Trima Accel® 82520Recalls & reported problems
Terumo Bct, Inc. · Disposable Tubing Sets, Auto RBC, Plasma Set · Class 2
Quick summary
This Terumo BCT Trima Accel disposable tubing set is used to collect blood cell components, separating red cells and plasma from a donor. Most of the 53 reports describe problems with collected product quality — red blood cells breaking down (hemolysis) and leftover white blood cells showing up in the product — rather than device breakdowns; many reports simply lack detail about what went wrong. Two injuries and no deaths were reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 49
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 49 adverse event reports that list this exact device (DI 05020583825208), received between May 2, 2019 and Dec 26, 2025 — classified as 2 injuries, 47 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (23 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem43
- Insufficient Device Problem Information23
- High Test Results13
- No Apparent Adverse Event5
- Contamination of Device Ingredient or Reagent3
- High Readings3
- Incorrect, Inadequate or Imprecise Result or Readings2
- Device Ingredient or Reagent Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions41
- No Patient Involvement5
- Itching Sensation2
- Rash2
- Hemolysis1
- Respiratory Insufficiency1
- Tinnitus1
Recent adverse event reports
The 25 most recent of 49 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 26, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE RED BLOOD CELL (RBC) PRODUCT. DONOR UNIT ID: (B)(6). THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE UNIT PROCESSING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23896470 ↗MalfunctionDec 11, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE REB BLOOD CELL (RBC) PRODUCT. DONOR UNIT #: (B)(6). THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING; THEREFORE, NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE PLATELET COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23779445 ↗MalfunctionNov 4, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE RED BLOOD CELL (RBC) PRODUCT. DONOR UNIT #: (B)(6) THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING; THEREFORE, NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23466684 ↗MalfunctionOct 30, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE RED BLOOD CELL (RBC) PRODUCT. DONOR UNIT #: (B)(6) THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING; THEREFORE, NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23426576 ↗MalfunctionOct 30, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE RED BLOOD CELL PRODUCT (RBC). DONOR UNIT #: (B)(6) THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE PLATELET COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23426365 ↗MalfunctionOct 30, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE RED BLOOD CELL (RBC) PRODUCT. DONOR UNIT #: (B)(6) THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING; THEREFORE, NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE PLATELET COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23426287 ↗MalfunctionAug 14, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER REPORTED THAT DURING A DOUBLE RED BLOOD CELL PROCEDURE (AUTOMATED 2RBC) HEMOLYSIS WAS OBSERVED AFTER MIDWAY LINE SWITCH (AFTER BAG 1 WAS FILLED, AFTER LINE 2 COLLECTED SOME BLOOD). IT WAS NOTICED IN THE TUBING LINE/RESERVOIR. NO CLOTS WERE OBSERVED IN THE CHANNEL OR CHANNEL LINES. ANOTHER TECHNICIAN RESPONDED TO AN ALARM, BUT PER THE CUSTOMER IT WAS UNCLEAR WHICH ALARM NUMBER PROMPTED THEM TO CHECK; IT WAS SHORTLY AFTER THE LINES WERE SWITCHED TO COLLECT THE SECOND BAG. THEY NOTICED THE RESERVOIR CONTAINED SEPARATED PLASMA AND RED CELLS TOWARDS THE BOTTOM OF THE RESERVOIR. UPON RETURN TO THE DONOR, IT WAS NOTICED THERE WAS SETTLED RED CELLS AT THE BOTTOM OF THE RETURN LINE AND A "STRIPEY" PATTERN OF FLUIDS RETURNING TO THE DONOR. SEEING "CLUMPED" RED CELLS. THEY ENDED THE DONATION IMMEDIATELY AS IT SEEMED MORE LIKE HEMOLYSIS THAN RED CELL SPILLOVER. NO HEMOLYSIS TESTING WAS PERFORMED. THE PLASMA COLOR WAS YELLOW IN THE RESERVOIR, LIGHT PINK/ORANGE IN THE LINE BUT IT MOSTLY LOOKED SEPARATED. NO MEDICAL INTERVENTION WAS REQUIRED FOR THIS EVENT. FULL PATIENT ID: (B)(6) THE COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: Hemolysis
MAUDE report 22806935 ↗MalfunctionFeb 18, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE IN RED BLOOD CELL (RBC) PRODUCT. DONOR UNIT ID #: (B)(6) THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING; THEREFORE, NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21399665 ↗MalfunctionDec 17, 2024Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE RED BLOOD CELL (RBC) PRODUCT. DONOR UNIT ID #: (B)(4). THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING; THEREFORE, NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE PLATELET COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20953937 ↗MalfunctionSep 26, 2023No Apparent Adverse Event; Insufficient Device Problem Information; No Apparent Adverse Event; Insufficient Device Problem Information
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE PLATELET PRODUCT. THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE PLATELET COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17818374 ↗MalfunctionSep 18, 2023Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE DOUBLE RED BLOOD CELL (RBC) PRODUCT. DONOR UNIT #: (B)(6) THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17766618 ↗MalfunctionSep 14, 2023Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE DOUBLE RED CELL COLLECTIONS. CUSTOMER DID NOT RESPOND AFTER 3 ATTEMPTS FOR ADDITIONAL PROCEDURAL INFORMATION. THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17745422 ↗MalfunctionJul 20, 2023Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER REPORTED THAT DURING AN APHERESIS DOUBLE RED BLOOD CELL PROCEDURE AN AIR BUBBLE SMALLER THAN A PEA WAS DELIVERED TO THE DONOR AFTER A NEW NEEDLE WAS STERILE CONNECTED TO THE DRAW LINE. THEY DID PRIME THE LINE WITH 3ML OF BLOOD PER THE OPERATOR'S MANUAL. THE DONOR OBSERVED THE AIR BUBBLE AND QUESTIONED IF IT WAS SAFE AND WHY IT HAPPENED. TERUMO BCT CUSTOMER SUPPORT INFORMED THE CUSTOMER THAT A SMALL AMOUNT OF AIR IS OK BECAUSE THE BUBBLE WILL DISSIPATE AS IT ENTERS THE BLOOD STREAM. PATIENT AGE WAS NOT PROVIDED BY CUSTOMER. FULL PATIENT ID: (B)(6). PATIENT IS IN STABLE CONDITION AND WAS RELEASED TO SELF, NO COMPLICATIONS. THE COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17365099 ↗InjuryJun 14, 2023Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER REPORTED THAT DURING A THE FIRST DONATION BY APHERESIS THE PATIENT STARTED TO PRESENT REDNESS ON THE FACE, ARMS, SHOULDERS, AND NECK AND ALSO REPORTED ITCHING AND A SENSATION OF MILD HEAT. THE CUSTOMER REPORTED THAT 100MG OF HYDROCORTISONE WAS ADMINISTERED INTRAVENOUSLY AND 10MG LORATADINE ORALLY. THE EVENT WAS RESOLVED IN 1 HOUR AND THE DONOR IS ASYMPTOMATIC. THE DONOR IS IN STABLE CONDITION. THE CUSTOMER DECLINED TO PROVIDE THE PATIENT IDENTIFIER. THE COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: Itching Sensation; Rash
MAUDE report 17130676 ↗InjuryJun 14, 2023Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER REPORTED THAT 25 MINUTES AFTER STARTING APHERESIS THE DONOR BEGAN TO PRESENT REDNESS ON THE FACE, ARMS, SHOULDERS, AND NECK, AND REPORTED THAT THE DONOR WAS ALSO ITCHING AND HAD A SENSATION OF HEAT ALONG WITH MILD SHORTNESS OF BREATH AND TINNITUS. LORATADINE WAS ADMINISTERED ORALLY AND THE DONOR VOMITED AND THEN 100MG OF HYDROCORTISONE WAS ADMINISTERED INTRAVENOUSLY AND A 10MG DOSE OF LORATADINE WAS REPEATED ORALLY. THE CUSTOMER REPORTED THAT THE SYMPTOMS BEGAN TO SUBSIDE AFTER 5 MINUTES. THE CUSTOMER REPORTED AFTER 24 HOURS THE DONOR REPORTED A GENERALIZED SLIGHTLY ITCHY RASH AND IT WAS INDICATED THAT 10 MG OF CETIRIZINE WAS TAKEN. THE PATIENT IS IN STABLE CONDITION. THE CUSTOMER DECLINED TO PROVIDE THE CUSTOMER IDENTIFIER. THE COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: Itching Sensation; Rash; Respiratory Insufficiency; Tinnitus
MAUDE report 17130050 ↗MalfunctionJun 8, 2023Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE RED BLOOD CELL (RBC) PRODUCT. DONOR UNIT #: (B)(6) THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17088889 ↗MalfunctionJun 6, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE RED BLOOD CELL (RBC) PRODUCT. DONOR UNIT #: (B)(6) THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE PLATELET COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17072016 ↗MalfunctionJun 5, 2023Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE RED BLOOD CELL (RBC) PRODUCT. DONOR UNIT #: (B)(6) THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE PLATELET COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17067040 ↗MalfunctionMay 10, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE DOUBLE RED BLOOD CELL (RBC) PRODUCT. DONOR UNIT #: (B)(6) THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16904372 ↗MalfunctionMay 8, 2023Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE PLATELET PRODUCT. THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE WBC COUNT IS NOT AVAILABLE AT THIS TIME THE PLATELET COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16887850 ↗MalfunctionMar 28, 2023Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE PLATELET PRODUCT. DONOR UNIT #: (B)(6) THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE PLATELET COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16628710 ↗MalfunctionMar 14, 2023High Test Results; No Apparent Adverse Event; No Apparent Adverse Event
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE PLATELET PRODUCT. THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE PLATELET COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16538219 ↗MalfunctionMar 14, 2023Incorrect, Inadequate or Imprecise Result or Readings; No Apparent Adverse Event; No Apparent Adverse Event
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE PLATELET PRODUCT. THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE PLATELET COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16538182 ↗MalfunctionJan 6, 2023No Apparent Adverse Event; Insufficient Device Problem Information; No Apparent Adverse Event; Insufficient Device Problem Information
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE PLATELET PRODUCT. DONOR UNIT #: W28682208875200T THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE PLATELET COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16115989 ↗MalfunctionJan 3, 2023Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE PLATELET PRODUCT. DONOR UNIT#: (B)(6) THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBC TESTING, THEREFORE NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. THE PLATELET COLLECTION SET IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16091411 ↗
Other Separator, automated, blood cell, diagnostic devices
- Trima Accel® 82406Terumo Bct, Inc. · Separator, automated, blood cell, diagnostic470 reports
- Trima Accel® 82446Terumo Bct, Inc. · Separator, automated, blood cell, diagnostic1 death300 reports
- Trima Accel® 81000Terumo Bct, Inc. · Separator, automated, blood cell, diagnostic150 reports
- Spectra Optia® 12220Terumo Bct, Inc. · Separator, Automated, Blood Cell, Diagnostic14 deaths147 reports
- Spectra Optia® 12320Terumo Bct, Inc. · Separator, automated, blood cell, diagnostic2 deaths119 reports
- Trima Accel® 80400Terumo Bct, Inc. · Separator, automated, blood cell, diagnostic109 reports
- Spectra Optia® 10310Terumo Bct, Inc. · Separator, automated, blood cell, diagnostic6 deaths103 reports
- HAEMONETICS PCS 300 PLASMA COLLECTION SYSTEM PCS-300-USHaemonetics Corporation · Separator, automated, blood cell, diagnostic18 deaths101 reports