EVIS EUS ENDOSCOPIC ULTRASOUND CENTER OLYMPUS EU-ME2Recalls & reported problems

Olympus Medical Systems Corp. · EU-ME2 EU-ME2 ULTRASOUND PROCESSOR · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04953170343490), received between Jul 3, 2024 and Jul 3, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Communication or Transmission Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Communication or Transmission Problem2
  • No Display/Image2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 3, 2024No Display/Image; Communication or Transmission Problem; No Display/Image; Communication or Transmission Problem

    IT WAS REPORTED THAT THE DURING A DIAGNOSTIC EBUS-TBNA (ENDOBRONCHIAL ULTRASOUND GUIDED TRANSBRONCHIAL NEEDLE ASPIRATION) PROCEDURE, PROCESSOR MALFUNCTION OCCURRED, WHERE THE IMAGE FROZE AND ERROR MESSAGE US1605 OCCURRED FOR CELL BOARD MALFUNCTION. AS SUCH, THE PROCEDURE WAS CUT SHORT. THE SAMPLE OBTAINED WAS NOT ENOUGH AND PATIENT MAY NEED TO BE RESCHEDULED. SEVERAL LYMPH NOTED THAT NEEDED TO BE SAMPLED ONLY ONE WAS SAMPLED BEFORE THE MALFUNCTION. THE PROCEDURE WAS NOT ABLE TO BE COMPLETED AS PLANNED. THE PATIENT NEEDS TO BE RESCHEDULED FOR REPEAT PROCEDURE. ADDITIONAL FOLLOW-UP HAS BEEN REQUESTED BUT UNAVAILABLE THE TIME OF THE REPORT. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19667400 ↗
  • InjuryJul 3, 2024No Display/Image; Communication or Transmission Problem; No Display/Image; Communication or Transmission Problem

    IT WAS REPORTED THAT THE DURING A DIAGNOSTIC EBUS-TBNA (ENDOBRONCHIAL ULTRASOUND GUIDED TRANSBRONCHIAL NEEDLE ASPIRATION) PROCEDURE, PROCESSOR MALFUNCTION OCCURRED, WHERE THE IMAGE FROZE AND ERROR MESSAGE US1605 OCCURRED FOR CELL BOARD MALFUNCTION. AS SUCH, THE PROCEDURE WAS CUT SHORT. THE SAMPLE OBTAINED WAS NOT ENOUGH AND PATIENT MAY NEED TO BE RESCHEDULED. SEVERAL LYMPH NOTED THAT NEEDED TO BE SAMPLED ONLY ONE WAS SAMPLED BEFORE THE MALFUNCTION. THE PROCEDURE WAS NOT ABLE TO BE COMPLETED AS PLANNED. THE PATIENT NEEDS TO BE RESCHEDULED FOR REPEAT PROCEDURE. ADDITIONAL FOLLOW-UP HAS BEEN REQUESTED BUT UNAVAILABLE THE TIME OF THE REPORT. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19667390 ↗