Single Use Injector NM-401L-0523 (REF N6140430)Recalls & reported problems
Olympus Medical Systems Corp. · Single Use Injector · Class 2
All models of Single Use Injector (29)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04953170422645), received between Sep 28, 2021 and Dec 2, 2025 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Obstruction of Flow” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Obstruction of Flow5
- Failure to Fire3
- Difficult or Delayed Activation1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 2, 2025Difficult or Delayed Activation; Difficult or Delayed Activation
IT WAS REPORTED THAT THE SINGLE USE INJECTOR MALFUNCTIONED DURING THE INJECTION OF BOTOX INTO THE VOCAL FOLDS. THE SYRINGE PLUNGER WAS REPORTED TO HAVE BEEN FORCED BACK. THE DOCTOR ASKED THE STAFF TO REATTACH THE SYRINGE TO THE INJECTOR TO ATTEMPT THE INJECTION A SECOND TIME. AGAIN, THE SYRINGE WAS FULLY PRESSED, APPEARING TO DELIVER THE MEDICATION, BUT THE SYRINGE PLUNGER BOUNCED BACK WHEN THE USER'S THUMB WAS REMOVED FROM PLUNGER. THERE WERE NO REPORTS PATIENT HARM.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 23698041 ↗MalfunctionFeb 17, 2022Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED TO OLYMPUS THAT DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE THE FLUID WOULD NOT FLOW THROUGH THE SINGLE-USE INJECTOR NEEDLE WHEN ADVANCED OUTSIDE OF THE SHEATH. WHEN RETRACTED INTO THE SHEATH, FLUID FLOWED THROUGH THE NEEDLE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WAS NO REPORT OF PATIENT DEATH, HARM, OR INJURIES. THE FOLLOWING PATIENT IDENTIFIERS ARE RELATED EVENTS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13552142 ↗MalfunctionFeb 17, 2022Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED TO OLYMPUS THAT DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE THE FLUID WOULD NOT FLOW THROUGH THE SINGLE-USE INJECTOR NEEDLE WHEN ADVANCED OUTSIDE OF THE SHEATH. WHEN RETRACTED INTO THE SHEATH, FLUID FLOWED THROUGH THE NEEDLE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WAS NO REPORT OF PATIENT DEATH, HARM, OR INJURIES. THE FOLLOWING PATIENT IDENTIFIERS ARE RELATED EVENTS: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13552104 ↗MalfunctionFeb 17, 2022Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED TO OLYMPUS THAT DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE THE FLUID WOULD NOT FLOW THROUGH THE SINGLE-USE INJECTOR NEEDLE WHEN ADVANCED OUTSIDE OF THE SHEATH. WHEN RETRACTED INTO THE SHEATH, FLUID FLOWED THROUGH THE NEEDLE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WAS NO REPORT OF PATIENT DEATH, HARM, OR INJURIES. THE FOLLOWING PATIENT IDENTIFIERS ARE RELATED EVENTS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13552008 ↗MalfunctionFeb 17, 2022Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED TO OLYMPUS THAT DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE THE FLUID WOULD NOT FLOW THROUGH THE SINGLE-USE INJECTOR NEEDLE WHEN ADVANCED OUTSIDE OF THE SHEATH. WHEN RETRACTED INTO THE SHEATH, FLUID FLOWED THROUGH THE NEEDLE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WAS NO REPORT OF PATIENT DEATH, HARM, OR INJURIES. THE FOLLOWING PATIENT IDENTIFIERS ARE RELATED EVENTS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13551912 ↗MalfunctionFeb 17, 2022Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED TO OLYMPUS THAT DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE THE FLUID WOULD NOT FLOW THROUGH THE SINGLE-USE INJECTOR NEEDLE WHEN ADVANCED OUTSIDE OF THE SHEATH. WHEN RETRACTED INTO THE SHEATH, FLUID FLOWED THROUGH THE NEEDLE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WAS NO REPORT OF PATIENT DEATH, HARM, OR INJURIES. THE FOLLOWING PATIENT IDENTIFIERS ARE RELATED EVENTS: (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13551189 ↗MalfunctionSep 28, 2021Failure to Fire; Failure to Fire
AS REPORTED FOR THIS EVENT BY THE CUSTOMER, THE OLEATE OR SERUM WOULD NOT COME OUT THROUGH THE NEEDLE LUMEN WITH THE NEEDLE OUT. THE LIQUID WAS COMING OUT WITH THE NEEDLE SHEATHED. THIS DEVICE IS USED TO INJECT ETHANOLAMINE OLEATE BY INTRA-MUCOSAL BRONCHIAL PUNCTURE, WITH THE INTENTION OF CLOSING SMALL BRONCHOPLEURAL FISTULAS. IT WAS NOT POSSIBLE TO USE THE DEVICE. TWO OTHER DEVICES FROM DIFFERENT BOXES WERE TRIED AND HAD THE SAME ISSUE. THERE IS NO REPORTED HARM TO ANY PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12540522 ↗MalfunctionSep 28, 2021Failure to Fire; Failure to Fire
AS REPORTED FOR THIS EVENT BY THE CUSTOMER, THE OLEATE OR SERUM WOULD NOT COME OUT THROUGH THE NEEDLE LUMEN WITH THE NEEDLE OUT. THE LIQUID WAS COMING OUT WITH THE NEEDLE SHEATHED. THIS DEVICE IS USED TO INJECT ETHANOLAMINE OLEATE BY INTRA-MUCOSAL BRONCHIAL PUNCTURE, WITH THE INTENTION OF CLOSING SMALL BRONCHOPLEURAL FISTULAS. IT WAS NOT POSSIBLE TO USE THE DEVICE. TWO OTHER DEVICES FROM DIFFERENT BOXES WERE TRIED AND HAD THE SAME ISSUE. THERE IS NO REPORTED HARM TO ANY PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12540511 ↗MalfunctionSep 28, 2021Failure to Fire; Failure to Fire
AS REPORTED FOR THIS EVENT BY THE CUSTOMER, THE OLEATE OR SERUM WOULD NOT COME OUT THROUGH THE NEEDLE LUMEN WITH THE NEEDLE OUT. THE LIQUID WAS COMING OUT WITH THE NEEDLE SHEATHED. THIS DEVICE IS USED TO INJECT ETHANOLAMINE OLEATE BY INTRA-MUCOSAL BRONCHIAL PUNCTURE, WITH THE INTENTION OF CLOSING SMALL BRONCHOPLEURAL FISTULAS. IT WAS NOT POSSIBLE TO USE THE DEVICE. TWO OTHER DEVICES FROM DIFFERENT BOXES WERE TRIED AND HAD THE SAME ISSUE. THERE IS NO REPORTED HARM TO ANY PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12540493 ↗
Other endoscopic injection needle, gastroenterology-urology devices
- AcuJect G22526Cook Incorporated · endoscopic injection needle, gastroenterology-urology1 recall1 report
- n.a. 27182AKarl Storz GmbH & Co. KG · endoscopic injection needle, gastroenterology-urology1 recall0 reports
- n.a. 27030NKarl Storz GmbH & Co. KG · endoscopic injection needle, gastroenterology-urology1 recall0 reports
- injeTAK Adjustable Tip Needle 70cm DIS201Laborie Medical Technologies Canada ULC · endoscopic injection needle, gastroenterology-urology1 recall0 reports
- injeTAK Adjustable Tip Needle 35cm DIS199Laborie Medical Technologies Canada ULC · endoscopic injection needle, gastroenterology-urology1 recall0 reports
- Single Use Injector NM-401L-0425Olympus Medical Systems Corp. · endoscopic injection needle, gastroenterology-urology28 reports
- Single Use Injector NM-401L-0423Olympus Medical Systems Corp. · endoscopic injection needle, gastroenterology-urology27 reports
- Single Use Injector NM-401L-0425Olympus Medical Systems Corp. · endoscopic injection needle, gastroenterology-urology26 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page