Long Clip HX-610-090L (REF N6012730)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2 · Implant
All models of Long Clip (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04953170415159), received between Sep 29, 2022 and Sep 18, 2025 — classified as 4 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Detachment of Device or Device Component2
- Failure to Clean Adequately1
- Mechanical Jam1
- Separation Failure1
- Separation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Hemorrhage/Blood Loss/Bleeding2
- Hematemesis1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjurySep 18, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN UNSPECIFIED PROCEDURE, THE CLIP WAS REMOVED FROM THE SHEATH FOR USE ON A LESION INSIDE THE PATIENT¿S BODY, BUT IT DETACHED IMMEDIATELY AND BECAME UNUSABLE. THE DETACHED CLIP WAS RETRIEVED FROM THE PATIENT¿S BODY AND DISPOSED OF BY THE HOSPITAL. THERE WERE NO FURTHER REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23089685 ↗MalfunctionOct 7, 2024Separation Failure; Separation Failure
IT WAS REPORTED THE LONG CLIP THAT GRIPS THE MUCOUS MEMBRANE HAD NOT BEEN ABLE TO BE DETACHED AND IT WAS REMOVED BY HAND BY THE DOCTOR. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20388156 ↗MalfunctionOct 2, 2024Separation Problem; Separation Problem
IT WAS REPORTED THAT THE LONG CLIP COULD NOT BE FIRED OR WITH DIFFICULTIES. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE, WITH NO DELAY. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20354401 ↗InjurySep 6, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED, THAT THE CLIP FELL OFF INSIDE THE PATIENT'S BODY AND WAS RETRIEVED. IT IS UNKNOWN, WHERE IT FELL OFF, BUT IT WAS NOT INSIDE THE ABDOMINAL CAVITY. THE PROCEDURE WAS COMPLETED, AFTER REPLACING IT WITH A SIMILAR DEVICE. THERE WAS NO HEALTH DAMAGE TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20158510 ↗MalfunctionAug 25, 2024Failure to Clean Adequately; Failure to Clean Adequately
IT WAS REPORTED A HEMOSTASIS CLIP WAS SUCTIONED WITHIN THE PATIENT¿S STOMACH AND CAUSED SUCTION ISSUES DURING A DIAGNOSTIC ENDOSCOPIC PROCEDURE AND THE CUSTOMER SUSPECTS THE CLIP MAY HAVE REMAINED WITHIN THE GASTROINTESTINAL VIDEOSCOPE AFTER A PREVIOUS PROCEDURE AND IT WAS NOT REPROCESSED CORRECTLY. THE CLIP WAS SUCTIONED AND REMOVED DURING THE PROCEDURE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20061166 ↗InjuryAug 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) RECEIVED A LITERATURE TITLED "TECHNICAL OUTCOMES AND POSTPROCEDURAL COURSES OF MUCOSAL INCISION-ASSISTED BIOPSY FOR POSSIBLE GASTRIC GASTROINTESTINAL STROMAL TUMORS: A SERIES OF 48 CASES (WITH VIDEO)." MUCOSAL INCISION-ASSISTED BIOPSY (MIAB) HAS BEEN INTRODUCED AS AN ALTERNATIVE TO ENDOSCOPIC ULTRASOUND-GUIDED FINE NEEDLE ASPIRATION FOR TISSUE SAMPLING OF SUBEPITHELIAL LESIONS. HOWEVER, THERE HAVE BEEN FEW REPORTS ON MIAB, AND THE EVIDENCE IS LACKING, PARTICULARLY IN SMALL LESIONS. IN THIS CASE SERIES, WE INVESTIGATED THE TECHNICAL OUTCOMES AND POSTPROCEDURAL INFLUENCES OF MIAB FOR GASTRIC SUBEPITHELIAL LESIONS 10 MM OR GREATER IN SIZE. TYPE OF ADVERSE EVENTS/NUMBER OF PATIENTS: POSTOPERATIVE BLEEDING (N=1). NO INTRAOPERATIVE ADVERSE EVENTS OCCURRED, WHEREAS ONE PATIENT WITH SUBEPITHELIAL LESION GREATER THAN 20MM EXPERIENCED HAEMATEMESIS THE DAY AFTER MIAB. IN THIS CASE, NO BLEEDING WAS FOUND ON EMERGENCY ENDOSCOPY, AND THE PATIENT HAD NO SYMPTOMS THEREAFTER WITHOUT TRANSFUSION OR HEMOSTATIC TREATMENT. THIS EVENT REQUIRES 4 REPORTS. THE PATIENT IDENTIFIERS ARE AS FOLLOWS: (B)(6) : GIF-Q260J, (B)(6) : D-201-11804, (B)(6) : FD-410LR, (B)(6) : HX-6
Patient problems: Hematemesis; Hemorrhage/Blood Loss/Bleeding
MAUDE report 17450076 ↗MalfunctionMar 2, 2023Mechanical Jam; Mechanical Jam
ADDENDUM MAR 7, 2023: THE EVENT OCCURRED DURING HEMOSTASIS. THERE WAS NO TISSUE INJURY DUE TO THE FORCIBLE REMOVAL OF THE CLIP DEVICE. THERE IS NO INFORMATION IF THE FORCIBLE REMOVAL OF THE CLIP DEVICE LED TO ANY BLEEDING. THERE WAS NO DAMAGE TO THE CLIP FIXING DEVICE DUE TO THE FORCIBLE REMOVAL OF THE CLIP DEVICE. THERE WAS ALMOST NO DELAY IN THE COMPLETION OF THE PROCEDURE. IT IS NOT KNOWN IF THE PATIENT NEEDED MORE ANESTHESIA. THE DEVICE WAS INSPECTED BEFORE USE WITH NO ANOMALIES NOTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16466721 ↗InjurySep 29, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS REVIEWED THE FOLLOWING LITERATURE TITLED "GEL IMMERSION ENDOSCOPIC MUCOSAL RESECTION (EMR) FOR SUPERFICIAL NONAMPULLARY DUODENAL EPITHELIAL TUMORS MAY REDUCE PROCEDURE TIME COMPARED WITH UNDERWATER EMR (WITH VIDEO)". LITERATURE SUMMARY: THIS RETROSPECTIVE COHORT STUDY WAS AIMED TO EVALUATE THE ENDOSCOPIC OUTCOMES OF UNDERWATER ENDOSCOPIC MUCOSAL RESECTION (UEMR) AND GEL IMMERSION EMR (GIEMR). TOTAL 23 PATIENTS WERE INCLUDED IN THIS STUDY. GIEMR SIGNIFICANTLY REDUCED THE PROCEDURE TIME COMPARED WITH UEMR. THERE WAS NO SIGNIFICANT DIFFERENCE BETWEEN THE UEMR AND GIEMR GROUPS FOR EN BLOC RESECTION RATE AND R0 RESECTION RATE. THERE WAS ONE CASE OF INTRAOPERATIVE BLEEDING IN THE GIEMR GROUP AND ONE IN THE UEMR GROUP; BOTH OF WHICH WERE CONTROLLED WITH ENDOSCOPIC CLIPS. NO SERIOUS COMPLICATIONS WERE OBSERVED IN EITHER GROUP. THIS STUDY CONCLUDED THAT GIEMR WAS SAFE AND HAD GOOD THERAPEUTIC RESULTS FOR SNADETS AND SHOWED THE POTENTIAL TO ACCELERATE THE PROCEDURE TIME COMPARED WITH UEMR. THIS METHOD MAY BE PARTICULARLY USEFUL IN AREAS WHERE IMMERSION IN WATER IS DIFFICULT. INTRAOPERATIVE BLEEDING WAS DEFINED AS BLEEDING DURING OR IMMEDIATELY AFTER ENDOSCOPIC RESECTION. TYPE OF ADVE
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 15501597 ↗
Other Hemostatic metal clip for the GI tract devices
- Resolution 360 Clip M00521230Boston Scientific Corporation · Hemostatic metal clip for the GI tract3 deaths2,291 reports
- Resolution Clip M00522610Boston Scientific Corporation · Hemostatic metal clip for the GI tract1,116 reports
- Instinct G58010Cook Incorporated · Hemostatic metal clip for the GI tract1 recall1,017 reports
- Resolution 360™ ULTRA Clip M00521400Boston Scientific Corporation · Hemostatic metal clip for the GI tract3 deaths496 reports
- Instinct G18343Cook Incorporated · Hemostatic metal clip for the GI tract244 reports
- MANTIS™ Clip M00521420Boston Scientific Corporation · Hemostatic metal clip for the GI tract195 reports
- Single Use Repositionable Clip HX-202UROlympus Medical Systems Corp. · Hemostatic metal clip for the GI tract1 recall118 reports
- colonic FTRD Set 14Ovesco Endoscopy AG · Hemostatic Metal Clip For The Gi Tract2 deaths95 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page