Skip to content

Single Use BML Handle V MAJ-2502Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04953170433658), received between Apr 30, 2024 and Jul 17, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 17, 2025Break; Mechanical Jam; Break; Mechanical Jam

    IT WAS REPORTED THAT THE SUBJECT DEVICE EXHIBITED THE FOLLOWING ISSUE: AFTER THE GALL STONE WAS CRUSHED, THE HANDLE COULD NOT BE RELEASED FROM POSITION ¿2¿ AND WAS JAMMED IN THE RATCHET POSITION. HANDLE BROKE WHILST TRYING TO DISENGAGE RATCHET WHILST INSIDE PATIENT. THE ISSUE OCCURRED DURING A THERAPEUTIC ERCP (ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY). THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22535511 ↗
  • MalfunctionJun 24, 2025Break; Break

    IT WAS REPORTED THAT DURING THE ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ECPW) PROCEDURE, THE STONE HAD ALREADY BEEN CAUGHT AND WAS ABOUT TO BE CRUSHED, THE KNOB FOR CRUSHING/CLOSING THE BASKET BROKE OFF WITH THE SUBJECT DEVICE STUCK. THE PROCEDURE WAS COMPLETED BY REPLACING IT WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22322844 ↗
  • InjuryApr 30, 2024Break; Break

    IT WAS REPORTED THAT DURING LITHOTRIPSY, THE ANCHORING OF THE LITHOTRIPTER BASKET WAS TORN OUT AND THE HANDLE WAS BROKEN. EMERGENCY LITHOTRIPSY HAD TO PERFORMED USING A METAL COIL, DURING WHICH NOT ONLY WAS THE STONE NOT FRAGMENTED, BUT INSTEAD OF A FRACTURE ON THE CUP ITSELF, ONLY ONE OF THE WIRES AT THE TOP WAS TORN. FURTHER BASKETS THEN GOT CAUGHT IN THE OLYMPUS BASKET, AND A CHOLANGIOSCOPY WITH LASER LITHOTRIPSY COULD ONLY PARTIALLY FRAGMENT THE STONE. THE PATIENT HAD TO UNDERGO EMERGENCY SURGERY AFTER A 3.5-HOUR ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WITH BASKETS THAT COULD NOT BE REMOVED. THE BASKET WAS REMOVED DURING THE EMERGENCY OPERATION. REPORTEDLY, APART FROM THE STONE ITSELF AND FAILURE OF THE PRIMARY LITHOTRIPSY, THE MAIN PROBLEM WAS THAT THE BASKET WAS NOT SUITABLE FOR EMERGENCY LITHOTRIPSY VIA SPIRAL. IT WAS FURTHER ELABORATED, THERE WAS A SMOOTH PASSAGE TO THE PARS DESCENDENS DUODENI, WHERE A PLASTIC STENT IN-SITU WAS REMOVED. THE PARS DESCENDENS DUODENI WAS RE-ACCESSED AND IT WAS NOTED THAT THE OSTIUM OF THE BILE DUCT WAS LYING DEEP IN A DIVERTICULUM. CANNULATION USING AN ERCP CATHETER AND WIRE WAS DONE WITH WIRE ADVANCEMENT TO THE INTRAHEPATIC. CON

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19216055 ↗

Other Lithotriptor, Biliary Mechanical devices

All Lithotriptor, Biliary Mechanical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page