Single Use Injector NM-400U-0425 (REF N6139730)Recalls & reported problems
Olympus Medical Systems Corp. · Single Use Injector · Class 2
All models of Single Use Injector (29)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 04953170422577), received between Nov 8, 2022 and Oct 31, 2025 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Failure to Deliver” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Deliver4
- Failure to Infuse4
- Infusion or Flow Problem4
- Activation, Positioning or Separation Problem2
- Failure to Align2
- Activation Problem1
- Mechanics Altered1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 31, 2025Activation, Positioning or Separation Problem
OLYMPUS INJECTOR FORCE 25G NEEDLE WOULD NOT DEPLOY OUT OF SHEATH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23435743 ↗MalfunctionNov 25, 2024
INSERTED INJECTOR NEEDLE DOWN SCOPE, TRIED TO DEPLOY NEEDLE FROM SHEATH TO INJECT, BUT NEEDLE WOULD NOT COME OUT. REMOVED INJECTOR NEEDLE FROM SCOPE, ATTEMPTED TO DEPLOY NEEDLE OUTSIDE OF SCOPE, AND NEEDLE WOULD NOT DEPLOY EITHER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20771114 ↗MalfunctionJan 12, 2024
INJECTION NEEDLE FAILED TO DEPLOY. DEVICE APPEARS TO BE MISSING A COMPONENT THAT CONNECTS THE NEEDLE TO THE HANDLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18502131 ↗MalfunctionAug 17, 2023
NJECTORFORCE MAX/OLYMPUS SINGLE USE INJECTOR FAILED TO ADVANCE NEEDLE OUT OF THE DEVICE WHEN ASSISTANT ATTEMPTED TO ACTIVATE THE MECHANISM TO "NEEDLE OUT". NEW INJECTOR OPENED, TESTED THE FUNCTION ON THIS NEW DEVICE PRIOR TO INSERTING INTO GASTROSCOPE- DEVICE WORKING CORRECTLY/NEEDLE WAS ADVANCING PROPERLY WHEN MECHANISM INITIATED. DEVICE INSERTED INTO SCOPE, DEVICE CATHETER IN VIEW ON SCREEN, DEVICE FAILED TO WORK AGAIN. RESISTANCE MET WHEN TRYING TO ADVANCE THE NEEDLE INTO VIEW. THIRD INJECTOR OPENED, TESTED MECHANISM- CONFIRMED IT WAS WORKING CORRECTLY, INSERTED INTO SCOPE AND THIS TIME NEEDLE DEVICE ADVANCING PROPERLY AND INJECTION WAS ABLE TO BE PERFORMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17563468 ↗MalfunctionJun 27, 2023Activation, Positioning or Separation Problem
NEEDLE FLUSHED WITH TATTOO. NEEDLE WAS RETRACTED. 25GA X 4MM INJECTION NEEDLE WAS INTRODUCED IN THE SCOPE. UPON ATTEMPTING TO USE THE NEEDLE IT WOULD NOT DEPLOY. MULTIPLE ATTEMPTS TO REPOSITION AND DEPLOY HAVE BEEN ATTEMPTED. THE DEVICE WAS REMOVED FROM THE SCOPE AND WOULD NOT DEPLOY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17208563 ↗MalfunctionJan 20, 2023Failure to Deliver; Failure to Infuse; Infusion or Flow Problem; Failure to Deliver; Failure to Infuse; Infusion or Flow Problem
USER FACILITY REPORTED WITH AN ISSUE OF "INK SPREAD EVERYWHERE EXCEPT DESIRED LOCATION". CUSTOMER STATED "THE INK IS SPREADING EVERYWHERE EXCEPT WHERE THE INK NEEDS TO GO. ACCORDING TO THE REPORTER, THE ISSUE OCCURRED DURING A STANDARD PROCEDURE FOR A DIAGNOSTIC COLONOSCOPY PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. NO HARM WAS REPORTED. NO PATIENT HARM, NO USER INJURY REPORTED . PER THE REPORT , THE PROBLEM OCCURRED/FAILED TWICE ON THE SAME MODEL WITH THE SAME LOT (21V). NO FURTHER INFORMATION PROVIDED. THIS EVENT INCLUDES TWO (2 ) REPORTS TO CAPTURE THE TWO REPORTED FAILURE: REPORT WITH PATIENT IDENTIFIER (B)(6) (NM-400U-0425, LOT 21V). REPORT WITH PATIENT IDENTIFIER (B)(6) (NM-400U-0425, LOT 21V). THIS REPORT BEING SUBMITTED IS FOR REPORT WITH PATIENT IDENTIFIER (B)(6) (NM-400U-0425, LOT 21V).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16218469 ↗MalfunctionJan 20, 2023Failure to Deliver; Failure to Infuse; Infusion or Flow Problem; Failure to Deliver; Failure to Infuse; Infusion or Flow Problem
USER FACILITY REPORTED WITH AN ISSUE OF "INK SPREAD EVERYWHERE EXCEPT DESIRED LOCATION". CUSTOMER STATED "THE INK IS SPREADING EVERYWHERE EXCEPT WHERE THE INK NEEDS TO GO. ACCORDING TO THE REPORTER, THE ISSUE OCCURRED DURING A STANDARD PROCEDURE FOR A DIAGNOSTIC COLONOSCOPY PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. NO HARM WAS REPORTED. NO PATIENT HARM, NO USER INJURY REPORTED . PER THE REPORT , THE PROBLEM OCCURRED/FAILED TWICE ON THE SAME MODEL WITH THE SAME LOT (21V). NO FURTHER INFORMATION PROVIDED. THIS EVENT INCLUDES TWO (2 ) REPORTS TO CAPTURE THE TWO REPORTED FAILURE: REPORT WITH PATIENT IDENTIFIER (B)(6) (NM-400U-0425, LOT 21V). REPORT WITH PATIENT IDENTIFIER (B)(6) (NM-400U-0425, LOT 21V). THIS REPORT BEING SUBMITTED IS FOR REPORT WITH PATIENT IDENTIFIER (B)(6) (NM-400U-0425, LOT 21V).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16218449 ↗MalfunctionJan 18, 2023Failure to Deliver; Failure to Infuse; Infusion or Flow Problem; Failure to Deliver; Failure to Infuse; Infusion or Flow Problem
USER FACILITY REPORTED WITH AN ISSUE OF "INK SPREADING". CUSTOMER STATED "THE INK IS SPREADING EVERYWHERE EXCEPT WHERE THE INK NEEDS TO GO. ACCORDING TO THE REPORTER, THE ISSUE OCCURRED DURING A STANDARD PROCEDURE FOR A DIAGNOSTIC COLONOSCOPY PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. NO HARM WAS REPORTED. NO PATIENT HARM, NO USER INJURY REPORTED . THIS EVENT INCLUDES 3 REPORTS: REPORT WITH PATIENT IDENTIFIER (B)(6) (NM-400U-0425 LOT 1YV). REPORT WITH PATIENT IDENTIFIER (B)(6) (NM-400U-0425 LOT 1ZV). REPORT WITH PATIENT IDENTIFIER (B)(6) (NM-400U-0425 LOT 07V 20). THIS REPORT IS FOR REPORT WITH PATIENT IDENTIFIER (B)(6) (NM-400U-0425 LOT 1ZV).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16192141 ↗MalfunctionJan 18, 2023Failure to Deliver; Failure to Infuse; Infusion or Flow Problem; Failure to Deliver; Failure to Infuse; Infusion or Flow Problem
USER FACILITY REPORTED WITH AN ISSUE OF "INK SPREADS". CUSTOMER STATED "THE INK IS SPREADING EVERYWHERE EXCEPT WHERE THE INK NEEDS TO GO. ACCORDING TO THE REPORTER, THE ISSUE OCCURRED DURING A STANDARD PROCEDURE FOR A DIAGNOSTIC COLONOSCOPY PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. NO HARM WAS REPORTED. NO PATIENT HARM, NO USER INJURY REPORTED . THIS EVENT INCLUDES 3 REPORTS: REPORT WITH PATIENT IDENTIFIER (B)(6) (NM-400U-0425 LOT 1YV). REPORT WITH PATIENT IDENTIFIER (B)(6) (NM-400U-0425 LOT 1ZV). REPORT WITH PATIENT IDENTIFIER (B)(6) (NM-400U-0425 LOT 07V 20). THIS REPORT IS FOR REPORT WITH PATIENT IDENTIFIER (B)(6) (NM-400U-0425 LOT 1YV).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16191993 ↗MalfunctionJan 17, 2023Failure to Align; Failure to Align
THE CUSTOMER REPORTED TO OLYMPUS THAT THE SINGLE USE INJECTOR WAS NOT INJECTING LIQUID PROPERLY INTO THE TARGET TISSUE. THE ISSUE OCCURRED DURING THE DIAGNOSTIC PROCEDURE (COLONOSCOPY), AND THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16184021 ↗MalfunctionJan 17, 2023Failure to Align; Failure to Align
THE CUSTOMER REPORTED TO OLYMPUS THAT THE SINGLE USE INJECTOR WAS NOT INJECTING LIQUID PROPERLY INTO THE TARGET TISSUE. THE ISSUE OCCURRED DURING THE DIAGNOSTIC PROCEDURE (COLONOSCOPY), AND THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16184003 ↗MalfunctionNov 8, 2022Mechanics Altered; Activation Problem
NEEDLE WOULD NOT COME OUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15754460 ↗
Other endoscopic injection needle, gastroenterology-urology devices
- AcuJect G22526Cook Incorporated · endoscopic injection needle, gastroenterology-urology1 recall1 report
- n.a. 27182AKarl Storz GmbH & Co. KG · endoscopic injection needle, gastroenterology-urology1 recall0 reports
- n.a. 27030NKarl Storz GmbH & Co. KG · endoscopic injection needle, gastroenterology-urology1 recall0 reports
- injeTAK Adjustable Tip Needle 70cm DIS201Laborie Medical Technologies Canada ULC · endoscopic injection needle, gastroenterology-urology1 recall0 reports
- injeTAK Adjustable Tip Needle 35cm DIS199Laborie Medical Technologies Canada ULC · endoscopic injection needle, gastroenterology-urology1 recall0 reports
- Single Use Injector NM-401L-0425Olympus Medical Systems Corp. · endoscopic injection needle, gastroenterology-urology28 reports
- Single Use Injector NM-401L-0423Olympus Medical Systems Corp. · endoscopic injection needle, gastroenterology-urology27 reports
- Single Use Injector NM-401L-0425Olympus Medical Systems Corp. · endoscopic injection needle, gastroenterology-urology26 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page