RayOne EMV RAO200E (REF RAO200E/+19.0)Recalls & reported problems

Rayner Intraocular Lenses Limited · Class 3 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
15
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 05029867006685), received between May 20, 2022 and Oct 24, 2025 — classified as 15 malfunctions.

The most frequently reported specific device problem is “Manufacturing, Packaging or Shipping Problem” (8 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 13, 2022Open, Classified

    The outer package is mislabeled and the package contains a different IOL strength.

    Z-1338-2022 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Manufacturing, Packaging or Shipping Problem8
  • Break2
  • Product Quality Problem2
  • Crack1
  • Material Opacification1
  • Optical Distortion1
  • Optical Problem1

Patient problems

  • Visual Impairment11
  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 24, 2025Crack; Crack

    ON 2ND OCTOBER 2025, RAYNER RECEIVED NOTIFICATION FROM ITS TAIWAN DISTRIBUTOR OF AN EVENT THAT OCCURRED DURING USE OF A RAYONE EMV RAO200E. THE EVENT DESCRIPTION PROVIDED STATES THAT IMMEDIATELY FOLLOWING IMPLANTATION INTO THE EYE A CRACK WAS OBSERVED IN THE OPTIC NECESSITATING LENS EXPLANTATION AND EXCHANGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23377144 ↗
  • MalfunctionMay 28, 2025Optical Problem; Optical Problem

    ON 15TH MAY 2025, RAYNER RECEIVED NOTIFICATION FROM A US HEALTHCARE FACILITY OF AN EVENT THAT OCCURRED DURING USE OF A RAYONE EMV RAO200E. THE EVENT DESCRIPTION PROVIDED STATES THAT THE PATIENT REFRACTIVE OUTCOME WAS NOT AS EXPECTED POST-OPERATIVELY NECESSITATING AN ADDITIONAL SURGERY TO EXPLANT AND EXCHANGE THE IOL.

    Patient problems: Visual Impairment

    MAUDE report 22094781 ↗
  • MalfunctionJan 22, 2025Break; Break

    ON 13TH JANUARY 2025, RAYNER RECEIVED NOTIFICATION FROM A US HEALTHCARE FACILITY OF AN EVENT THAT OCCURRED DURING USE OF A RAYONE EMV RAO200E. THE EVENT DESCRIPTION PROVIDED STATES THAT THE TRAILING HAPTIC BROKE DURING IMPLANTATION NECESSITATING EXPLANTATION OF THE IOL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21206730 ↗
  • MalfunctionDec 11, 2024Optical Distortion; Optical Distortion

    ON 5TH DECEMBER 2024, RAYNER RECEIVED NOTIFICATION FROM A US HEALTHCARE PROFESSIONAL OF AN EVENT THAT OCCURRED FOLLOWING IMPLANTATION OF A RAYONE EMV RAO200E. THE EVENT DESCRIPTION PROVIDED STATES THAT THE PATIENT'S POST-OPERATIVE REFRACTIVE OUTCOME WAS NOT AS EXPECTED.

    Patient problems: Visual Impairment

    MAUDE report 20901515 ↗
  • MalfunctionMar 3, 2024Material Opacification; Material Opacification

    ON 13TH FEBRUARY 2024, RAYNER RECEIVED NOTIFICATION FROM A US HEALTHCARE FACILITY OF AN EVENT THAT OCCURRED FOLLOWING IMPLANTATION OF A RAYONE EMV RAO200E. THE EVENT DESCRIPTION PROVIDED STATES THAT POST-OPERATIVELY OPACIFICATION OF THE IOL WAS OBSERVED AND THAT VISUAL ACUITY DECLINED AS A RESULT NECESSITATING IOL EXPLANTATION.

    Patient problems: Visual Impairment

    MAUDE report 18823456 ↗
  • MalfunctionOct 3, 2022Break; Break

    ON (B)(6) 2022, RAYNER RECEIVED NOTIFICATION FROM A US HEALTHCARE FACILTIY OF AN EVENT THAT OCCURRED DURING USE OF A RAYONE EMV RAO200E. THE EVENT DESCRIPTION PROVIDED STATES THAT THE LENS HAPTIC WAS STUCK IN THE INSERTER AND TORE AFTER ATTEMPT TO PLACE IT IN THE EYE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15527857 ↗
  • MalfunctionMay 25, 2022Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    ON (B)(6) 2022, RAYNER INTRAOCULAR LENSES LIMITED RECEIVED NOTIFICATION FROM A US HEALTHCARE FACILITY THAT A LENS SUBJECT TO RECALL (REF. (B)(4)/17/2022-001-R) HAD BEEN IMPLANTED. POST-OPERATIVELY, THE PATIENT HAS PRESENTED WITH A REFRACTIVE ERROR.

    Patient problems: Visual Impairment

    MAUDE report 14495398 ↗
  • MalfunctionMay 25, 2022Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    ON 19TH MAY 2022, RAYNER INTRAOCULAR LENSES LIMITED RECEIVED NOTIFICATION FROM A US HEALTHCARE FACILITY THAT A LENS SUBJECT TO RECALL (REF. 3012304651-05/17/2022-001-R) HAD BEEN IMPLANTED. POST-OPERATIVELY, THE PATIENT HAS PRESENTED WITH A REFRACTIVE ERROR.

    Patient problems: Visual Impairment

    MAUDE report 14495351 ↗
  • MalfunctionMay 25, 2022Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    ON (B)(6) 2022, RAYNER INTRAOCULAR LENSES LIMITED RECEIVED NOTIFICATION FROM A US HEALTHCARE FACILITY THAT A LENS SUBJECT TO RECALL (REF. 3012304651-05/17/2022-001-R) HAD BEEN IMPLANTED. POST-OPERATIVELY, THE PATIENT HAS PRESENTED WITH A REFRACTIVE ERROR.

    Patient problems: Visual Impairment

    MAUDE report 14495023 ↗
  • MalfunctionMay 25, 2022Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    ON 19TH MAY 2022, RAYNER INTRAOCULAR LENSES LIMITED RECEIVED NOTIFICATION FROM A US HEALTHCARE FACILITY THAT A LENS SUBJECT TO RECALL (REF. 3012304651-05/17/2022-001-R) HAD BEEN IMPLANTED. POST-OPERATIVELY, THE PATIENT HAS PRESENTED WITH A REFRACTIVE ERROR.

    Patient problems: Visual Impairment

    MAUDE report 14495000 ↗
  • MalfunctionMay 25, 2022Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    ON (B)(6) 2022, RAYNER INTRAOCULAR LENSES LIMITED RECEIVED NOTIFICATION FROM A US HEALTHCARE FACILITY THAT A LENS SUBJECT TO RECALL (REF. 3012304651-05/17/2022-001-R) HAD BEEN IMPLANTED. POST-OPERATIVELY, THE PATIENT HAS PRESENTED WITH A REFRACTIVE ERROR.

    Patient problems: Visual Impairment

    MAUDE report 14494992 ↗
  • MalfunctionMay 25, 2022Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    ON 19TH MAY 2022, RAYNER INTRAOCULAR LENSES LIMITED RECEIVED NOTIFICATION FROM A US HEALTHCARE FACILITY THAT A LENS SUBJECT TO RECALL (REF. 3012304651-05/17/2022-001-R) HAD BEEN IMPLANTED. POST-OPERATIVELY, THE PATIENT HAS PRESENTED WITH A REFRACTIVE ERROR.

    Patient problems: Visual Impairment

    MAUDE report 14494991 ↗
  • MalfunctionMay 25, 2022Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    ON (B)(6) 2022, RAYNER INTRAOCULAR LENSES LIMITED RECEIVED NOTIFICATION FROM A US HEALTHCARE FACILITY THAT A LENS SUBJECT TO RECALL (REF. (B)(4)/17/2022-001-R) HAD BEEN IMPLANTED. POST-OPERATIVELY, THE PATIENT HAS PRESENTED WITH A REFRACTIVE ERROR.

    Patient problems: Visual Impairment

    MAUDE report 14494730 ↗
  • MalfunctionMay 20, 2022Product Quality Problem; Product Quality Problem

    ON 16TH MAY 2022, RAYNER INTRAOCULAR LENSES LIMITED RECEIVED NOTIFICATION FROM A US HEALTHCARE FACILITY THAT A LENS SUBJECT TO RECALL (REF. (B)(4) HAD BEEN IMPLANTED. POST-OPERATIVELY, THE PATIENT HAS PRESENTED WITH A REFRACTIVE ERROR.

    Patient problems: Visual Impairment

    MAUDE report 14461440 ↗
  • MalfunctionMay 20, 2022Product Quality Problem; Manufacturing, Packaging or Shipping Problem; Product Quality Problem; Manufacturing, Packaging or Shipping Problem

    ON (B)(6) 2022, RAYNER INTRAOCULAR LENSES LIMITED RECEIVED NOTIFICATION FROM ITS US SALES TEAM THAT A RAYONE EMV RAO200E +19.0D BATCH 012183745 SYSTEM PACK CONTAINED THE INNER BLISTER TRAY (PRELOADED INJECTOR AND IOL) FOR THE RAYONE EMV RAO200E +21.0 D BATCH 012183845. THE PRODUCT AS SUBJECT TO THIS REPORT WAS NOT USED; HOWEVER, IT WAS IDENTIFIED WITHIN THE INITIAL NOTIFICATION THAT PRIOR IMPLANTATION OF A LENS FROM THE SAME BATCH HAD OCCURRED. INVESTIGATION BY RAYNER IDENTIFIED THAT THE ISSUE AFFECTED ALL UNITS IN THE RAYONE EMV RAO200E +19.0D BATCH 012183745 AND RAYONE EMV RAO200E +21.0 D BATCH 012183845. THE ACUTE CONSEQUENCE OF THE HEALTH HAZARD TO THE PATIENT IS A REFRACTIVE SURPRISE; A DEVIATION IN TARGET REFRACTION THAT COULD BE NOTICED BY THE PATIENT FOLLOWING SURGERY (MANIFESTING IN THIS CASE AS BLURRED DISTANCE VISION DUE TO THE MYOPIC REFRACTIVE ERROR) WHICH WOULD BE CONFIRMED WHEN THE PATIENT'S SURGICAL/OPTICAL TEAM DETERMINES THE PATIENT'S REFRACTION. THE MOST PROBABLE RESULT FOR PATIENTS WHO HAVE UNDERGONE IMPLANTATION OF A +21.0 D EMV LENS AS OPPOSED TO THE REQUIRED +19.0 D IS A 2.0 D MYOPIC OUTCOME. THE RAYONE EMV RAO200E +19.0D BATCH 012183745 IS SUBJECT TO A VOLUN

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14461037 ↗