SENSAR IOL AR40e (REF AR40E00225)Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · AMO Sensar AR40e Lens · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Sensar AR40e intraocular lens is implanted in the eye during cataract surgery to replace the eye's natural lens. The 36 reports mostly describe the tiny supporting arms (haptics) breaking, cracking or tearing during implantation, sometimes leaving pieces in the eye and requiring the lens to be removed and replaced; 14 reports describe injuries and 22 describe malfunctions. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 35
- Deaths reported
- 0
- Injuries reported
- 14
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 05050474502154), received between Mar 15, 2017 and May 8, 2026 — classified as 14 injuries, 21 malfunctions.
The most frequently reported specific device problem is “Break” (16 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break16
- Adverse Event Without Identified Device or Use Problem6
- Material Fragmentation5
- Device Dislodged or Dislocated2
- Contamination1
- Difficult to Insert1
- Inaccurate Delivery1
- Manufacturing, Packaging or Shipping Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions18
- No Code Available4
- No Known Impact Or Consequence To Patient4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
- Vitrectomy3
- Capsular Bag Tear2
- Conjunctivitis1
- Excessive Tear Production1
Recent adverse event reports
The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 8, 2026Break
IT WAS REPORTED THAT A NON-PRELOADED MONOFOCAL INTRAOCULAR LENS HAPTIC BROKE DURING INSERTION AND WAS NOTICED AFTER THE LENS HAD BEEN IMPLANTED; CONSEQUENTLY, THE LENS WAS REMOVED DURING THE SAME PROCEDURE. DIRECTIONS FOR USE HAD BEEN FOLLOWED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25126674 ↗MalfunctionFeb 6, 2026Break
IT WAS REPORTED THAT THE STANDALONE INTRAOCULAR LENS (IOL) IS DEFECTIVE. NO FURTHER INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24284934 ↗MalfunctionFeb 2, 2026Break
IT WAS REPORTED THAT A LENS HAPTIC FRACTURED AFTER IMPLANTATION. THE SURGEON REMOVED THE LENS AND IMPLANTED A REPLACEMENT. THE PATIENT RECOVERED WELL. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24242648 ↗MalfunctionOct 30, 2025Break; Break
IT WAS REPORTED THAT THE JOHNSON AND JOHNSON (JNJ) INTRAOCULAR LENS (IOL) HAD A BROKEN HAPTIC. IT WAS EXPLAINED THAT THE TRAILING HAPTIC BROKE DURING IMPLANTATION. THE IOL WAS FULLY INSERTED. IT WAS CUT AND REMOVED. A VITRECTOMY WAS PERFORMED. THE PROCEDURE WAS COMPLETED WITH A BACKUP IOL. THE DIRECTIONS FOR USE WERE FOLLOWED. THERE WAS NO SECONDARY SURGERY OR UNPLANNED SUTURES. THE PATIENT HAS FULLY RECOVERED. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23432701 ↗MalfunctionSep 5, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE SUPERIOR LOOP OF A NON-PRELOADED MONOFOCAL INTRAOCULAR LENS (IOL) BECAME DETACHED DURING THE INSERTION PROCESS. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22984207 ↗InjuryAug 13, 2025Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT DURING A PROCEDURE THE INTRAOCULAR LENS WAS IMPLANTED BUT THE LENS DID NOT SEAT AS INTENDED, AND UPON REMOVAL, IT CAUSED THE PATIENT'S ENTIRE IRIS TO BE EXTRACTED. NO REPLACEMENT LENS WAS IMPLANTED. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Eye Injury
MAUDE report 22790815 ↗MalfunctionJun 13, 2025Break; Break
IT WAS REPORTED THAT THE INTRAOCULAR LENS (IOL) WAS BROKEN. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22207958 ↗InjuryMay 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CASE INVOLVING THE AR40E MODEL INTRAOCULAR LENS (IOL) TURNED INTO A COMPLICATED ANTERIOR VITRECTOMY DUE TO THE PATIENT'S ANATOMY. THE DOCTOR HAD TO USE FOUR (04) LENSES AND THREE (03) WERE DESTROYED IN THE SURGICAL PROCESS. THE SURGERY WAS COMPLETED DURING THE SAME PROCEDURE. PATIENT PROGNOSIS POST PROCEDURE IS UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 22111030 ↗InjuryMay 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CASE INVOLVING THE AR40E MODEL INTRAOCULAR LENS (IOL) TURNED INTO A COMPLICATED ANTERIOR VITRECTOMY DUE TO THE PATIENT'S ANATOMY. THE DOCTOR HAD TO USE FOUR (04) LENSES AND THREE (03) WERE DESTROYED IN THE SURGICAL PROCESS. THE SURGERY WAS COMPLETED DURING THE SAME PROCEDURE. PATIENT PROGNOSIS POST PROCEDURE IS UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 22110960 ↗MalfunctionMar 18, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE HAPTIC OF THE NON-PRELOADED INTRAOCULAR LENS (IOL) WAS DETACHED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21635632 ↗MalfunctionNov 12, 2024Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem
IT WAS REPORTED THAT THE THREE PIECE MONOFOCAL NON PRELOADED INTRAOCULAR LENS (IOL) WAS RECEIVED WITH BROKEN SEAL. THERE WAS NO CONTACT WITH PATIENT'S EYE. AS PER ADDITIONAL INFORMATION RECEIVED, IT IS UNKNOWN WHETHER THERE WAS ANY BREACH OR POTENTIAL BREACH IN STERILITY OF THE IOL. ONLY ONE BOX WAS INVOLVED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20668273 ↗MalfunctionApr 10, 2024Scratched Material; Scratched Material
IT WAS REPORTED THAT A MONOFOCAL INTRAOCULAR LENS WAS FULLY INSERTED AND REMOVED IN THE SAME PROCEDURE DUE TO A DENT IN THE LENS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING DIFFERENT MODEL MONOFOCAL IOL. THERE WAS NO INJURY, AND SURGICAL AND MEDICAL INTERVENTIONS WERE REQUIRED. THE PATIENT IS DOING WELL. THE PRODUCT IS BEING RETUNED. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19077109 ↗MalfunctionDec 27, 2023Break; Break
IT WAS REPORTED THAT THE HAPTIC OF THE NON PRELOADED MONOFOCAL INTRAOCULAR LENS(IOL) WAS KINKED, WHEN IT WAS IMPLANTED. ADDITIONAL INFORMATION INDICATED THAT THE LENS WAS FULLY INSERTED AND THE HAPTIC WAS BROKEN BUT NOT DETACHED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING BACKUP LENS. THERE WAS NO USE ERROR AND THERE WAS NO PATIENT INJURY. THERE WAS NO MEDICAL/SURGICAL INTERVENTION REPORTED. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18399847 ↗MalfunctionDec 13, 2023Break; Break
IT WAS REPORTED THERE WAS HAPTIC BENT. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18321455 ↗MalfunctionOct 3, 2023Break; Break
IT WAS REPORTED THAT THE PATIENT WAS UNDERGOING A PLANNED SCLERAL FIXATED SECONDARY INTRAOCULAR LENS (IOL) IMPLANTATION PROCEDURE IN THEIR LEFT EYE. WHEN THE IOL WAS INTRODUCED INTO THE ANTERIOR CHAMBER OF THE PATIENT'S EYE, THE TRAILING HAPTIC BROKE. THE IOL WAS REMOVED AND REPLACED WITH A NON-JOHNSON&JOHNSON LENS. THERE WERE MEDICATIONS PRESCRIBED AS PART OF STANDARD OF CARE, AND PLANNED SUTURES WERE PLACED. ACCOUNT INDICATED THAT THE PATIENT WAS WELL POST-OPERATIVELY AND VISION WAS OKAY. THERE WAS NO DELAY IN TREATMENT OR OTHER INTERVENTIONS PERFORMED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17864749 ↗MalfunctionAug 17, 2023Break; Break
THE CUSTOMER REPORTED OPEN, WASTED, LENS DAMAGE WITH AN INTRAOCULAR LENS (IOL) FOR THE LEFT EYE OF A PATIENT. IT IS UNKNOWN IF THERE WAS PATIENT CONTACT. IT WAS INDICATED THAT THE ISSUE IS NOT RELATED TO USE ERROR. THERE WERE NO MEDICAL AND OR SURGICAL INTERVENTIONS REQUIRED. THE IOL IS NOT RETURNING AS IT WAS DISCARDED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17569528 ↗InjuryAug 31, 2022Material Fragmentation; Device Dislodged or Dislocated; Material Fragmentation; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE INTRAOCULAR LENS (IOL) LEADING HAPTIC WAS BROKEN AND WAS NOT VISIBLE ON INSERTION IN THE PATIENT¿S RIGHT EYE, IOL WAS DEFECTIVE. UPON FURTHER FOLLOW-UP, WE LEARNED THAT THE IOL WAS FULLY INSERTED. ACCOUNT INDICATED THAT THE LENS DROPPED TO THE BACK OF THE EYE RESULTING IN VITRECTOMY. THE LENS WAS REMOVED AND REPLACED IN THE SAME PROCEDURE DURING WHICH A COMPETITOR LENS WAS IMPLANTED. SUTURES WERE PLACED. NO TREATMENT OR PRESCRIPTION WAS GIVEN BY THE DOCTOR OUTSIDE THE REGULAR STANDARD OF CARE. THE PATIENT IS FINE PRESENTLY AND IS BEING MONITORED. THE LENS IS AVAILABLE FOR EVALUATION. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 15325723 ↗MalfunctionJul 18, 2022Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING THE IMPLANTATION OF THE INTRAOCULAR LENS (IOL), THE HAPTIC BROKE OFF. THE IOL WAS REMOVED DURING THE SAME PROCEDURE AND ANOTHER LENS OF THE SAME MODEL AND DIOPTER WAS IMPLANTED. THE PATIENT IS DOING WELL WITH NO FURTHER COMPLICATIONS. NO FURTHER DETAILS WERE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15047994 ↗MalfunctionMay 4, 2022Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT UPON FURTHER CLARIFICATION, AN INTRAOCULAR LENS (IOL) WAS REMOVED AND REPLACED DUE TO THE AMPUTATION OF THE HAPTIC. IT WAS NOT DURING ASSEMBLY, IT WAS MOUNTED, AND AT THE TIME OF INSERTING THE LENS THE AMPUTATION WAS VERIFIED. NO FURTHER INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14274509 ↗InjuryApr 5, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT HAD INTRAOCULAR LENS (IOL) IMPLANT IN THE RIGHT EYE. TWO WEEKS LATER, THE PATIENT RETURNED DUE TO REDNESS AND LACRIMATION (TEARING) IN THE RIGHT EYE. EXAMINATION SHOWED MILD CONJUNCTIVAL CONGESTION IN THE RIGHT EYE, WEAKLY POSITIVE CILIARY TENDERNESS, FLOCCULENT EXUDATES IN FRONT OF THE ANTERIOR CHAMBER INTRAOCULAR LENS, AND SLUGGISH LIGHT REFLEX. MEDICATION WAS PRESCRIBED. ATROPINE EYE GEL EYE DROPS THREE TIMES A DAY; SUBCONJUNCTIVAL INJECTION OF DEXAMETHASONE INJECTION 2.5 MG ONCE; TOBRAMYCIN DEXAMETHASONE EYE DROPS THREE TIMES A DAY; TOBRAMYCIN DEXAMETHASONE EYE OINTMENT ONCE EVERY NIGHT. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Conjunctivitis; Excessive Tear Production; Red Eye(s)
MAUDE report 14018360 ↗InjuryJan 5, 2022Break; Break
IT WAS REPORTED THAT THE HAPTIC OF THE INTRAOCULAR LENS (IOL) WAS BROKEN DURING IMPLANTATION OF THE LENS INTO PATIENT'S OPERATIVE EYE. THE INCISION WAS ENLARGED AND THE LENS WAS REMOVED AND REPLACED DURING THE SAME PROCEDURE. ADDITIONALLY, SUTURES WERE REQUIRED AND VITRECTOMY WAS PERFORMED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13163046 ↗InjurySep 15, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE INTRAOCULAR LENS (IOL), MODEL AR40E, WAS REMOVED AND REPLACED FROM THE PATIENT'S LEFT EYE DURING THE SAME PROCEDURE DUE TO CAPSULE TEAR. A COMPETITOR LENS WAS IMPLANTED AS THE REPLACEMENT. ACCOUNT PROVIDED THAT THE OUTCOME DID NOT SIGNIFICANTLY INTERFERE WITH THE PATIENT'S DAILY LIFE ACTIVITIES. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Capsular Bag Tear
MAUDE report 12478551 ↗MalfunctionAug 13, 2021Break; Break
IT WAS REPORTED THAT DOCTOR DIDN'T LIKE THE LOOK OF A AR40E MODEL INTRAOCULAR LENS(IOL). NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12321244 ↗InjuryJul 6, 2021Break; Inaccurate Delivery; Break; Inaccurate Delivery
CUSTOMER PARTIALLY INSERTED AN INTRAOCULAR LENS (IOL) ON PATIENT'S RIGHT EYE (OD) WHEN THE PHYSICIAN NOTICED ONE OF ITS HAPTICS WAS BROKEN AND COULD NOT BE FULLY IMPLANTED. LENS WAS REMOVED ON THE SAME PROCEDURE. MEDICATION WAS PRESCRIBED. PATIENT WAS TEMPORARILY INJURED AND DAILY ACTIVITIES WAS AFFECTED. THE PATIENT HAD ALREADY USED THE PRODUCT BEFORE. THE PATIENT NEEDED TO RETURN THE DAY AFTER TO PUT IN ANOTHER LENS. THE CUSTOMER UNDERSTAND PATIENT HAD HIS ACTIVITIES AFFECTED. NOTHING WITH A HIGH LEVEL OF CONCERN, OR THAT COULD BRING ANY DISTRESS. THE MEDICATION PRESCRIBED WAS PART OF THE POST OPERATIVE ROUTINE. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12124787 ↗InjuryMay 7, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT LENS WAS NOT IMPLANTED FOR INTRAOPERATIVE COMPLICATION. AFTER FOLLOW-UP WE LEARNED THAT DURING THE SURGERY THERE WAS A POSTERIOR CAPSULAR RUPTURE. IT WAS STATED THAT THE LENS DID NOT GUARANTEE STABILITY AS IT MOVED IN THE VITREOUS SO FOR THAT REASON IT WAS REMOVED. NO PATIENT IMPACT REPORTED. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Capsular Bag Tear
MAUDE report 11791791 ↗
Other intraocular lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page