NAVICROSS NC35900Recalls & reported problems
Terumo Corporation · NAVICROSS Support Catheter · Class 2
All models of NAVICROSS (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04987350709264), received between Jun 19, 2018 and Jun 27, 2019 — classified as 3 injuries.
The most frequently reported specific device problem is “Material Separation” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Separation3
Patient problems
- Device Embedded In Tissue or Plaque3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJun 27, 2019Material Separation; Material Separation
THE USER FACILITY REPORTED THAT TWO OF THE NAVICROSS RADIOPAQUE MARKERS CAME OFF THE DEVICE DURING USE AND REMAINED IN THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED ON JUNE 3, 2019. PER HEALTH (B)(6) REPORT ((B)(4)): "DURING PROCEDURES ENDOVASCULAR FOR (ILIAC STENTING) TWO SEPARATE RADIO OPAQUE MARKERS CAME OFF, TWO DIFFERENT NAVICROSS SUPPORT CATHETERS INTO THE PATIENT." ADDITIONAL INFORMATION WAS RECEIVED ON 26JUN2019. THE INCIDENT WAS REGARDING TWO NAVICROSS CATHETERS TO THE SAME PATIENT. NO SURGICAL INTERVENTION WAS REQUIRED TO TREAT THE PATIENT FOR THE RADIO OPAQUE MARKERS. AFTER THE SECOND NAVICROSS CATHETER LOST ITS RADIO OPAQUE MARKERS, THEY SWITCHED TO A DIFFERENT PRODUCT. A COOK MEDICAL J-WIRE, GUIDE WIRE OR BENSON WIRE WAS BEING USED ALONG WITH THE NAVICROSS.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 8740883 ↗InjuryJun 27, 2019Material Separation; Material Separation
THE USER FACILITY REPORTED THAT TWO OF THE NAVICROSS RADIOPAQUE MARKERS CAME OFF THE DEVICE DURING USE AND REMAINED IN THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED ON JUNE 3, 2019. PER HEALTH CANADA REPORT ((B)(4)): "DURING PROCEDURES ENDOVASCULAR FOR (ILIAC STENTING) TWO SEPARATE RADIO OPAQUE MARKERS CAME OFF, TWO DIFFERENT NAVICROSS SUPPORT CATHETERS INTO THE PATIENT." ADDITIONAL INFORMATION WAS RECEIVED ON JUNE 26, 2019. THE INCIDENT WAS REGARDING TWO NAVICROSS CATHETERS TO THE SAME PATIENT. NO SURGICAL INTERVENTION WAS REQUIRED TO TREAT THE PATIENT FOR THE RADIO OPAQUE MARKERS. AFTER THE SECOND NAVICROSS CATHETER LOST ITS RADIO OPAQUE MARKERS, THEY SWITCHED TO A DIFFERENT PRODUCT. A COOK MEDICAL J-WIRE, GUIDE WIRE OR BENSON WIRE WAS BEING USED ALONG WITH THE NAVICROSS.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 8740882 ↗InjuryJun 19, 2018Material Separation; Material Separation
ADDITIONAL INFORMATION WAS RECEIVED JUNE 25, 2018. THE GOLD MARKER WAS NOT REMOVED FROM THE PATIENT. THE PATIENT WAS REPORTED TO BE IN STABLE CONDITION.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 7614768 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page