Terumo Corporation · MISAGO RX Self-expanding Peripheral Stent · Class 3 · Implant

All models of R2P MISAGO (13)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04987350727763), received between Jun 7, 2025 and Jun 7, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Activation Failure” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Failure2
  • Device Dislodged or Dislocated2
  • Material Fragmentation2

Patient problems

  • Foreign Body Embolism2
  • Pain2
  • Reduced Blood Flow2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 7, 2025Material Fragmentation; Device Dislodged or Dislocated; Activation Failure; Material Fragmentation; Device Dislodged or Dislocated; Activation Failure

    THE USER FACILITY DURING A LOWER EXTREMITY ANGIOGRAM WITH INTERVENTION VIA CONTRALATERAL FEMORAL ACCESS, A PHYSICIAN SUCCESSFULLY DEPLOYED A 7MM X 150CM MISAGO STENT IN THE SUPERFICIAL FEMORAL ARTERY (SFA) USING A .035 ROSEN WIRE. A SECOND STENT, A 7MM X 120CM MISAGO, WAS THEN PREPARED FOR IMPLANTATION. ACCORDING TO THE TECHNOLOGIST, THE STENT WAS FLUSHED WITH A SYRINGE PRIOR TO USE. UPON INITIATING DEPLOYMENT BY ROLLING THE WHEEL ON THE DELIVERY SYSTEM, THE PROXIMAL SHAFT OF THE MISAGO DELIVERY SYSTEM BENT OR KINKED AT A 90-DEGREE ANGLE AT THE STRAIN RELIEF OF THE HANDLE, WHICH PREVENTED FULL DEPLOYMENT OF THE STENT. THE PHYSICIAN ATTEMPTED TO RETRIEVE THE PARTIALLY DEPLOYED STENT, BUT IT BEGAN TO FRAGMENT WITHIN THE PATIENT. THE SHEATH HAD TO BE REMOVED DUE TO THE COMPROMISED STENT. WIRE ACCESS WAS MAINTAINED, AND A NEW SHEATH WAS INTRODUCED. A FRAGMENT OF THE STENT WAS LOCATED WITHIN THE TISSUE TRACT AT THE ARTERIOTOMY SITE AND WAS SUCCESSFULLY REMOVED BEFORE THE NEW SHEATH WAS PLACED. THE PHYSICIAN PROCEEDED TO PERFORM ANGIOPLASTY ON THE SFA AND DEPLOYED A PROTEGE STENT. HOWEVER, A FRAGMENT OF THE MISAGO STENT REMAINED LODGED IN THE ILIAC ARTERY. THE TECHNOLOGIST REPORTED THAT

    Patient problems: Foreign Body Embolism; Pain; Reduced Blood Flow

    MAUDE report 22162298 ↗
  • InjuryJun 7, 2025Material Fragmentation; Device Dislodged or Dislocated; Activation Failure; Material Fragmentation; Device Dislodged or Dislocated; Activation Failure

    THE USER FACILITY DURING A LOWER EXTREMITY ANGIOGRAM WITH INTERVENTION VIA CONTRALATERAL FEMORAL ACCESS, A PHYSICIAN SUCCESSFULLY DEPLOYED A 7MM X 150CM MISAGO STENT IN THE SUPERFICIAL FEMORAL ARTERY (SFA) USING A .035 ROSEN WIRE. A SECOND STENT, A 7MM X 120CM MISAGO, WAS THEN PREPARED FOR IMPLANTATION. ACCORDING TO THE TECHNOLOGIST, THE STENT WAS FLUSHED WITH A SYRINGE PRIOR TO USE. UPON INITIATING DEPLOYMENT BY ROLLING THE WHEEL ON THE DELIVERY SYSTEM, THE PROXIMAL SHAFT OF THE MISAGO DELIVERY SYSTEM BENT OR KINKED AT A 90-DEGREE ANGLE AT THE STRAIN RELIEF OF THE HANDLE, WHICH PREVENTED FULL DEPLOYMENT OF THE STENT. THE PHYSICIAN ATTEMPTED TO RETRIEVE THE PARTIALLY DEPLOYED STENT, BUT IT BEGAN TO FRAGMENT WITHIN THE PATIENT. THE SHEATH HAD TO BE REMOVED DUE TO THE COMPROMISED STENT. WIRE ACCESS WAS MAINTAINED, AND A NEW SHEATH WAS INTRODUCED. A FRAGMENT OF THE STENT WAS LOCATED WITHIN THE TISSUE TRACT AT THE ARTERIOTOMY SITE AND WAS SUCCESSFULLY REMOVED BEFORE THE NEW SHEATH WAS PLACED. THE PHYSICIAN PROCEEDED TO PERFORM ANGIOPLASTY ON THE SFA AND DEPLOYED A PROTEGE STENT. HOWEVER, A FRAGMENT OF THE MISAGO STENT REMAINED LODGED IN THE ILIAC ARTERY. THE TECHNOLOGIST REPORTED THAT

    Patient problems: Foreign Body Embolism; Pain; Reduced Blood Flow

    MAUDE report 22162287 ↗