement reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50803608 ↗MalfunctionSep 14, 2026Poor Quality Image; Overheating of Device
The customer reported dark image and lamp burn out during unspecified procedure involving an Olympus xenon light source. There were no reports of patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50752666 ↗MalfunctionSep 2, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50566023 ↗MalfunctionSep 1, 2026No Display/Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50235681 ↗MalfunctionAug 26, 2026Material Integrity Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50457074 ↗MalfunctionAug 26, 2026Break; Key or Button Unresponsive/not Working
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50452033 ↗MalfunctionAug 14, 2026Failure to Power Up; Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50253915 ↗MalfunctionAug 11, 2026Loose or Intermittent Connection
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50208230 ↗MalfunctionAug 7, 2026Communication or Transmission Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50175233 ↗MalfunctionJul 29, 2026Installation-Related Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50071433 ↗MalfunctionJul 22, 2026Break; Output Problem
THE CUSTOMER REPORTED THAT THE POWER BUTTON WAS BROKEN AND STUCK, PREVENTING THE OLYMPUS XENON LIGHT SOURCE FROM BEING TURNED ON DURING INSPECTION PRIOR TO USE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25902584 ↗MalfunctionJul 15, 2026Electrical /Electronic Property Problem
THE CUSTOMER REPORTED THAT LAMP USAGE INDICATOR READS 500 HOURS, AND THE SPARE LAMP MAY HAVE DEPLOYED DURING AN UNSPECIFIED PROCEDURE INVOLVING AN OLYMPUS XENON LIGHT SOURCE. THE CUSTOMER CONTACTED OLYMPUS TECHNICAL ASSISTANCE SUPPORT (TAC) VIA PHONE. USING THE INSTRUCTION MANUAL AVAILABLE IN ENDOWISE AS A REFERENCE, BIOMED SHANE HALL POWERED ON THE XENON LIGHT SOURCE ALONG WITH THE REMAINING OLYMPUS TOWER COMPONENTS. HE VERIFIED THAT THE LAMP USAGE INDICATOR ON THE XENON LIGHT SOURCE DISPLAYED 500 HOURS. NO ERROR MESSAGES APPEARED ON THE CONNECTED MONITOR, THE SPARE LAMP INDICATOR REMAINED OFF, AND THE LIGHT OUTPUT FROM THE ENDOSCOPE CONNECTED TO THE XENON LIGHT SOURCE WAS ADEQUATE. THE PART NUMBER WAS ALSO PROVIDED AS THE RECOMMENDED REPLACEMENT LAMP. HE INDICATED THAT HE PLANNED TO ORDER IT AT A LATER DATE. THE CUSTOMER INFORMED TAC OF THE EVENT. THE DEVICE WILL NOT BE RETURNED TO OLYMPUS FOR EVALUATION. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25827620 ↗MalfunctionJul 8, 2026No Display/Image; Loose or Intermittent Connection
THE CUSTOMER REPORTED NO IMAGE AND A LOOSE BOTTOM PLUG DURING AN INSPECTION INVOLVING AN OLYMPUS XENON LIGHT SOURCE. THERE WAS NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25748846 ↗MalfunctionJun 26, 2026Communication or Transmission Problem; Output Problem
THE CUSTOMER RETURNED THE XENON LIGHT SOURCE TO THE SERVICE CENTER FOR EVALUATION RELATED TO A SEPARATE ISSUE. DURING TESTING, THE REPAIR CENTER TECHNICIAN FOUND THE BAD FAN NOT WORKING AND SCOPE DETECTOR NOT FUNCTIONAL, CAUSE TRACED TO COMPONENT FAILURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25636706 ↗MalfunctionJun 9, 2026Key or Button Unresponsive/not Working
THE CUSTOMER REPORTED THE STANDBY ON LIGHT BUTTON WAS NOT FUNCTIONING ON THE LIGHT SOURCE, DURING THERAPEUTIC URETEROSCOPY PROCEDURE WHICH WAS COMPLETED USING A SIMILAR DEVICE INVOLVING AN OLYMPUS XENON LIGHT SOURCE. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25449058 ↗MalfunctionJun 3, 2026Poor Quality Image
THE CUSTOMER REPORTED DARK IMAGE DISPLAYED DURING AN UNSPECIFIED PROCEDURE INVOLVING AN OLYMPUS XENON LIGHT SOURCE. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25382943 ↗MalfunctionMay 31, 2026Optical Problem
THE CUSTOMER REPORTED THAT THE LAMP WAS DIM AND NEEDED TO BE REPLACED ON AN OLYMPUS XENON LIGHT SOURCE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25345185 ↗MalfunctionMay 23, 2026No Display/Image; Poor Quality Image
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25264921 ↗MalfunctionMay 23, 2026Communication or Transmission Problem
IT WAS REPORTED THAT THE XENON LIGHT SOURCE HAD SCOPE ERROR CODE (E200). THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25264504 ↗MalfunctionMay 20, 2026Erratic or Intermittent Display; Connection Problem; Mechanics Altered
IT WAS REPORTED THAT THE XENON LIGHT SOURCE HAD THE FRONT PANEL BLINKING, THE SWITCH AT THE 12:00 POSITION WAS NOT WORKING WHEN IT WAS EXERCISED, THE PLASTIC TAB INSIDE THE SOCKET WAS STICKY AND WAS NOT MOVING FREELY, THE CONNECTION WAS LOOSE, AND WHEN CONNECTING THE UNIVERSAL LIGHT GUIDE ADAPTER, THE CONNECTION DID NOT STAY SECURED AND FELT LOOSE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25229502 ↗MalfunctionMay 19, 2026Power Problem
IT WAS REPORTED THE XENON LIGHT SOURCE WAS FLASHING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25220029 ↗MalfunctionMay 14, 2026Connection Problem
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE XENON LIGHT SOURCE HAD NOT FUNCTIONAL SCOPE DETECTOR ON SCOPE SOCKET. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25171251 ↗
Other BRONCHOSCOPE (FLEXIBLE OR RIGID) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
- TRACHEAL INTUBATION FIBERSCOPE OLYMPUS LF-DPOlympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1 recall332 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page