Skip to content

PENTAX Medical EC38-i20cLRecalls & reported problems

Hoya Corporation · PENTAX Medical Video Colonoscope EC38-i20cL · Class 2

All models of PENTAX Medical (34)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Pentax Medical video colonoscope is a flexible scope used to view the colon. Its 182 reports include no deaths and 2 injuries. Many describe fluid or blood leaking into the scope, a blocked lens, and no image on screen. There is 1 open recall, started in January 2025, over reddish or dark images and a tip that can get hot during use.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
185
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 185 adverse event reports that list this exact device (DI 04961333252169), received between Sep 18, 2023 and Sep 30, 2026 — classified as 2 injuries, 183 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (53 reports).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K231249); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 6 reportsApr 2024: 0 reportsMay 2024: 5 reportsJun 2024: 1 reportsJul 2024: 4 reportsAug 2024: 1 reportsSep 2024: 3 reportsOct 2024: 6 reportsNov 2024: 7 reportsDec 2024: 2 reportsJan 2025: 7 reportsFeb 2025: 8 reportsMar 2025: 3 reportsApr 2025: 4 reportsMay 2025: 4 reportsJun 2025: 4 reportsJul 2025: 10 reportsAug 2025: 5 reportsSep 2025: 9 reportsOct 2025: 3 reportsNov 2025: 3 reportsDec 2025: 5 reportsJan 2026: 5 reportsFeb 2026: 12 reportsMar 2026: 5 reportsApr 2026: 8 reportsMay 2026: 19 reportsJun 2026: 4 reportsJul 2026: 10 reportsAug 2026: 10 reportsSep 2026: 7 reportsOct 2026: 0 reports20242025202619
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 29, 2025Class IIOpen, Classified

    During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.

    Z-1236-2025 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak53
  • Optical Obstruction41
  • No Display/Image38
  • Erratic or Intermittent Display15
  • Obstruction of Flow11
  • Poor Quality Image9
  • Break5
  • Material Split, Cut or Torn5

Patient problems

  • No Clinical Signs, Symptoms or Conditions183
  • Foreign Body In Patient2
  • Gastrointestinal Hemorrhage1

Recent adverse event reports

The 25 most recent of 185 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Erratic or Intermittent Display

    The time of event is unknown.There was no report of patient harm.Video image failure.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942343 ↗
  • MalfunctionSep 29, 2026Material Split, Cut or Torn

    The time of event is unknown.There was no report of patient harm.Insertion flexible tube (IFT) Steel braid piercing.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50927543 ↗
  • MalfunctionSep 25, 2026No Display/Image

    The time of event is unknown.There was no report of patient harm.Video image (Blackout).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50893568 ↗
  • MalfunctionSep 24, 2026No Display/Image

    The time of event is unknown.There was no report of patient harm.Video image failure (Blackout).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50877471 ↗
  • MalfunctionSep 14, 2026Optical Obstruction

    The time of event is unknown.There was no report of patient harm.Video image failure(Shadow in image).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50749376 ↗
  • MalfunctionSep 8, 2026Optical Obstruction

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50661471 ↗
  • MalfunctionSep 2, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50558337 ↗
  • MalfunctionAug 28, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50490853 ↗
  • MalfunctionAug 28, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50490847 ↗
  • MalfunctionAug 18, 2026Optical Obstruction

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50292699 ↗
  • MalfunctionAug 17, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50269334 ↗
  • MalfunctionAug 13, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50232156 ↗
  • MalfunctionAug 12, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50217984 ↗
  • MalfunctionAug 12, 2026Erratic or Intermittent Display

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50217977 ↗
  • MalfunctionAug 3, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50111527 ↗
  • MalfunctionAug 3, 2026Erratic or Intermittent Display

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50111519 ↗
  • MalfunctionAug 3, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50111517 ↗
  • MalfunctionJul 30, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50079010 ↗
  • MalfunctionJul 29, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50064965 ↗
  • MalfunctionJul 29, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50064958 ↗
  • MalfunctionJul 28, 2026Optical Obstruction

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50049586 ↗
  • MalfunctionJul 21, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25884947 ↗
  • MalfunctionJul 21, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25884937 ↗
  • MalfunctionJul 15, 2026Optical Obstruction

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25822219 ↗
  • MalfunctionJul 15, 2026Obstruction of Flow

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25822214 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K231249. One submission often covers many models, so check the product and lot details in each record.

  • Z-2329-2025 · Jul 16, 2025 · Open, Classified
    Pentax Medical Video Processor; Model Number: EPK-i8020c;
  • Z-1238-2025 · Jan 29, 2025 · Open, Classified
    Pentax Medical Video Processor- Intended to be used with PENTAX Medical endoscopes, monitors and other peripheral devices for endoscopic diagnosis, treatment, and video observation
  • Z-1237-2025 · Jan 29, 2025 · Open, Classified
    PENTAX Medical Video Upper GI Scope-Intended to provide optical visualization of (via a video monitor), and therapeutic access to the lower gastrointestinal tract Model: EG29-i20c

Other Endoscope, Accessories, Image Post-Processing For Color Enhancement devices

All Endoscope, Accessories, Image Post-Processing For Color Enhancement devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page