Single Use Injector NM-401L-0525 (REF N6140530)Recalls & reported problems
Olympus Medical Systems Corp. · Single Use Injector · Class 2
All models of Single Use Injector (29)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 04953170422652), received between Mar 2, 2022 and Jul 19, 2024 — classified as 11 malfunctions.
The most frequently reported specific device problem is “No Flow” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- No Flow4
- Physical Resistance/Sticking4
- Activation Failure1
- Failure to Charge1
- Failure to Discharge1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 19, 2024Failure to Charge; Failure to Charge
THE PROCEDURE WAS CYSTOSCOPY WITH BOTOX INJECTION THAT WAS COMPLETED USING A SIMILAR DEVICE. THERE WAS A LITTLE DELAY AND NO ANESTHESIA INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19785206 ↗MalfunctionJul 15, 2024Failure to Discharge; Failure to Discharge
IT WAS REPORTED THAT THE INJECTOR AND SHEATH SET WERE UNABLE TO INJECT LIQUID. THE ISSUE OCCURRED DURING A DIAGNOSTIC PROCEDURE; HOWEVER, THE PROCEDURE WAS COMPLETED. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19739883 ↗MalfunctionJun 11, 2024Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THE SINGLE USE INJECTOR HAD RESISTANCE DURING AN INJECTION OF BOTOX OR DEXAMETHASONE AND WHILE DISENGAGING AND FLUSHING THE DEVICE, THE FLUID WENT OUT WITHOUT RESISTANCE. THE ISSUE OCCURRED DURING A THERAPEUTIC PROCEDURE. THE PROCEDURE WAS STOPPED 3 TIMES IN ORDER TO CHANGE OUT THE NEEDLE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19516034 ↗MalfunctionMay 27, 2024Physical Resistance/Sticking; Physical Resistance/Sticking
THE CUSTOMER REPORTED TO OLYMPUS THEY EXPERIENCED RESISTANCE WHEN ENGAGING THE OLYMPUS 25 G NEEDLE TO INJECT BOTOX AND DEXAMETHASONE IN VOCAL CHORDS DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE. THE PROCEDURE HAD TO BE STOPPED THREE TIMES TO CHANGE THE NEEDLE WITH THE SAME PATIENT. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19402410 ↗MalfunctionMay 24, 2024Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THE SINGLE USE INJECTOR HAD RESISTANCE DURING AN INJECTION OF BOTOX OR DEXAMETHASONE AND WHILE DISENGAGING AND FLUSHING THE DEVICE, THE FLUID WENT OUT WITHOUT RESISTANCE. THE ISSUE OCCURRED DURING A THERAPEUTIC PROCEDURE. THE PROCEDURE WAS STOPPED 3 TIMES IN ORDER TO CHANGE OUT THE NEEDLE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19395901 ↗MalfunctionMay 23, 2024Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THE SINGLE USE INJECTOR HAD RESISTANCE DURING AN INJECTION OF BOTOX OR DEXAMETHASONE AND WHILE DISENGAGING AND FLUSHING THE DEVICE, THE FLUID WENT OUT WITHOUT RESISTANCE. THE ISSUE OCCURRED DURING A THERAPEUTIC PROCEDURE. THE PROCEDURE WAS STOPPED 3 TIMES IN ORDER TO CHANGE OUT THE NEEDLE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19388771 ↗MalfunctionJul 17, 2023Activation Failure; Activation Failure
THE CUSTOMER REPORTED TO OLYMPUS THAT THE SINGLE USE INJECTOR LOCKS AND YOU CAN¿T INJECT ANYTHING WHEN YOU FULLY ENGAGE THE NEEDLE. SO, WE MUST USE A SECOND ONE AND WASTE WHAT WE ALREADY TRIED TO INJECT. THERE WAS NO REPORT OF PATIENT HARM OR USER INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17339407 ↗MalfunctionJul 8, 2022No Flow; No Flow
AS REPORTED FOR THIS EVENT BY THE CUSTOMER, DURING A DIAGNOSTIC LARGE INTESTINE EXAMINATION PROCEDURE WHEN THE DEVICE WAS USED INSIDE THE PATIENT BODY, THE LIQUID DID NOT COME OUT AND WAS NOT SENT TO THE PATIENT BODY. ANOTHER DEVICE OF THE SAME MODEL NUMBER WAS TESTED OUTSIDE THE PATIENT BODY. THE SECOND DEVICE DID NOT EJECT ANY LIQUID EITHER. A THIRD DEVICE WAS USED AS REPLACEMENT AND THE PROCEDURE COMPLETED WITH NO MORE ISSUES. THERE IS NO HARM OR ADVERSE IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14960968 ↗MalfunctionMar 2, 2022No Flow; No Flow
AS REPORTED, THE ENT DEVICE WAS REPORTED TO FAIL UPON INITIAL CLINICAL USE. THE OPERATOR CONNECTED THE NM-401L-0525 SINGLE USE INJECTOR AND TRIED TO PASS FLUID THROUGH THE NEEDLE, COULD NOT PUSH ANY FLUID, IT WAS BLOCKED. THE CUSTOMER TRIED THREE DIFFERENT INJECTORS FROM THIS LOT WITH NO SUCCESS. CUSTOMER DID NOT COMPLETE THE INJECTION PART OF THE CASE. THERE WAS NO PATIENT HARM OR INJURY REPORTED DUE TO THE EVENT. NO USER INJURY REPORTED THIS REPORT IS RELATED TO REPORT WITH PATIENT IDENTIFIER (B)(6) AND (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13654293 ↗MalfunctionMar 2, 2022No Flow; No Flow
AS REPORTED, THE ENT DEVICE WAS REPORTED TO FAIL UPON INITIAL CLINICAL USE. THE OPERATOR CONNECTED THE NM-401L-0525 SINGLE USE INJECTOR AND TRIED TO PASS FLUID THROUGH THE NEEDLE, COULD NOT PUSH ANY FLUID, IT WAS BLOCKED. THE CUSTOMER TRIED THREE DIFFERENT INJECTORS FROM THIS LOT WITH NO SUCCESS. CUSTOMER DID NOT COMPLETE THE INJECTION PART OF THE CASE. THERE WAS NO PATIENT HARM OR INJURY REPORTED DUE TO THE EVENT. NO USER INJURY REPORTED THIS REPORT IS RELATED TO REPORT WITH PATIENT IDENTIFIER (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13654285 ↗MalfunctionMar 2, 2022No Flow; No Flow
AS REPORTED, THE ENT DEVICE WAS REPORTED TO FAIL UPON INITIAL CLINICAL USE. THE OPERATOR CONNECTED THE NM-401L-0525 SINGLE USE INJECTOR AND TRIED TO PASS FLUID THROUGH THE NEEDLE, COULD NOT PUSH ANY FLUID, IT WAS BLOCKED. THE CUSTOMER TRIED THREE DIFFERENT INJECTORS FROM THIS LOT WITH NO SUCCESS. CUSTOMER DID NOT COMPLETE THE INJECTION PART OF THE CASE. THERE WAS NO PATIENT HARM OR INJURY REPORTED DUE TO THE EVENT. NO USER INJURY REPORTED. THIS REPORT IS RELATED TO REPORT WITH PATIENT IDENTIFIER (B)(6)AND (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13654280 ↗
Other endoscopic injection needle, gastroenterology-urology devices
- AcuJect G22526Cook Incorporated · endoscopic injection needle, gastroenterology-urology1 recall1 report
- n.a. 27182AKarl Storz GmbH & Co. KG · endoscopic injection needle, gastroenterology-urology1 recall0 reports
- n.a. 27030NKarl Storz GmbH & Co. KG · endoscopic injection needle, gastroenterology-urology1 recall0 reports
- injeTAK Adjustable Tip Needle 70cm DIS201Laborie Medical Technologies Canada ULC · endoscopic injection needle, gastroenterology-urology1 recall0 reports
- injeTAK Adjustable Tip Needle 35cm DIS199Laborie Medical Technologies Canada ULC · endoscopic injection needle, gastroenterology-urology1 recall0 reports
- Single Use Injector NM-401L-0425Olympus Medical Systems Corp. · endoscopic injection needle, gastroenterology-urology28 reports
- Single Use Injector NM-401L-0423Olympus Medical Systems Corp. · endoscopic injection needle, gastroenterology-urology27 reports
- Single Use Injector NM-401L-0425Olympus Medical Systems Corp. · endoscopic injection needle, gastroenterology-urology26 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page