Skip to content

SENSAR IOL AR40E (REF AR40E00025)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · ACRY 13.5mm 6OPT MOD C PMMA OPTIEDGE · Class 3 · Implant

All models of SENSAR IOL (112)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 05050474514065), received between Sep 7, 2018 and Apr 22, 2025 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Failure to Unfold or Unwrap2
  • Material Split, Cut or Torn2
  • Deformation Due to Compressive Stress1
  • Failure to Eject1
  • Material Fragmentation1
  • Mechanical Jam1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Known Impact Or Consequence To Patient2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 22, 2025Break; Mechanical Jam; Break; Mechanical Jam

    IT WAS REPORTED THAT THE LEADING HAPTIC OF THE PRELOADED MONOFOCAL INTRAOCULAR LENS (IOL) KICKED AFTER THE LENS WAS IMPLANTED IN THE ANTERIOR CHAMBER (AC), MOST LIKELY BECAUSE IT CAUGHT THE PLUNGER WHILE IN THE CARTRIDGE. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21884437 ↗
  • MalfunctionNov 22, 2023Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT WHEN INSERTING A MONOFOCAL INTRAOCULAR LENS (IOL) INTO THE PATIENT'S OPERATIVE EYE, THE SURGEON NOTICED THAT THE TRAILING HAPTIC HAD BROKEN OFF. THE SURGEON REMOVED THE LENS AND REPLACED IT WITH A BACK UP LENS. THERE WAS NO ISSUE. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18196721 ↗
  • MalfunctionJul 18, 2023Use of Device Problem; Material Split, Cut or Torn; Use of Device Problem; Material Split, Cut or Torn

    IT WAS REPORTED DURING IMPLANTATION THE NON-PRE-LOADED AR40E MODEL INTRAOCULAR LENS (IOL) WAS INSERTED INTO THE PATIENT¿S OCULAR SINISTER (LEFT EYE) AND REMOVED DUE TO TORN IOL DURING THE PROCEDURE. IT WAS ALSO MENTIONED ONLY THE CARTRIDGE TIP CONTACT HAD PATIENT CONTACT WITH PATIENT'S OS. THE IOL WAS WASTED BECAUSE THE TECH LOADED THE LENS INTO MACHINE INCORRECTLY. THE INCISION WAS ENLARGED HOWEVER THERE WAS NO PATIENT INJURY, NO SUTURE(S) AND NO VITRECTOMY. THE REPLACEMENT LENS WAS ANOTHER AR40E 4.0 DIOPTER IOL. PATIENT OUTCOME WAS REPORTED AS FULLY RECOVERED. THE SUSPECT IOL IS NOT AVAILABLE FOR RETURN AS IT WAS DESTROYED. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 17347337 ↗
  • MalfunctionApr 23, 2021Break; Break

    IT WAS REPORTED THAT AN INTRAOCULAR LENS (IOL) TRAILING HAPTIC BROKE DURING INSERTION. THE LENS WAS REMOVED FROM THE PATIENT'S EYE AND ANOTHER LENS INSERTED DURING THE SAME PROCEDURE. THERE WAS PATIENT CONTACT, BUT NO INJURY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11720293 ↗
  • InjuryMar 27, 2021Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT AFTER AN AR40E INTRAOCULAR LENS (IOL) WAS INSERTED INTO THE PATIENT'S RIGHT (OD) EYE, THE SURGEON OBSERVED A CRACK IN THE IOL. THE LENS WAS REMOVED AND REPLACED WITH AN UNKNOWN BACK UP LENS DURING THE SAME SURGERY. A VITRECTOMY WAS PERFORMED, HOWEVER THERE WAS NO COMPLICATION NOTED. THE REMOVED LENS HAS BEEN DISCARDED, AND THE PATIENT IS REPORTED TO BE DOING FINE. NO FURTHER INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11579856 ↗
  • MalfunctionSep 9, 2020Break; Failure to Unfold or Unwrap; Failure to Eject; Break; Failure to Unfold or Unwrap; Failure to Eject

    IT WAS REPORTED THAT AN INTRAOCULAR LENS (IOL) WAS PARTIALLY INSERTED INTO THE PATIENT¿S RIGHT DUE TO A BENT HAPTIC AND THE LENS WAS NOT UNFOLDING CORRECTLY. THE LENS WAS REMOVED AND REPLACED WITH ANOTHER LENS OF THE SAME MODEL AND DIOPTER DURING THE SAME PROCEDURE. IT WAS STATED THAT THERE WAS NO INCISION ENLARGEMENT, NO VITRECTOMY OR SUTURES REQUIRED. NO OTHER INFORMATION WAS AVAILABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10506055 ↗
  • MalfunctionJul 17, 2020Failure to Unfold or Unwrap; Failure to Unfold or Unwrap

    IT WAS REPORTED THAT AN INTRAOCULAR LENS (IOL) WAS INSERTED, THE HAPTIC FOLDED, HOWEVER, THE LENS WOULD NOT OPEN. THE LENS WAS REMOVED AND REPLACED WITH THE SAME MODEL AND DIOPTER LENS WITHOUT INCIDENT. THROUGH FOLLOW-UP, IT WAS CONFIRMED THERE WAS NO VITRECTOMY, INCISION ENLARGEMENT OR SUTURES REQUIRED. NO ADDITIONAL INFORMATION WAS PROVIDED TO JOHNSON & JOHNSON SURGICAL VISION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10292335 ↗
  • MalfunctionSep 7, 2018Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    IT WAS REPORTED THAT THE HAPTIC OF THE INTRAOCULAR LENS (IOL) KINKED DURING INSERTION INTO THE PATIENT'S OPERATIVE EYE. REPORTEDLY, THE IOL WAS REMOVED AND REPLACED IN THE SAME PROCEDURE. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7854800 ↗

Other intraocular lens devices

All intraocular lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page