Skip to content

R2P MISAGO SXR08040RRecalls & reported problems

Terumo Corporation · MISAGO RX Self-expanding Peripheral Stent · Class 3 · Implant

All models of R2P MISAGO (13)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04987350727640), received between Apr 27, 2023 and Nov 7, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Difficult to Insert1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 7, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT WHILE LOADING THE STENT ONTO THE GUIDEWIRE, THE STENT WAS ACCIDENTALLY RELEASED DUE TO THE ROLLER BALL BEING DEPRESSED UPON OPENING THE PACKAGE. A SECOND MISAGO STENT WAS USED SUCCESSFULLY IN THE PROCEDURE. THERE WAS NO INJURY TO THE PATIENT AND NO BLOOD LOSS. THE PROCEDURE PERFORMED WAS A LEG AV FISTULA. ADDITIONAL INFORMATION RECEIVED ON 15 OCT 2024: THEY DID NOT DEPRESS THE ROLLER BALL BUT WERE UNSURE OF WHAT OCCURRED BETWEEN OPENING THE PACKAGE AND STARTING TO USE IT. THIS INCIDENT DID NOT CAUSE ANY HARM DURING THE PROCEDURE OR TO THE PATIENT. THE STENT NEVER ENTERED THE SHEATH OR THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20627005 ↗
  • MalfunctionApr 27, 2023Difficult to Insert; Physical Resistance/Sticking; Difficult to Insert; Physical Resistance/Sticking

    THE USER FACILITY REPORTED THAT THE DOCTOR PERFORMED A RADIAL CASE WITH AN INTERVENTION OF THE PROXIMAL SUPERFICIAL FEMORAL ARTERY (SFA). WHEN THE DOCTOR WENT TO PASS THE STENT INTO THE INVOLVED DESTINATION SLENDER GUIDING SHEATH THERE WAS RESISTANCE AT THE HUB OF THE SHEATH. THE PROXIMAL END OF THE STENT WAS EXAMINED AND NOTICED THAT THE END OF THE STENT WAS OUT OF THE PROTECTIVE SHEATH. THE MARKERS HAD STARTED TO FLARE, AND THE THUMBWHEEL HAD NOT BEEN DEPRESSED OR ROLLED BACKWARDS. THE PROXIMAL STENT EDGE HAD SHIFTED PRIOR TO THE INTRODUCTION INTO THE GUIDING SHEATH. THE THUMBWHEEL HAD BEEN VERIFIED THAT IT WAS NOT ALTERED BY THE TECH DURING PREP OR PRE DEPLOYMENT. THE STENT NEVER MADE IT INTO THE PATIENT AS IT MET RESISTANCE IN THE SHEATH. THE CASE WAS COMPLETED WITH ANOTHER STENT. THE PATIENT WAS IN FINE CONDITION. THE PROCEDURE OUTCOME WAS SUCCESSFUL. THE PATIENT WAS NOT INJURED, MEDICAL OR SURGICAL INTERVENTION WAS NOT REQUIRED. THE EVENT OCCURRED INTRA-OPERATIVE. THE ARTERY OF THE LOWER EXTREMITY WAS NOT PREVIOUSLY TREATED. THE LESION LOCATION WAS AT THE SFA, WITH MODERATE CALCIFICATION AND NO TORTUOUSNESS. THE LESION LENGTH WAS APPROXIMATELY 30MM, WITH 80% DIAMETER STENOSIS

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16826930 ↗

Other STENT, SUPERFICIAL FEMORAL ARTERY devices

All STENT, SUPERFICIAL FEMORAL ARTERY devices →