Fully Automated Urine Particle Analyzer UF-5000 BN344411Recalls & reported problems
Sysmex Corporation · Fully Automated Urine Particle Analyzer UF-5000 · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04987562440139), received between Jun 13, 2019 and Aug 10, 2020 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “High Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- High Test Results1
- Improper or Incorrect Procedure or Method1
- Smoking1
Patient problems
- No Known Impact Or Consequence To Patient2
- No Patient Involvement1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryAug 10, 2020Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
A USER IN THE (B)(6) REPORTED A URINE CULTURE WAS DELAYED DUE TO FALSE LOW WHITE BLOOD CELL (WBC) AND BACTERIA (BACT) RESULTS. THE SAMPLE WAS ANALYZED, AND LOW WBC AND BACT RESULTS WERE GENERATED. THE RESULTS WERE RELEASED TO THE LABORATORY INFORMATION SYSTEM (LIS). THE NEXT DAY, THE USER REPORTED WBC QUALITY CONTROL (QC) WAS RECOVERING LOW OUT OF RANGE. DURING INVESTIGATION, IT WAS FOUND THE UF-CELLPACK CR REAGENT HAD INADVERTENTLY BEEN EMPTIED AND FILLED WITH DEIONIZED WATER ON ALL ANALYZERS DURING WEEKLY MAINTENANCE. WEEKLY MAINTENANCE CONSISTS OF EMPTYING AND REPLENISHING THE RINSE WATER, AND THE RINSE WATER BOTTLE IS SIMILAR IN APPEARANCE TO THE UF-CELLPACK CR REAGENT. THE SAMPLE WAS REANALYZED ON A DIFFERENT ANALYZER AFTER THE UF-CELLPACK CR REAGENT WAS PROPERLY INSTALLED. THE ANALYZER GENERATED HIGHER WBC AND BACT RESULTS. NO PATIENT HARM WAS ALLEGED DUE TO THE CULTURE DELAY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10385348 ↗InjuryJan 31, 2020High Test Results; High Test Results
AN OPERATOR IN DENMARK ALLEGED A CEREBROSPINAL FLUID (CSF) SAMPLE WAS ANALYZED AND GENERATED FALSELY ELEVATED POLYMORPHONUCLEAR LEUKOCYTES (PMN) AND FALSELY DECREASED MONONUCLEAR CELLS (MN). THERE WERE NO ERROR MESSAGES OR FLAGS WITH THE RESULTS. THESE RESULTS WERE NOT VERIFIED WITH MICROSCOPIC EVALUATION. DUE TO DUE TO THE PATIENT'S CLINICAL SIGNS AND SYMPTOMS AND RESULTS FROM THE CSF SAMPLE, THE PATIENT RECEIVED ANTIBIOTICS AND ANTI-VIRAL TREATMENT BEGINNING ON (B)(6) 2020. ADDITIONAL TESTS FOR BLOOD CULTURE AND CSF CULTURE WERE INITIATED. SIX DAYS LATER, ANOTHER CSF SAMPLE WAS COLLECTED AND ANALYZED, WHICH GENERATED LOWER PMN RESULTS AND HIGHER MN RESULTS AS COMPARED TO THE INITIAL CSF VALUES. THESE RESULTS WERE CONFIRMED BY MICROSCOPIC EVALUATION. ANTIBIOTIC TREATMENT WAS DISCONTINUED AND STEROID TREATMENT WAS STARTED BASED ON THE SECOND CSF SAMPLE RESULTS. NO PATIENT HARM WAS REPORTED BASED ON THE ADMINISTRATION OF ANTIBIOTIC TREATMENT, DELAY OF STEROID TREATMENT OR THE SECOND CSF COLLECTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9655919 ↗MalfunctionJun 13, 2019Smoking; Smoking
IN (B)(6), THE USER OF A SYSMEX UF-5000 AUTOMATED URINE PARTICLE ANALYZER WITH A CV-11, URINE SAMPLE TRANSPORTATION UNIT, SERIAL NUMBER (SN) (B)(4), REPORTED SPARKS, SMOKE AND A BURNING ODOR COMING FROM THE CV-11 FOLLOWING A SPILL FROM A RACK OF URINE SAMPLES. THE USER REPORTED THAT THE CV-11 HAD CEASED OPERATION UNEXPECTEDLY, PROMPTING A RE-INITIALIZATION. UPON RE-INITIALIZATION, THE SAMPLE RACK WAS PUSHED TOWARDS THE LEFT RACK POOL AREA THAT COLLECTS SAMPLES AFTER ANALYSIS AND TIPPED OVER, SPILLING THE URINE ONTO THE CONVEYOR. THE USER IMMEDIATELY CLEANED THE AREA, BUT WITHIN AN HOUR OF THE SPILL, THE USER NOTED SPARKS, SMOKE AND A BURNING ODOR COMING FROM THE CV-11. A FIRE EXTINGUISHER WAS DEPLOYED AT THE SOURCE OF THE SMOKE. THERE WERE NO REPORTS OF HARM TO THE USERS OR PATIENTS DUE TO THE EVENT. A NEW CV-11 WAS INSTALLED TO RESOLVE THE ISSUE.
Patient problems: No Patient Involvement
MAUDE report 8697169 ↗
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page