Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04987350722188), received between Feb 23, 2016 and Mar 17, 2016 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Activation, Positioning or Separation Problem2

Patient problems

  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 17, 2016Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem

    THE USER FACILITY REPORTED THAT THE MISAGO DEVICE WOULD NOT FULLY DEPLOY. FOLLOW UP COMMUNICATION WITH THE USER FACILITY CONFIRMED THE FOLLOWING INFORMATION: THE PROCEDURE WAS FOR A PERIPHERAL CASE; THE MISAGO DEVICE WAS PREPARED; THE TIP WAS FLUSHED WITH SALINE BUT IT DID NOT DRIP OUT OF THE CATHETER; THE STENT WAS LOADED OVER THE .035 GLIDEWIRE; RESISTANCE WAS FELT IMMEDIATELY WHEN GOING THROUGH THE VALVE OF THE 30CM COOK SHEATH; THE DOCTOR TOOK AN IMAGE, THE STENT WAS PARTIALLY DEPLOYED; THE DOCTOR DECIDED TO FINISH DEPLOYING THE STENT, AND IT CAME OUT TWISTED AT BOTH ENDS; AND (THE DOCTOR ATTEMPTED TO BALLOON THE STENT, AND DEPLOYED A 100CM STENT THROUGH THE TWISTED STENT AND BALLOONED SEVERAL TIMES TO FIX THE TWISTED STENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5506788 ↗
  • MalfunctionFeb 23, 2016Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem

    THE USER FACILITY REPORTED THAT THE MISAGO DEVICE DEPLOYED INSIDE OF THE DESTINATION DEVICE. FOLLOW UP COMMUNICATION WITH THE USER FACILITY CONFIRMED THE FOLLOWING INFORMATION: DURING DEPLOYMENT OF THE MISAGO STENT IT STARTED TO DRIFT PROXIMALLY FROM THE INITIAL POSITION; THE DEVICE WAS MOVE BACK INTO THE INVOLVED DESTINATION AND DEPLOYED; THE STENT WAS STRETCHED TO APPROXIMATELY 100MM POSSIBLY DUE TO SOME PULLING FORCE TO THE MISAGO DEVICE; THE STENT WAS LEFT IMPLANTED IN THE PATIENT; THE PROCEDURE WAS COMPLETED; AND THE PATIENT WAS NOT IMPACTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5453529 ↗

Other Stent, Superficial Femoral Artery devices

All Stent, Superficial Femoral Artery devices →