EVES EUS ULTRASOUND BRONCHOFIBERVIDEOSCOPE OLYMPUS BF-UC190FRecalls & reported problems

Olympus Medical Systems Corp. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus EUS ultrasound bronchoscope is a flexible scope with an ultrasound probe used to view the airways and nearby tissue. The 998 reports often describe the display or image failing, communication faults, and parts detaching or chipping. There are 64 injury reports and no deaths. Reports rose from 227 in the prior 12 months to 340 in the last 12. 2 recalls are open, covering water left in the scope's channel after repair, which could allow contamination, and a scope that can get stuck in a breathing tube connector.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,082
Deaths reported
0
Injuries reported
68
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1,082 adverse event reports that list this exact device (DI 04953170399831), received between Jan 7, 2021 and Sep 30, 2026 — classified as 68 injuries, 1,014 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (271 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 16 reportsDec 2023: 21 reportsJan 2024: 12 reportsFeb 2024: 17 reportsMar 2024: 20 reportsApr 2024: 21 reportsMay 2024: 15 reportsJun 2024: 15 reportsJul 2024: 19 reportsAug 2024: 7 reportsSep 2024: 23 reportsOct 2024: 13 reportsNov 2024: 13 reportsDec 2024: 8 reportsJan 2025: 12 reportsFeb 2025: 5 reportsMar 2025: 27 reportsApr 2025: 7 reportsMay 2025: 26 reportsJun 2025: 9 reportsJul 2025: 17 reportsAug 2025: 31 reportsSep 2025: 24 reportsOct 2025: 48 reportsNov 2025: 40 reportsDec 2025: 38 reportsJan 2026: 46 reportsFeb 2026: 7 reportsMar 2026: 17 reportsApr 2026: 49 reportsMay 2026: 31 reportsJun 2026: 0 reportsJul 2026: 31 reportsAug 2026: 78 reportsSep 2026: 87 reportsOct 2026: 0 reports20242025202687
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 25, 2023Class IIOpen, Classified

    Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

    Z-0328-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Jul 27, 2023Class IIOpen, Classified

    Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.

    Z-2494-2023 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • No Display/Image271
  • Communication or Transmission Problem187
  • Detachment of Device or Device Component170
  • Failure to Clean Adequately86
  • Poor Quality Image74
  • Material Fragmentation73
  • Break49
  • Device Reprocessing Problem41

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,053
  • Foreign Body In Patient15
  • Insufficient Information4
  • Subclinical Infection4
  • No Consequences Or Impact To Patient2
  • Pulmonary Emphysema2
  • Upper Respiratory Tract Injury2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 25 most recent of 1,082 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026No Display/Image; Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952642 ↗
  • MalfunctionSep 30, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951814 ↗
  • MalfunctionSep 30, 2026Detachment of Device or Device Component

    It was observed that during the device inspection, the Bronchofibervideoscope exhibited the C-BODY ¿ was chipped at probe unit tip. There were no reports of patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951296 ↗
  • MalfunctionSep 30, 2026Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950534 ↗
  • MalfunctionSep 30, 2026Erratic or Intermittent Display

    The customer returned the Bronchofibervideoscope for a separate issue. During the initial evaluation of the returned device, an intermittent/flickering image was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950087 ↗
  • MalfunctionSep 30, 2026No Display/Image; Poor Quality Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949987 ↗
  • MalfunctionSep 30, 2026Material Fragmentation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949138 ↗
  • MalfunctionSep 30, 2026No Display/Image; Communication or Transmission Problem

    The customer reported an Error code E226 and when the customer shut it down the scope still did not work. The issue occurred during preparation for use and before the patient was in the procedure room. The issue involved an Olympus Ultrasound Bronchofibervideoscope. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946090 ↗
  • MalfunctionSep 30, 2026Detachment of Device or Device Component

    The customer reported the Distal End was partially detached during preparation for use and before the patient was in the procedure room. The issue involved an Olympus Ultrasound Bronchofibervideoscope. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945564 ↗
  • MalfunctionSep 30, 2026Detachment of Device or Device Component

    The customer reported the BRONCHOFIBERVIDEOSCOPE'S sheath came off the distal tip. The issue was found during reprocessing..

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941794 ↗
  • MalfunctionSep 29, 2026Erratic or Intermittent Display

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941333 ↗
  • MalfunctionSep 29, 2026Scratched Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938311 ↗
  • MalfunctionSep 29, 2026Contamination

    The customer returned the Bronchofibervideoscope for another issue. The device was returned unrepaired to customer as biohazard. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935953 ↗
  • MalfunctionSep 29, 2026No Display/Image; Communication or Transmission Problem

    The customer reported the scope was giving an unspecified communication error message during preparation for use and before the patient was in the procedure room. The issue involved an Olympus Ultrasound Bronchofibervideoscope. Additionally, during the device analysis, the Olympus Service Repair Technician indicated that a B30 scope communication error message and no image was found. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50931831 ↗
  • MalfunctionSep 29, 2026No Display/Image

    It was reported that the camera image was not showing. Unsure as to when the unspecified procedure occurred involving an Olympus Bronchofibervideoscope. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50930963 ↗
  • MalfunctionSep 28, 2026Device Reprocessing Problem

    The customer reported that device failed in the Medivator, involving an Olympus bronchofibervideoscope. There was no patient injury involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923509 ↗
  • MalfunctionSep 28, 2026Material Fragmentation

    The customer returned the Bronchofibervideoscope for a separate issue. During the initial evaluation of the returned device, a chip on the distal end was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920965 ↗
  • MalfunctionSep 28, 2026Material Fragmentation

    The customer returned the Bronchofibervideoscope for a separate issue. During the initial evaluation of the returned device, a chip on distal end was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50919288 ↗
  • MalfunctionSep 28, 2026Detachment of Device or Device Component

    The customer reported the channel that connects to the handle was coming loose and coming apart. The issue occurred during reprocessing. The issue involved an Olympus Ultrasound Bronchofibervideoscope. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50913924 ↗
  • MalfunctionSep 25, 2026Material Separation; Unintended Movement

    The customer reported that there was a gap and movement on the introduction portion of the scope that occurred during a diagnostic Endobronchial Ultrasound (EBUS) involving an Olympus Broncho fiber videoscope. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50899719 ↗
  • MalfunctionSep 25, 2026Material Split, Cut or Torn

    The customer returned the Ultrasound Bronchofibervideoscope to the service center for a separate issue. During the device analysis, the Olympus Service Repair Technician indicated that there was a cut on the probe unit pink rubber. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50897808 ↗
  • MalfunctionSep 25, 2026Detachment of Device or Device Component

    It was observed that during the device inspection, the Bronchofibervideoscope exhibited C-body was chipped. There were no reports of patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50896051 ↗
  • MalfunctionSep 24, 2026Material Fragmentation

    The customer returned the Bronchofibervideoscope for a separate issue. During the initial evaluation of the returned device, a chipped probe tip was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50886594 ↗
  • MalfunctionSep 24, 2026Material Fragmentation

    The customer returned the Bronchofibervideoscope for a separate issue. During the initial evaluation of the returned device, it was observed that the distal end was chipped. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50886051 ↗
  • MalfunctionSep 24, 2026Material Fragmentation

    The customer returned the Ultrasound Bronchofibervideoscope to the service center for a separate issue. During the device analysis, the Olympus Service Repair Technician indicated that a chip at the Pink Probe was found. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50882264 ↗

Other Ultrasound bronchoscope devices

All Ultrasound bronchoscope devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page