Swann Morton Ltd. No. 15 CARBON STERILE BLADESRecalls & reported problems
Swann-Morton Limited · Blade,scalpel, single use. A component of a surgical instrument (scalpel) that is designed to be attached to the scalpel handle. It is capable of cutting through tissue when moved with downward pressure. This device is disposable. · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 05033955002053), received between Sep 14, 2023 and Jul 7, 2025 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information7
- Fracture5
- Break2
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Insufficient Information3
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 7, 2025Break; Break
THE FOLLOWING DESCRIPTION WAS PROVIDED BY THE HEALTHCARE FACILITY. "THE RESIDENT WAS PERFORMING SURGERY ON THE PATIENTS LEFT FOOT (PODIATRY CASE) WHEN THE KNIFE BLADE BROKE IN HALF. SHE WAS ABLE TO TAKE THE PIECE OUT IMMEDIATELY". THE FOLLOWING REMEDIAL ACTIONS WERE STATED BY THE HEALTHCARE FACILITY. "SHE WAS ABLE TO TAKE THE PIECE OUT IMMEDIATELY. XRAY WILL BE TAKEN AT THE END OF THE CASE DUE TO THE NATURE OF THE CASE".
Patient problems: Insufficient Information
MAUDE report 22411340 ↗MalfunctionJun 17, 2025Break; Break
THE FOLLOWING DESCRIPTION WAS PROVIDED BY THE HEALTHCARE FACILITY. "TODAY WE HAD A 15-BLADE SNAP IN HALF DURING A SPINAL CASE". IT WAS STATED NO PERSON WAS INJURED DUE TO THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22234711 ↗MalfunctionMay 1, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
THE FOLLOWING DESCRIPTION WAS PROVIDED BY THE HEALTHCARE FACILITY. "TODAY WE HAD A 15-BLADE SNAP IN HALF DURING A SPINAL CASE. THE SURGEON SAID HE HAS NEVER HAD THAT HAPPEN BEFORE". IT WAS STATED BY THE HEALTHCARE FACILITY THAT. "TO ASSURE THAT ALL THE PIECES OF THE BLADE WERE REMOVED FROM THE PATIENT, THE PATIENT UNDERWENT X-RAY IMAGING THAT WOULD OTHERWISE NOT HAVE BEEN EXPOSED TO, THIS DELAYED THE CASE FOR AN EXTRA 5-8 MINUTES".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21941051 ↗MalfunctionJan 15, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
THE FOLLOWING DESCRIPTION WAS PROVIDED BY HEALTHCARE FACILITY. "A #15 KNIFE BLADE IN ONE OF OUR PACKS BROKE IN HALF WHEN A SURGEON MADE HIS INCISION. THE SURGERY WAS A NASAL PROCEDURE, AND WHEN THE BLADE BROKE IN HALF IT CAUSED AN INJURY TO THE PATIENTS LIP".
Patient problems: Insufficient Information
MAUDE report 21159631 ↗MalfunctionSep 2, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
THE FOLLOWING DESCRIPTION WAS PROVIDED BY THE HEALTHCARE FACILITY, "COMPONENT BROKE WHEN SURGEON WENT TO MAKE AN INCISION ON THE PATIENT".
Patient problems: Insufficient Information
MAUDE report 20117022 ↗MalfunctionAug 21, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
THE FOLLOWING DESCRIPTION WAS PROVIDED BY THE HEALTHCARE FACILITY, "BLADE BROKE OFF IN THE PATIENT HAS SOME OF IT WAS NEVER FOUND" "MULTIPLE HAND PACKS WITH 15 BLADES HAVE BEEN BROKE MULTIPLE TIMES JUST USING A NEEDLE HOLDER TO PLACE IT ON A KNIFE HANDLE. BLADES ARE BREAKING OFF IN PREPARATION FOR AND DURING SURGERIES." ALTHOUGH IT WAS STATED THAT PART OF THE BLADE BROKE IN THE PATIENT AND WAS NEVER FOUND, IT WAS ALSO STATED THAT NO INJURY OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20012290 ↗MalfunctionAug 20, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
THE FOLLOWING DESCRIPTION WAS PROVIDED BY THE HEALTHCARE FACILITY, "HAND PACKS WITH 15 BLADES HAVE BEEN BROKE MULTIPLE TIMES JUST USING A NEEDLE HOLDER TO PLACE IT ON A KNIFE HANDLE. BLADES ARE BREAKING OFF IN PREPARATION FOR AND DURING SURGERIES. IT WAS REMOVED FROM THE STERILE FIELD. NOT IN THE PATIENT. THIS IS HAPPENING WHILE THE SCRUB TECHS ARE PLACING THE BLADES ON KNIFE HANDLES."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20007100 ↗MalfunctionMay 20, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
THE FOLLOWING DESCRIPTION OF THE INCIDENT WAS PROVIDED BY THE HEALTHCARE FACILITY, "BLADE #15 WAS BROKEN DURING USE. NO PATIENT OR STAFF INJURY. NO OTHER DETAILS AND HAPPENED ONCE ONLY". NO REMEDIAL ACTIONS WERE STATED BY THE HEALTHCARE FACILITY, ALTHOUGH IT WAS STATED THAT THERE WAS "NO PATIENT INJURY OR STAFF INJURY", AND THAT IT "ONLY HAPPENED ONCE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19355304 ↗MalfunctionMar 5, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
PLEASE FIND THE FOLLOWING DESCRIPTION PROVIDED BY THE HEALTHCARE FACILITY, "3 TIMES THIS WEEK (3 DIFFERENT CASES) WHEN USING THE 15 BLADES THAT CAME IN THE PACK, THE BLADES HAVE BROKEN". IT WAS STATED THAT THERE WAS NO INJURY OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18835302 ↗MalfunctionJan 9, 2024Fracture; Fracture
PLEASE SEE THE DESCRIPTION PROVIDED BY THE HEALTHCARE FACILITY"#15 BLADE BLADE BROKE OFF WHILE IN USE AND EXTENDED SURGICAL TIME WITH DIFFICULTY IN RETRIEVAL OF THE PIECE BROKEN OFF IN PATIENT WOUND. PHYSICIAN STATED HE UTILIZED MINIMAL ACTION IN BLADE MANIPULATION BEFORE THE BREAK. THE PATIENT HAD NO POST-OP COMPLICATIONS AND IS DOING WELL".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18474147 ↗MalfunctionJan 9, 2024Fracture; Fracture
PLEASE SEE DESCRIPTION PROVIDED BY THE HEALTHCARE FACILITY. "BLADE BREAKING DURING USE ON AN ANKLE CASE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18474136 ↗MalfunctionOct 6, 2023Fracture; Fracture
THE DESCRIPTION PROVIDED BY THE HEALTHCARE FACILITY WAS AS FOLLOWS: "BLADE BROKE DURING USE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17884876 ↗MalfunctionSep 14, 2023Fracture; Fracture
PLEASE SEE THE DESCRIPTION PROVIDED BY THE HEALTHCARE FACILITY, "BLADE BROKE IN HALF - PHYSICIAN WAS USING TO CREATE AN INCISION, PROCEDURE NOT SPECIFIED".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17750691 ↗MalfunctionSep 14, 2023Fracture; Fracture
PLEASE SEE THE DESCRIPTION PROVIDED BY THE HEALTHCARE FACILITY, "BLADE BROKE IN HALF, PROCEDURE NOT SPECIFIED".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17750690 ↗
Other Blade, Scalpel devices
- Medline CISION15CSMedline Industries, Inc. · BLADE, SCALPEL1 recall2 reports
- Medline CISION10CSMedline Industries, Inc. · BLADE, SCALPEL1 recall1 report
- Medline CISION15CSRMedline Industries, Inc. · BLADE, SCALPEL1 recall1 report
- Medline CISION12SSMedline Industries, Inc. · BLADE, SCALPEL1 recall0 reports
- Medline CISION10CSRMedline Industries, Inc. · BLADE, SCALPEL1 recall0 reports
- Medline CISION21CSMedline Industries, Inc. · BLADE, SCALPEL1 recall0 reports
- SPSS Cartridges 73-8020Southmedic Incorporated · BLADE, SCALPEL1 recall0 reports
- Medline CISION11SSMedline Industries, Inc. · BLADE, SCALPEL1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page