Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 05037881027043), received between Aug 28, 2024 and Jun 27, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Insufficient Device Problem Information1
  • Material Integrity Problem1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding1
  • Insufficient Information1
  • Reduced Blood Flow1
  • Rupture1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 27, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    EVENT NAME: AORTIC PSEUDOCOARCTATION. EXTEND-001 (B)(6), PATIENT NO. (B)(6) ON POST-OPERATIVE DAY 2 (B)(6) 2025), AORTIC PSEUDOCOARCTATION AT THE JUNCTION BETWEEN THE STENT GRAFT FOR THE THORAFLEX AND THE DACRON COMPONENT NOTED ON CHEST COMPUTED TOMOGRAPHY SCAN. TREATMENT OF THE EVENT WAS SURGICAL PATIENT UNDERWENT BALLOON ANGIOPLASTY ON (B)(6) 2025. THE SERIOUS ADVERSE EVENT (SAE) WAS CONSIDERED SEVERE. THE SAE IS RESOLVED WITH TREATMENT, AND RESOLVED WITHOUT SEQUELAE. SUBJECT CONTINUES IN THE STUDY.

    Patient problems: Insufficient Information; Reduced Blood Flow

    MAUDE report 22343903 ↗
  • MalfunctionMar 12, 2025Material Integrity Problem; Material Integrity Problem

    LEAKAGE FROM A THORAFLEX HYBRID ANTE-FLO FROM THE SUTURE LINE CONNECTING THE CUFF TO THE MAIN BODY - GREEN SUTURE LINE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 21582557 ↗
  • InjuryAug 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CLINICAL TRIAL: EXTEND-001 (B)(4). THORAFLEX HYBRID WAS IMPLANTED ON (B)(6) 2023. NO DISTAL SEAL WAS INTENDED DUE TO A PLANNED EVAR (UNDERSTOOD TO BE ENDOVASCULAR AORTIC REPAIR) EXTENSION AND NO DISTAL SEAL WAS OBTAINED. NO ENDOLEAK WAS PRESENT AT THE END OF THE PROCEDURE. DELIVERY OF THE THORAFLEX HYBRID WAS SUCCESSFUL, DEPLOYED IN PLANNED LOCATION AND WITHDRAWAL OF DELIVERY SYSTEM WAS SUCCESSFUL. NO DEVICE DEFICIENCIES OCCURRED DURING THE INDEX TH (UNDERSTOOD TO BE THORAFLEX HYBRID) PROCEDURE. EVENT DETAILS - RUPTURE (ONSET DATE 17-JUL-24, RESOLVED ON (B)(6) 2024). CONTAINED RUPTURE OF THE DESCENDING THORACIC AORTA DISTAL TO PREVIOUS ENDOVASCULAR AORTIC REPAIR. THE CLINICIAN/SITE HAVE REPORTED THIS EVENT AS NOT RELATED TO PROCEDURE AND NOT RELATED TO DEVICE. DURING THEIR REGULAR REVIEW OF EXTEND ADVERSE EVENTS, THE STUDY MEDICAL MONITOR HAS REVIEWED AND MARKED THIS EVENT AS POSSIBLY RELATED TO THE DEVICE. SURGICAL PROCEDURE CARRIED OUT - TEVAR ON (B)(6) 2024. TERUMO CATALOGUE NUMBER 2312050409. OUTCOME OF EVENT HAS BEEN RECORDED AS RESOLVED WITH TREATMENT.

    Patient problems: Rupture

    MAUDE report 20083291 ↗

Other Hybrid stent graft, thoracic aortic lesion treatment devices

All Hybrid stent graft, thoracic aortic lesion treatment devices →

Sources and updates