AMO ADMINISTRATION SET OM2505191Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · ADMINISTRATION SET · Class 1
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 05050474520691), received between Jan 30, 2024 and Aug 22, 2024 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Component Missing” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Component Missing14
- Delivered as Unsterile Product14
- Misassembled14
Patient problems
- No Clinical Signs, Symptoms or Conditions14
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 22, 2024Misassembled; Delivered as Unsterile Product; Component Missing; Misassembled; Delivered as Unsterile Product; Component Missing
IT WAS REPORTED THAT SEVERAL IV ADMINISTRATION SETS HAD LOOSE/MISSING TIPS ON THE LEUR-LOCK END AND CUSTOMER WAS UNABLE TO USE THEM DUE TO STERILITY. THROUGH FOLLOW UP CUSTOMER CONFIRMED THAT THERE WAS A BREACH IN THE STERILE BARRIER. SOME WERE NOTED UPON OPENING THE PRODUCT AND A COUPLE WERE NOTICED BEFORE OPENING, WHILE LOOKING AT THE PACKAGING. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS 10 OF 10 REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20023982 ↗MalfunctionAug 21, 2024Misassembled; Delivered as Unsterile Product; Component Missing; Misassembled; Delivered as Unsterile Product; Component Missing
IT WAS REPORTED THAT SEVERAL IV ADMINISTRATION SETS HAD LOOSE/MISSING TIPS ON THE LEUR-LOCK END AND CUSTOMER WAS UNABLE TO USE THEM DUE TO STERILITY. THROUGH FOLLOW UP CUSTOMER CONFIRMED THAT THERE WAS A BREACH IN THE STERILE BARRIER. SOME WERE NOTED UPON OPENING THE PRODUCT AND A COUPLE WERE NOTICED BEFORE OPENING, WHILE LOOKING AT THE PACKAGING. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS 6 OF 10 REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20023164 ↗MalfunctionAug 21, 2024Misassembled; Delivered as Unsterile Product; Component Missing; Misassembled; Delivered as Unsterile Product; Component Missing
IT WAS REPORTED THAT SEVERAL IV ADMINISTRATION SETS HAD LOOSE/MISSING TIPS ON THE LEUR-LOCK END AND CUSTOMER WAS UNABLE TO USE THEM DUE TO STERILITY. THROUGH FOLLOW UP CUSTOMER CONFIRMED THAT THERE WAS A BREACH IN THE STERILE BARRIER. SOME WERE NOTED UPON OPENING THE PRODUCT AND A COUPLE WERE NOTICED BEFORE OPENING, WHILE LOOKING AT THE PACKAGING. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS 5 OF 10.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20017603 ↗MalfunctionAug 21, 2024Misassembled; Delivered as Unsterile Product; Component Missing; Misassembled; Delivered as Unsterile Product; Component Missing
IT WAS REPORTED THAT SEVERAL IV ADMINISTRATION SETS HAD LOOSE/MISSING TIPS ON THE LEUR-LOCK END AND CUSTOMER WAS UNABLE TO USE THEM DUE TO STERILITY. THROUGH FOLLOW UP CUSTOMER CONFIRMED THAT THERE WAS A BREACH IN THE STERILE BARRIER. SOME WERE NOTED UPON OPENING THE PRODUCT AND A COUPLE WERE NOTICED BEFORE OPENING, WHILE LOOKING AT THE PACKAGING. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS 8 OF 10.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20015502 ↗MalfunctionAug 20, 2024Misassembled; Delivered as Unsterile Product; Component Missing; Misassembled; Delivered as Unsterile Product; Component Missing
IT WAS REPORTED THAT SEVERAL IV ADMINISTRATION SETS HAD LOOSE/MISSING TIPS ON THE LEUR-LOCK END AND CUSTOMER WAS UNABLE TO USE THEM DUE TO STERILITY. THROUGH FOLLOW UP CUSTOMER CONFIRMED THAT THERE WAS A BREACH IN THE STERILE BARRIER. SOME WERE NOTED UPON OPENING THE PRODUCT AND A COUPLE WERE NOTICED BEFORE OPENING, WHILE LOOKING AT THE PACKAGING. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS 7 OF 10 REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20005450 ↗MalfunctionAug 20, 2024Misassembled; Delivered as Unsterile Product; Component Missing; Misassembled; Delivered as Unsterile Product; Component Missing
IT WAS REPORTED THAT SEVERAL IV ADMINISTRATION SETS HAD LOOSE/MISSING TIPS ON THE LEUR-LOCK END AND CUSTOMER WAS UNABLE TO USE THEM DUE TO STERILITY. THROUGH FOLLOW UP CUSTOMER CONFIRMED THAT THERE WAS A BREACH IN THE STERILE BARRIER. SOME WERE NOTED UPON OPENING THE PRODUCT AND A COUPLE WERE NOTICED BEFORE OPENING, WHILE LOOKING AT THE PACKAGING. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS 9 OF 10.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20003178 ↗MalfunctionAug 20, 2024Misassembled; Delivered as Unsterile Product; Component Missing; Misassembled; Delivered as Unsterile Product; Component Missing
IT WAS REPORTED THAT SEVERAL IV ADMINISTRATION SETS HAD LOOSE/MISSING TIPS ON THE LEUR-LOCK END AND CUSTOMER WAS UNABLE TO USE THEM DUE TO STERILITY. THROUGH FOLLOW UP CUSTOMER CONFIRMED THAT THERE WAS A BREACH IN THE STERILE BARRIER. SOME WERE NOTED UPON OPENING THE PRODUCT AND A COUPLE WERE NOTICED BEFORE OPENING, WHILE LOOKING AT THE PACKAGING. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS 1 OF 10 REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20002784 ↗MalfunctionAug 20, 2024Misassembled; Delivered as Unsterile Product; Component Missing; Misassembled; Delivered as Unsterile Product; Component Missing
IT WAS REPORTED THAT SEVERAL IV ADMINISTRATION SETS HAD LOOSE/MISSING TIPS ON THE LEUR-LOCK END AND CUSTOMER WAS UNABLE TO USE THEM DUE TO STERILITY. THROUGH FOLLOW UP CUSTOMER CONFIRMED THAT THERE WAS A BREACH IN THE STERILE BARRIER. SOME WERE NOTED UPON OPENING THE PRODUCT AND A COUPLE WERE NOTICED BEFORE OPENING, WHILE LOOKING AT THE PACKAGING. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS 3 OF 10 REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20001534 ↗MalfunctionAug 20, 2024Misassembled; Delivered as Unsterile Product; Component Missing; Misassembled; Delivered as Unsterile Product; Component Missing
IT WAS REPORTED THAT BD INSYTE AUTOG BC UNABLE TO CONNECT TO LUER CONNECTION AND LEAKING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ANGIO 22G LOT NUMBER ENDING IN 9757 ARE NOT SCREWING ON NORMALLY AND THUS CAUSING LEAKING. THEY NOTICED THIS HAPPENED A COUPLE OF TIMES LAST WEEK (IT HAPPED TO ME AS WELL-THOUGH IT WAS USER ERROR).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20000473 ↗MalfunctionAug 20, 2024Misassembled; Delivered as Unsterile Product; Component Missing; Misassembled; Delivered as Unsterile Product; Component Missing
IT WAS REPORTED THAT SEVERAL IV ADMINISTRATION SETS HAD LOOSE/MISSING TIPS ON THE LEUR-LOCK END AND CUSTOMER WAS UNABLE TO USE THEM DUE TO STERILITY. THROUGH FOLLOW UP CUSTOMER CONFIRMED THAT THERE WAS A BREACH IN THE STERILE BARRIER. SOME WERE NOTED UPON OPENING THE PRODUCT AND A COUPLE WERE NOTICED BEFORE OPENING, WHILE LOOKING AT THE PACKAGING. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS 2 OF 10 REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20000432 ↗MalfunctionJan 30, 2024Misassembled; Delivered as Unsterile Product; Component Missing; Misassembled; Delivered as Unsterile Product; Component Missing
IT WAS REPORTED THAT SEVERAL IV ADMIN SETS HAD LOOSE/MISSING TIPS ON THE LEUR-LOCK END AND CUSTOMER WAS UNABLE TO USE THEM DUE TO STERILITY. THROUGH FOLLOW UP CUSTOMER CONFIRMED THAT THERE WAS A BREACH IN THE STERILE BARRIER. SOME WERE NOTED UPON OPENING THE PRODUCT AND A COUPLE WERE NOTICED BEFORE OPENING, WHILE LOOKING AT THE PACKAGING. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS 4 OF 4 REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18610849 ↗MalfunctionJan 30, 2024Misassembled; Delivered as Unsterile Product; Component Missing; Misassembled; Delivered as Unsterile Product; Component Missing
IT WAS REPORTED THAT SEVERAL IV ADMIN SETS HAD LOOSE/MISSING TIPS ON THE LEUR-LOCK END AND CUSTOMER WAS UNABLE TO USE THEM DUE TO STERILITY. THROUGH FOLLOW UP CUSTOMER CONFIRMED THAT THERE WAS A BREACH IN THE STERILE BARRIER. SOME WERE NOTED UPON OPENING THE PRODUCT AND A COUPLE WERE NOTICED BEFORE OPENING, WHILE LOOKING AT THE PACKAGING. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS 3 OF 4 REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18610823 ↗MalfunctionJan 30, 2024Misassembled; Delivered as Unsterile Product; Component Missing; Misassembled; Delivered as Unsterile Product; Component Missing
IT WAS REPORTED THAT SEVERAL IV ADMIN SETS HAD LOOSE/MISSING TIPS ON THE LEUR-LOCK END AND CUSTOMER WAS UNABLE TO USE THEM DUE TO STERILITY. THROUGH FOLLOW UP CUSTOMER CONFIRMED THAT THERE WAS A BREACH IN THE STERILE BARRIER. SOME WERE NOTED UPON OPENING THE PRODUCT AND A COUPLE WERE NOTICED BEFORE OPENING, WHILE LOOKING AT THE PACKAGING. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS 2 OF 4 REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18610750 ↗MalfunctionJan 30, 2024Misassembled; Delivered as Unsterile Product; Component Missing; Misassembled; Delivered as Unsterile Product; Component Missing
IT WAS REPORTED THAT SEVERAL IV ADMINISTRATION SETS HAD LOOSE/MISSING TIPS ON THE LEUR-LOCK END AND CUSTOMER WAS UNABLE TO USE THEM DUE TO STERILITY. THROUGH FOLLOW UP CUSTOMER CONFIRMED THAT THERE WAS A BREACH IN THE STERILE BARRIER. SOME WERE NOTED UPON OPENING THE PRODUCT AND A COUPLE WERE NOTICED BEFORE OPENING, WHILE LOOKING AT THE PACKAGING. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS REPORT IS 1 OF 4 REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18610469 ↗
Other DEVICE, IRRIGATION, OCULAR SURGERY devices
- MST Yellow Silicone Sleeve for 2.4-2.75 incision SIL-0001-YMicrosurgical Technology, Inc. · Device, Irrigation, Ocular Surgery1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.