ULTRASONIC PROBE UM-3R (UDI 04953170368561)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of ULTRASONIC PROBE (17)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 04953170368561), received between Jan 13, 2021 and Apr 9, 2026 — classified as 18 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash4
- Material Deformation4
- Material Twisted/Bent3
- Detachment of Device or Device Component2
- Fluid/Blood Leak2
- Material Puncture/Hole2
- Material Split, Cut or Torn2
- Scratched Material2
Patient problems
- No Clinical Signs, Symptoms or Conditions17
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 9, 2026Detachment of Device or Device Component; Material Deformation
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE ULTRASONIC PROBE EXHIBITED FOUND LOCK PIN WAS DAMAGED WITH CHIP OFF AND FOUND MANY KINKS ON THE PROBE¿S CABLE UNIT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24835014 ↗MalfunctionJan 2, 2026Break
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE ULTRASONIC PROBE EXHIBITED THE DISTAL END WAS DAMAGED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23943135 ↗MalfunctionDec 17, 2025Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THE ULTRASONIC PROBE HAD A LEAKAGE AT THE DISTAL END OF THE SENSOR. THE ISSUE OCCURRED DURING THE SET-UP. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23830266 ↗MalfunctionSep 9, 2025Leak/Splash; Material Puncture/Hole; Leak/Splash; Material Puncture/Hole
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE ULTRASONIC PROBE EXHIBITED A PINHOLE IN THE SHEATH IN THE DISTAL END. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23009273 ↗MalfunctionJul 11, 2025Material Frayed; Detachment of Device or Device Component; Material Split, Cut or Torn; Material Frayed; Detachment of Device or Device Component; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING INSPECTION FOR USE, THE PLASTIC COVERING ON THE ULTRASONIC PROBE WAS PULLED APART, AND IT WAS NOTED THAT THE WIRES WERE FRAYED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22485449 ↗MalfunctionDec 9, 2024Material Twisted/Bent; Material Twisted/Bent
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED BEND IN THE INSERTION SHEATH. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20878829 ↗MalfunctionNov 5, 2024Fracture; Fracture
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC PROBE EXHIBITED A FRACTURE AT THE DISTAL END OF THE TRANSDUCER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20609942 ↗MalfunctionOct 3, 2024Material Deformation; Material Deformation
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, AN INDENTATION AT THE TIP OF THE INSERTION PART WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT ON THIS REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20364065 ↗MalfunctionSep 16, 2024Material Deformation; Scratched Material; Material Deformation; Scratched Material
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC PROBE EXHIBITED SCRATCHES ON THE INSERTION SITE AND BENDING OF THE INSERTION TIP. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20226541 ↗MalfunctionAug 27, 2024Material Deformation; Material Deformation
IT WAS OBSERVED DURING THE DEVICE EVALUATION, THE ULTRASONIC PROBE HAD WRINKLES AT THE INSERTION SITE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20081721 ↗MalfunctionAug 6, 2024Leak/Splash; Material Puncture/Hole; Leak/Splash; Material Puncture/Hole
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC PROBE EXHIBITED ULTRASONIC MEDIA LEAKAGE DUE TO TIP PART BREAKAGE (HOLE). THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19919462 ↗MalfunctionJul 22, 2024Scratched Material; Scratched Material
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED SCRATCHES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19799906 ↗MalfunctionJun 28, 2024Stretched; Deformation Due to Compressive Stress; Stretched; Deformation Due to Compressive Stress
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED INDENTATION AND ELONGATION AT THE TIP OF THE INSERTION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19639349 ↗MalfunctionJun 21, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED DURING A DIAGNOSTIC PROCEDURE THE SUBJECT DEVICE HAD A LEAKAGE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19581600 ↗MalfunctionMay 10, 2024Material Twisted/Bent; Material Twisted/Bent
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC PROBE EXHIBITED TWISTING OF THE INSERTION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19286814 ↗MalfunctionMay 6, 2024Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT ULTRASOUND PROBE HAD A TEAR AT THE TIP OF THE INSERT AND LEAKAGE OF ULTRASOUND MEDIUM. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19247094 ↗MalfunctionMar 14, 2022Fluid/Blood Leak; Fluid/Blood Leak
THE USER FACILITY REPORTED THAT WHEN PREPARING THE SUBJECT DEVICE FOR USE, THE ULTRASOUND IMAGE WAS ABNORMAL. THERE WAS NO REPORT OF PATIENT/USER HARM OR INJURY DUE TO THE EVENT. THE DEVICE WAS SENT TO AN OLYMPUS SERVICE CENTER FOR EVALUATION. DEVICE INSPECTION AND TESTING CONFIRMED THE REPORTED IMAGE ISSUE. IN ADDITION, SERVICE OBSERVED ULTRASONIC MEDIUM (OIL) LEAKING FROM THE INSERTION TUBE. THIS REPORT IS BEING SUBMITTED FOR THE REPORTABLE MALFUNCTION OF THE ULTRASONIC MEDIUM LEAKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13763038 ↗MalfunctionJan 13, 2021Fluid/Blood Leak; Material Twisted/Bent; Fluid/Blood Leak; Material Twisted/Bent
IT WAS REPORTED THERE WAS OIL LEAKING FROM THE TIP OF THE ULTRASONIC PROBE. NO PATIENT HARM OR IMPACT WAS REPORTED DUE TO THIS INCIDENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11158521 ↗
Other Transducer, ultrasonic, diagnostic devices
- Site~Rite 8 Ultrasound System 9770603Bard Access Systems, Inc. · Transducer, ultrasonic, diagnostic620 reports
- ULTRASONIC PROBE UM-S20-17SOlympus Medical Systems Corp. · Transducer, ultrasonic, diagnostic2 deaths344 reports
- Transducer X7-2t 989605414122Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic6 recalls21 reports
- EPIQ Diagnostic Ultrasound System 989605488741Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic3 recalls165 reports
- bk3000, bk3500, bk5000, bkActiv 2300Bk Medical Aps · Transducer, Ultrasonic, Diagnostic6 recalls1 report
- Transducer L12-3 989605412072Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls1 report
- Transducer L15-7io 989605387852Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls0 reports
- Transducer eL18-4 EMT 989605431531Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page