SOLO OPOIA20STRRecalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · I/A HP 20 GA STRAIGHT SILICONE SLV TIP · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 11
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 05050474521070), received between Dec 12, 2017 and Dec 27, 2023 — classified as 11 injuries.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Flaked4
- Material Erosion4
- Restricted Flow rate4
- Sharp Edges4
- Suction Problem3
- Infusion or Flow Problem1
Patient problems
- Capsular Bag Tear11
- Vitrectomy4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Intraocular Pressure Increased1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryDec 27, 2023Suction Problem; Infusion or Flow Problem; Suction Problem; Infusion or Flow Problem
IT WAS REPORTED THAT THERE WAS A CASE OF SUDDEN POSTERIOR VESICLE DAMAGE WHICH OCCURRED DURING IRRIGATION/ASPIRATION (IA) STEP FOR REMOVAL OF OPHTHALMIC VISCOSURGICAL DEVICE (OVD) ON THE PATIENT'S RIGHT EYE. THE DOCTOR THINKS THERE IS A POSSIBILITY THAT THERE IS A PROBLEM WITH THE SUCTION OF IA DUE TO THE IA TIP. THE PHYSICIAN HAS BEEN USING SIGPRO FOR MORE THAN THREE YEARS, BUT THIS IS THE FIRST TIME TO HAVE THESE ISSUES. THROUGH FOLLOW-UP WE LEARNED, THAT AN ISTENT WAS USED AND POSTERIOR CAPSULE DAMAGE DURING OVD REMOVAL AFTER INTRAOCULAR LENS (IOL) IMPLANTATION. THE IOL IMPLANTED WAS A DFW375 SERIAL NUMBER: (B)(6) DUE TO PATIENT INJURY THIS LENS WAS EXPLANTED, AND THE FOLLOWING LENS WAS IMPLANTED: AR40E 21.5D SERIAL NUMBER: (B)(6). DIAMOX WAS PRESCRIBED DUE TO INCREASED INTRAOCULAR PRESSURE. A VITRECTOMY WAS PERFORMED AND THE INCISION TO THE EYE WAS ENLARGED, REQUIRING SUTURES. THE OPERATION ON THE PATIENT'S LEFT EYE WAS POSTPONED. NO OTHER COMPLICATIONS WERE REPORTED AND HEALON NA 0.85 WAS USED FOR THESE CASE WITHOUT ISSUES RELATED TO IT. NO FURTHER DETAILS WERE PROVIDED. A SEPARATE REPORT IS BEING SUBMITTED FOR THE OTHER JNJ DEVICE INVOLVED FOR THIS PATIENT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Capsular Bag Tear; Intraocular Pressure Increased
MAUDE report 18400905 ↗InjuryAug 24, 2021Break; Sharp Edges; Break; Sharp Edges
IT WAS REPORTED THAT POSTERIOR CAPSULE RAPTURE OCCURRED WHILE SURGEON WAS POLISHING WITH THE IRRIGATION/ASPIRATION HANDPIECE DURING PROCEDURE. AN ANTERIOR VITRECTOMY WAS PERFORMED AND PROCEDURE WAS SUCCESSFULLY COMPLETED. IT WAS ALLEGED THAT THE HANDPIECE TIP MIGHT HAVE HAD BURRS.
Patient problems: Capsular Bag Tear
MAUDE report 12362660 ↗InjuryAug 24, 2021Break; Sharp Edges; Break; Sharp Edges
IT WAS REPORTED THAT POSTERIOR CAPSULE RAPTURE OCCURRED WHILE SURGEON WAS POLISHING WITH THE IRRIGATION/ASPIRATION HANDPIECE DURING PROCEDURE. AN ANTERIOR VITRECTOMY WAS PERFORMED AND PROCEDURE WAS SUCCESSFULLY COMPLETED. IT WAS ALLEGED THAT THE HANDPIECE TIP MIGHT HAVE HAD BURRS.
Patient problems: Capsular Bag Tear
MAUDE report 12362463 ↗InjuryAug 24, 2021Break; Sharp Edges; Break; Sharp Edges
IT WAS REPORTED THAT POSTERIOR CAPSULE RAPTURE OCCURRED WHILE SURGEON WAS POLISHING WITH THE IRRIGATION/ASPIRATION HANDPIECE DURING PROCEDURE. AN ANTERIOR VITRECTOMY WAS PERFORMED AND PROCEDURE WAS SUCCESSFULLY COMPLETED. IT WAS ALLEGED THAT THE HANDPIECE TIP MIGHT HAVE HAD BURRS.
Patient problems: Capsular Bag Tear
MAUDE report 12362405 ↗InjuryAug 24, 2021Break; Sharp Edges; Break; Sharp Edges
IT WAS REPORTED THAT POSTERIOR CAPSULE RAPTURE OCCURRED WHILE SURGEON WAS POLISHING WITH THE IRRIGATION/ASPIRATION HANDPIECE DURING PROCEDURE. AN ANTERIOR VITRECTOMY WAS PERFORMED AND PROCEDURE WAS SUCCESSFULLY COMPLETED. IT WAS ALLEGED THAT THE HANDPIECE TIP MIGHT HAVE HAD BURRS.
Patient problems: Capsular Bag Tear
MAUDE report 12362362 ↗InjuryOct 5, 2018Material Erosion; Flaked; Restricted Flow rate; Material Erosion; Flaked; Restricted Flow rate
DURING A CATARACT PROCEDURE, A BROKEN POSTERIOR CAPSULE OCCURRED ON THE OPERATIVE EYE RESULTING IN A VITRECTOMY PROCEDURE. A DESCRIPTION FROM THE SURGERY CENTER INDICATED DURING THE IRRIGATION AND ASPIRATION PORTION, THERE WAS RESISTANCE THAT WAS CAUGHT AT THE PHACO TIP. IT WAS OBSERVED THAT THE GOLD COLOR SURFACE OF THE TIP SEEMED TO BE STRIPPING AWAY. THE BROKEN CAPSULE RESULTED IN A VITRECTOMY PROCEDURE AND THE ANTERIOR VITREOUS ZONE WAS RESECTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS 4 OF 4 REPORTS.
Patient problems: Capsular Bag Tear; Vitrectomy
MAUDE report 7939688 ↗InjuryOct 5, 2018Material Erosion; Flaked; Restricted Flow rate; Material Erosion; Flaked; Restricted Flow rate
DURING A CATARACT PROCEDURE, A BROKEN POSTERIOR CAPSULE OCCURRED ON THE OPERATIVE EYE RESULTING IN A VITRECTOMY PROCEDURE. A DESCRIPTION FROM THE SURGERY CENTER INDICATED DURING THE IRRIGATION AND ASPIRATION PORTION, THERE WAS RESISTANCE THAT WAS CAUGHT AT THE PHACO TIP. IT WAS OBSERVED THAT THE GOLD COLOR SURFACE OF THE TIP SEEMED TO BE STRIPPING AWAY. THE BROKEN CAPSULE RESULTED IN A VITRECTOMY PROCEDURE AND THE ANTERIOR VITREOUS ZONE WAS RESECTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS 3 OF 4 REPORTS.
Patient problems: Capsular Bag Tear; Vitrectomy
MAUDE report 7939680 ↗InjuryOct 5, 2018Material Erosion; Flaked; Restricted Flow rate; Material Erosion; Flaked; Restricted Flow rate
DURING A CATARACT PROCEDURE, A BROKEN POSTERIOR CAPSULE OCCURRED ON THE OPERATIVE EYE RESULTING IN A VITRECTOMY PROCEDURE. A DESCRIPTION FROM THE SURGERY CENTER INDICATED DURING THE IRRIGATION AND ASPIRATION PORTION, THERE WAS RESISTANCE THAT WAS CAUGHT AT THE PHACO TIP. IT WAS OBSERVED THAT THE GOLD COLOR SURFACE OF THE TIP SEEMED TO BE STRIPPING AWAY. THE BROKEN CAPSULE RESULTED IN A VITRECTOMY PROCEDURE AND THE ANTERIOR VITREOUS ZONE WAS RESECTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS 2 OF 4 REPORTS.
Patient problems: Capsular Bag Tear; Vitrectomy
MAUDE report 7939668 ↗InjuryOct 5, 2018Material Erosion; Flaked; Restricted Flow rate; Material Erosion; Flaked; Restricted Flow rate
DURING A CATARACT PROCEDURE, A BROKEN POSTERIOR CAPSULE OCCURRED ON THE OPERATIVE EYE RESULTING IN A VITRECTOMY PROCEDURE. A DESCRIPTION FROM THE SURGERY CENTER INDICATED DURING THE IRRIGATION AND ASPIRATION PORTION, THERE WAS RESISTANCE THAT WAS CAUGHT AT THE PHACO TIP. IT WAS OBSERVED THAT THE GOLD COLOR SURFACE OF THE TIP SEEMED TO BE STRIPPING AWAY. THE BROKEN CAPSULE RESULTED IN A VITRECTOMY PROCEDURE AND THE ANTERIOR VITREOUS ZONE WAS RESECTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS 1 OF 4 REPORTS.
Patient problems: Capsular Bag Tear; Vitrectomy
MAUDE report 7939358 ↗InjuryDec 12, 2017Suction Problem; Suction Problem
DURING A CATARACT EXTRACTION PROCEDURE, A HOSPITAL REPORTED THE POSTERIOR CHAMBER RUPTURED DURING THE IRRIGATION AND ASPIRATION SEGMENT. THE HOSPITAL INDICATED THE INTRAOCULAR LENS (IOL) WAS IMPLANTED IN THE CAPSULAR BAG AS THERE WAS NO VITREOUS THAT CAME OUT. THE PATIENT WAS DISCHARGED FROM THE HOSPITAL WITHOUT ANY PROBLEMS. THE HOSPITAL REQUESTED TO CHECK AND ADJUST THE VACUUM PRESSURE AND CHECK THE CONDITION OF THE TIP OF I/A. THIS REPORT IS FOR THE IRRIGATION/ASPIRATION (I/A TIP). A SEPARATE REPORT WILL BE SUBMITTED FOR THE PHACOFRAGMENTATION UNIT.
Patient problems: Capsular Bag Tear
MAUDE report 7111168 ↗InjuryDec 12, 2017Suction Problem; Suction Problem
DURING A CATARACT EXTRACTION PROCEDURE, A HOSPITAL REPORTED THE POSTERIOR CHAMBER RUPTURED DURING THE IRRIGATION AND ASPIRATION SEGMENT. THE HOSPITAL INDICATED THE IOL WAS IMPLANTED OUTSIDE THE CAPSULAR BAG AS THE POSTERIOR CAPSULE SPLIT. THE PATIENT WAS DISCHARGED FROM THE HOSPITAL WITHOUT ANY PROBLEMS. THE HOSPITAL REQUESTED TO CHECK AND ADJUST THE VACUUM PRESSURE AND CHECK THE CONDITION OF THE TIP OF I/A. THIS REPORT IS FOR THE IRRIGATION/ASPIRATION (I/A TIP). A SEPARATE REPORT WILL BE SUBMITTED FOR THE PHACOFRAGMENTATION UNIT.
Patient problems: Capsular Bag Tear
MAUDE report 7111023 ↗
Other Cannula, ophthalmic devices
- iTrack Advance iTrack-ADSNova Eye, Inc. · Cannula, Ophthalmic53 reports
- iTrack 250ANova Eye, Inc. · Cannula, Ophthalmic1 recall2 reports
- Vortex Surgical Inc VS0200.25Vortex Surgical, Inc. · Cannula, Ophthalmic1 recall0 reports
- Vortex Surgical Inc VS0275.27Vortex Surgical, Inc. · Cannula, Ophthalmic1 recall0 reports
- Vortex Surgical Inc VS0220.25Vortex Surgical, Inc. · Cannula, Ophthalmic1 recall0 reports
- Visitec® 585278Beaver-Visitec International, Inc. · Cannula, Ophthalmic1 recall0 reports
- Vortex Surgical Inc VS0205.23Vortex Surgical, Inc. · Cannula, Ophthalmic1 recall0 reports
- Vortex Surgical Inc VS0250.25Vortex Surgical, Inc. · Cannula, Ophthalmic1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026