R2P MISAGO SXR07120RRecalls & reported problems
Terumo Corporation · MISAGO RX Self-expanding Peripheral Stent · Class 3 · Implant
All models of R2P MISAGO (13)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04987350727763), received between Jun 7, 2025 and Jun 7, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Activation Failure” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Activation Failure2
- Device Dislodged or Dislocated2
- Material Fragmentation2
Patient problems
- Foreign Body Embolism2
- Pain2
- Reduced Blood Flow2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJun 7, 2025Material Fragmentation; Device Dislodged or Dislocated; Activation Failure; Material Fragmentation; Device Dislodged or Dislocated; Activation Failure
THE USER FACILITY DURING A LOWER EXTREMITY ANGIOGRAM WITH INTERVENTION VIA CONTRALATERAL FEMORAL ACCESS, A PHYSICIAN SUCCESSFULLY DEPLOYED A 7MM X 150CM MISAGO STENT IN THE SUPERFICIAL FEMORAL ARTERY (SFA) USING A .035 ROSEN WIRE. A SECOND STENT, A 7MM X 120CM MISAGO, WAS THEN PREPARED FOR IMPLANTATION. ACCORDING TO THE TECHNOLOGIST, THE STENT WAS FLUSHED WITH A SYRINGE PRIOR TO USE. UPON INITIATING DEPLOYMENT BY ROLLING THE WHEEL ON THE DELIVERY SYSTEM, THE PROXIMAL SHAFT OF THE MISAGO DELIVERY SYSTEM BENT OR KINKED AT A 90-DEGREE ANGLE AT THE STRAIN RELIEF OF THE HANDLE, WHICH PREVENTED FULL DEPLOYMENT OF THE STENT. THE PHYSICIAN ATTEMPTED TO RETRIEVE THE PARTIALLY DEPLOYED STENT, BUT IT BEGAN TO FRAGMENT WITHIN THE PATIENT. THE SHEATH HAD TO BE REMOVED DUE TO THE COMPROMISED STENT. WIRE ACCESS WAS MAINTAINED, AND A NEW SHEATH WAS INTRODUCED. A FRAGMENT OF THE STENT WAS LOCATED WITHIN THE TISSUE TRACT AT THE ARTERIOTOMY SITE AND WAS SUCCESSFULLY REMOVED BEFORE THE NEW SHEATH WAS PLACED. THE PHYSICIAN PROCEEDED TO PERFORM ANGIOPLASTY ON THE SFA AND DEPLOYED A PROTEGE STENT. HOWEVER, A FRAGMENT OF THE MISAGO STENT REMAINED LODGED IN THE ILIAC ARTERY. THE TECHNOLOGIST REPORTED THAT
Patient problems: Foreign Body Embolism; Pain; Reduced Blood Flow
MAUDE report 22162298 ↗InjuryJun 7, 2025Material Fragmentation; Device Dislodged or Dislocated; Activation Failure; Material Fragmentation; Device Dislodged or Dislocated; Activation Failure
THE USER FACILITY DURING A LOWER EXTREMITY ANGIOGRAM WITH INTERVENTION VIA CONTRALATERAL FEMORAL ACCESS, A PHYSICIAN SUCCESSFULLY DEPLOYED A 7MM X 150CM MISAGO STENT IN THE SUPERFICIAL FEMORAL ARTERY (SFA) USING A .035 ROSEN WIRE. A SECOND STENT, A 7MM X 120CM MISAGO, WAS THEN PREPARED FOR IMPLANTATION. ACCORDING TO THE TECHNOLOGIST, THE STENT WAS FLUSHED WITH A SYRINGE PRIOR TO USE. UPON INITIATING DEPLOYMENT BY ROLLING THE WHEEL ON THE DELIVERY SYSTEM, THE PROXIMAL SHAFT OF THE MISAGO DELIVERY SYSTEM BENT OR KINKED AT A 90-DEGREE ANGLE AT THE STRAIN RELIEF OF THE HANDLE, WHICH PREVENTED FULL DEPLOYMENT OF THE STENT. THE PHYSICIAN ATTEMPTED TO RETRIEVE THE PARTIALLY DEPLOYED STENT, BUT IT BEGAN TO FRAGMENT WITHIN THE PATIENT. THE SHEATH HAD TO BE REMOVED DUE TO THE COMPROMISED STENT. WIRE ACCESS WAS MAINTAINED, AND A NEW SHEATH WAS INTRODUCED. A FRAGMENT OF THE STENT WAS LOCATED WITHIN THE TISSUE TRACT AT THE ARTERIOTOMY SITE AND WAS SUCCESSFULLY REMOVED BEFORE THE NEW SHEATH WAS PLACED. THE PHYSICIAN PROCEEDED TO PERFORM ANGIOPLASTY ON THE SFA AND DEPLOYED A PROTEGE STENT. HOWEVER, A FRAGMENT OF THE MISAGO STENT REMAINED LODGED IN THE ILIAC ARTERY. THE TECHNOLOGIST REPORTED THAT
Patient problems: Foreign Body Embolism; Pain; Reduced Blood Flow
MAUDE report 22162287 ↗
Other STENT, SUPERFICIAL FEMORAL ARTERY devices
- EverFlex™ EVD35-06-150-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 death110 reports
- SUPERA S-60-150-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY108 reports
- SUPERA S-55-150-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY88 reports
- LifeStent® Solo Vascular Stent EX062003CLBard Peripheral Vascular, Inc. · STENT, SUPERFICIAL FEMORAL ARTERY80 reports
- BioMimics 3D Vascular Stent System 144700-12Veryan Medical Limited · Stent, Superficial Femoral Artery74 reports
- SUPERA S-50-120-120-P6Abbott Vascular Inc. · STENT, SUPERFICIAL FEMORAL ARTERY55 reports
- EverFlex™ EVD35-07-060-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 recall5 reports
- EverFlex™ EVD35-07-120-120Covidien LP · STENT, SUPERFICIAL FEMORAL ARTERY1 recall5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page