Adverse event reports
12
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 05050474502208), received between Jul 19, 2017 and Dec 11, 2025 — classified as 3 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Break” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break10
  • Adverse Event Without Identified Device or Use Problem2
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Code Available3
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 11, 2025Break; Break

    IT WAS REPORTED THAT THE JOHNSON AND JOHNSON (JNJ) INTRAOCULAR LENS (IOL) WAS INSERTED INTO THE PATIENTS EYE AND REMOVED DURING THE SAME PROCEDURE. THE NURSE WAS INFORMED THAT THE PRODUCT MADE CONTACT WITH THE PATIENT. HOWEVER, THE IOL WAS DAMAGED. THE LENS WAS REPLACED WITH A BACKUP OF THE SAME MODEL AND DIOPTER. NO FURTHER INFORMATION IS AVAILABLE REGARDING THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23785314 ↗
  • MalfunctionJun 11, 2025Break; Use of Device Problem; Break; Use of Device Problem

    THE SURGERY CENTER REPORTED BACK HAPTIC ALL TWISTED UP COMING OUT WITH A MONOFOCAL INTRAOCULAR LENS (IOL). THEY ALSO REPORTED LOADING ERROR. THERE WAS PATIENT CONTACT AND THE LENS WAS PLACED IN THE EYE. THE ISSUE WAS NOTED AFTER IMPLANTATION OF THE LENS. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22189339 ↗
  • MalfunctionApr 29, 2025Break; Break

    IT WAS REPORTED THAT A PRELOADED INTRAOCULAR LENS (IOL) HAS FAILED TO BE IMPLANTED AS ONE OF THE HAPTIC WAS BROKEN. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21931517 ↗
  • MalfunctionMar 18, 2022Break; Break

    IT WAS REPORTED THE INTRAOCULAR LENS (IOL) HAPTIC WAS BROKEN/TORN; THE EXTENT OF PATIENT CONTACT IS UNKNOWN. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13804116 ↗
  • MalfunctionJan 18, 2022Break; Break

    IT WAS REPORTED THAT AFTER INTRAOCULAR LENS (IOL) WAS IMPLANTED INTO THE EYE, THE HAPTIC WAS FRACTURED. IOL WAS REMOVED AND REPLACED. THE SURGERY WENT SMOOTHLY. PRODUCT RETURN NOT AVAILABLE. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13286847 ↗
  • MalfunctionJul 28, 2021Break; Break

    IT WAS REPORTED THAT A AR40E MODEL INTRAOCULAR LENS(IOL) HAD HAPTIC BROKE DURING MANIPULATION. ADDITIONAL DETAILS RECEIVED STATES THAT THE LENS BROKE DURING THE LOADING PROCESS AND WAS INSERTED INTO THE EYE. THE HAPTIC WAS NOTICED THAT IT WAS INDEED BROKEN ONCE THE LENS WAS INSERTED INTO THE EYE. THE DOCTOR HAD TO CUT THE LENS OUT OF THE EYE AND PROCEDURE WAS COMPLETED SUCCESSFULLY USING BACKUP LENS OF SAME MODEL AND DIOPTER. THERE WAS NO INJURY TO THE PATIENT AND NO MEDICAL/SURGICAL INTERVENTION SUCH AS INCISION ENLARGEMENT, VITRECTOMY OR SUTURES WERE REQUIRED. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12236262 ↗
  • MalfunctionJul 8, 2021Break; Break

    IT WAS REPORTED THAT DURING INTRAOCULAR LENS INJECTION, IT WAS FOUND THAT THE INTRAOCULAR LENS HAPTIC WAS BROKEN AND COULD NOT BE IMPLANTED. A NEW INTRAOCULAR LENS WAS USED AND THE SURGERY WAS UNEVENTFUL. PRODUCT NOT AVAILABLE FOR RETURN. IT IS UNCLEAR IF ISSUE OCCURRED WITH PATIENT CONTACT OR NOT. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12133286 ↗
  • InjuryApr 15, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT INITIALLY THERE WAS LOADING ERROR WITH SENSAR MONOFOCAL INTRAOCULAR LENS, SO SURGEON THOUGHT OF USING SENSAR MONOFOCAL INTRAOCULAR LENS WITH SAME MODEL AND SAME DIOPTER (MODEL AR40E +25.0 DIOPTER). HOWEVER, THIS LENS HAD BROKEN HAPTIC DURING SURGERY. SO, SURGEON IMPLANTED INITIAL LENS SUCCESSFULLY USING FORCEPS BY HAND AND NO CARTRIDGE WAS USED. DURING THE SAME PROCEDURE AN INCISION ENLARGEMENT AND SUTURES WERE REQUIRED. NO FURTHER INFORMATION PROVIDED. THIS REPORT WILL CAPTURE INFORMATION FOR LENS THAT REMAINS IMPLANTED. A SEPARATE REPORT IS BEING SUBMITTED FOR LENS THAT HAD A BROKEN HAPTIC ISSUE.

    Patient problems: No Code Available

    MAUDE report 8512879 ↗
  • InjuryApr 13, 2019Break; Break

    IT WAS REPORTED THAT INITIALLY THERE WAS LOADING ERROR WITH SENSAR MONOFOCAL INTRAOCULAR LENS, SO SURGEON THOUGHT OF USING SENSAR MONOFOCAL INTRAOCULAR LENS WITH SAME MODEL AND SAME DIOPTER (MODEL AR40E +25.0 DIOPTER). HOWEVER, THIS LENS HAD BROKEN HAPTIC DURING SURGERY. SO, SURGEON IMPLANTED INITIAL LENS SUCCESSFULLY USING FORCEPS BY HAND AND NO CARTRIDGE WAS USED. DURING THE SAME PROCEDURE INCISION ENLARGEMENT AND SUTURES WERE REQUIRED. NO FURTHER INFORMATION PROVIDED. THIS REPORT WILL CAPTURE INFORMATION FOR BROKEN HAPTIC ISSUE. ANOTHER REPORT IS BEING SUBMITTED FOR LENS THAT REMAINS IMPLANTED.

    Patient problems: No Code Available

    MAUDE report 8512000 ↗
  • MalfunctionAug 9, 2017Break; Break

    IT WAS REPORTED THAT DURING INSERTION INTO THE EYE, AN INTRAOCULAR LENS (IOL) WAS NOTICED BROKEN. PATIENT CONTACT WAS REPORTED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6780060 ↗
  • MalfunctionAug 9, 2017Break; Break

    IT WAS REPORTED THAT DURING INSERTION INTO THE EYE, AN INTRAOCULAR LENS (IOL) WAS NOTICED BROKEN. PATIENT CONTACT WAS REPORTED. REPORTEDLY, THE LENS WAS NOT IMPLANTED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6778811 ↗
  • InjuryJul 19, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED A AR40E 25.0 DIOPTER INTRAOCULAR LENS (IOL) WAS INSERTED AND REMOVED FROM A FEMALE PATIENT'S RIGHT EYE. ACCORDING TO THE PHYSICIAN THE LENS WAS FOLDED AND CHECKED BY TECHNICIAN AND REPRESENTATIVE. UPON ATTEMPT TO IMPLANT THE PHYSICIAN NOTICED THE EYE FILLED WITH BLOOD, IOL WAS REMOVED AND COULD NOT BE USED DUE TO THE PATIENT'S EYE. THROUGH FOLLOW-UP IT WAS LEARNED THAT THE DOCTOR POSSIBLY NICKED THE IRIS OR OTHER PARTS OF THE EYE WHILE DISSECTING BACK CONJUNCTIVA AND MAKING A HOLE IN THE SCLERA. THE LENS HAD NOTHING TO DO WITH THIS PART OF THE PROCEDURE. IOL WAS REMOVED AS A WHOLE. NO ISSUE WITH IOL, PATIENT ANATOMY AND COMPLICATION WITH PROCEDURE LED TO REMOVAL OF THE PERFECTLY IMPLANTED LENS. AT REMOVAL THE INCISION WAS ENLARGED AND SUTURED. THERE WAS NO VITRECTOMY. THE REASON THE PATIENT WAS THERE WAS BECAUSE THEY HAD ALREADY HAD A VITRECTOMY FROM A PREVIOUS SURGICAL COMPLICATION. PATIENT HAD NO ¿BAG¿ FOR NORMAL LENS PLACEMENT AND SO THE PHYSICIAN WAS TRYING TO SEW IN A LENS FOR THE PATIENT.

    Patient problems: No Code Available

    MAUDE report 6724072 ↗