UNFOLDER EMERALD EMERALDCRecalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · Emerald Cartridge-Thermoform TR, 30 bx · Class 1

Quick summary

This Unfolder Emerald cartridge, from Johnson & Johnson Surgical Vision, is a loading device for intraocular lenses used in eye surgery. Its 250 reports often describe cracked, broken or deformed cartridge tips, and some describe lenses damaged or partly inserted during surgery. The reports include 28 injuries and no deaths. Reports dropped from 7 in the previous 12 months to 1 in the most recent 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
250
Deaths reported
0
Injuries reported
28
Recalls naming it
0

Summary

The FDA's MAUDE database contains 250 adverse event reports that list this exact device (DI 05050474530027), received between Aug 31, 2016 and Jan 6, 2026 — classified as 28 injuries, 222 malfunctions.

The most frequently reported specific device problem is “Crack” (76 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 2 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 2 reportsFeb 2025: 1 reportsMar 2025: 3 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack76
  • Break51
  • Device Contamination with Chemical or Other Material33
  • Material Deformation30
  • Contamination26
  • Appropriate Device Problem Term/Code Not Available14
  • Device Difficult to Setup or Prepare12
  • Scratched Material11

Patient problems

  • No Consequences Or Impact To Patient78
  • No Clinical Signs, Symptoms or Conditions48
  • No Known Impact Or Consequence To Patient48
  • No Patient Involvement42
  • No Code Available14
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available12
  • Capsular Bag Tear6
  • Eye Injury3

Recent adverse event reports

The 25 most recent of 250 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 6, 2026Break

    IT WAS REPORTED THAT A MONOFOCAL INTRAOCULAR LENS (IOL) HAD ITS HAPTIC BENT DUE TO LENS INJECTOR MALFUNCTIONING. CONSEQUENTLY, THE LENS WAS ONLY PARTIALLY INSERTED. HOWEVER, IT WAS ALSO REPORTED THAT THE LENS WAS INSERTED AND REMOVED DURING THE SAME PROCEDURE. A REPLACEMENT AR40E 23.0 DIOPTER LENS WITH SERIAL NUMBER (B)(6) WAS SUCCESSFULLY IMPLANTED USING A NEW INJECTOR WITHOUT COMPLICATIONS. THE PROCEDURE WAS DELAYED BY APPROXIMATELY TWO MINUTES. THE DIRECTIONS FOR USE WERE FOLLOWED, AND NO UNPLANNED INCISION ENLARGEMENT, SUTURES, OR VITRECTOMY OCCURRED. DETAILS REGARDING PATIENT OUTCOME AND STATUS REMAINS UNKNOWN. NO ADDITIONAL INFORMATION ABOUT THE COMPLAINT IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23986085 ↗
  • MalfunctionMar 27, 2025Crack; Crack

    IT WAS REPORTED THAT THE CARTRIDGE TIP WAS NOT EVEN. IT LOOKED LIKE IT WAS MELTED. ACCOUNT INDICATED THAT THE TIP WAS IN CONTACT WITH THE PATIENT. THE SURGEON WAS ABLE TO IMPLANT THE LENS WITH THE DAMAGED CARTRIDGE AND THERE WERE NO PROBLEMS WITH THE PATIENT AFTERWARDS. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21715637 ↗
  • MalfunctionMar 17, 2025Break; Break

    IT WAS REPORTED THAT THE CARTRIDGE WAS DAMAGED WHICH CAUSED DAMAGE TO THE LENS. THE LENS WAS REMOVED AND REPLACED. THE PRODUCT WAS DISCARDED UPON COMPLETING THE PROCEDURE. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21627514 ↗
  • MalfunctionMar 17, 2025Crack; Crack

    IT WAS REPORTED THAT WHEN THE TIP OF THE CARTRIDGE WAS VIEWED UNDER THE MICROSCOPE, IT PRESENTED GROOVES (SHARP BURRS ALONG THE EDGE). DURING THE PROCEDURE, IT WAS NECESSARY TO MAKE AN INCISION LARGER THAN USUAL IN THE PATIENT'S EYE. THE MATERIAL WAS DISCARDED UPON COMPLETING THE PROCEDURE. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 21627312 ↗
  • InjuryFeb 7, 2025Break; Break

    IT WAS REPORTED THAT THE NON-PRELOADED INTRAOCULAR LENS (IOL) PRESENTED A HAPTIC DAMAGED UPON LENS IMPLANTATION IN THE PATIENT'S RIGHT EYE. PRE-OPERATIVE VISUAL ACUITY WAS 30 CM. THE INSTRUCTIONS FOR USE WERE FOLLOWED. ACCOUNT INDICATED THAT THE LENS WAS PARTIALLY INSERTED IN THE PATIENT'S EYE. THE INCISION WAS ENLARGED, AND THE PATIENT UNDERWENT A VITRECTOMY. THE PATIENT HAS FULLY RECOVERED. THROUGH FOLLOW-UP, WE LEARNED THAT THE CARTRIDGE TIP WAS NOT DAMAGED AND THERE WAS NO RESISTANCE WHEN HANDLING THE CARTRIDGE. NO ADDITIONAL SURGICAL INTERVENTION IS PLANNED. NO FURTHER INFORMATION WAS PROVIDED. A SEPARATE REPORT (MFR REPORT 3012236936-2025-0000238) WILL BE SUBMITTED FOR THE LENS MENTIONED ABOVE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 21339861 ↗
  • MalfunctionJan 31, 2025Contamination; Contamination

    IT WAS FOUND THERE WAS FOREIGN MATERIAL DURING INTRAOCULAR LENS IMPLANTATION. THERE WAS NO DELAY IN TREATMENT OR OTHER INTERVENTIONS PERFORMED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21285654 ↗
  • MalfunctionJan 10, 2025Break; Break

    PATIENT GENDER: FEMALE. THE NON PRE-LOADED AR40E MODEL INTRAOCULAR LENS (IOL) WAS DAMAGED DURING LOADING, TECHNICIAN ERROR WAS REPORTED AND ISSUE RELATED TO USE ERROR WAS CONFIRMED. THE EMERALD CARTRIDGE WAS THE ISSUE THAT CAUSED THE TECHNICIAN TO LOAD THE IOL INCORRECTLY HOWEVER, THE SPECIFIC CARTRIDGE ISSUE WAS NOT PROVIDED. THERE WAS PATIENT CONTACT HOWEVER, NO MEDICAL INTERVENTION AND NO SURGICAL INTERVENTION WAS PERFORMED DURING THE PROCEDURE. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21131995 ↗
  • MalfunctionDec 31, 2024Break; Material Fragmentation; Break; Material Fragmentation

    IT WAS REPORTED THAT THE NON-PRELOADED INTRAOCULAR LENS (IOL) HAPTIC WAS DETACHED INSIDE THE INJECTOR DURING IMPLANTATION. THE INJECTOR WAS DEFECTIVE AND THE LENS WAS IN CONTACT WITH THE PATIENT'S EYE. THROUGH FOLLOW-UP, WE LEARNED, THAT THE PATIENT IS WELL. THE LENS WAS REMOVED AND REPLACED WITH A NEW LENS WITH THE SAME DIOPTER. NO FURTHER INFORMATION WAS PROVIDED. NOTE: A SEPARATE REPORT WILL BE SUBMITTED UNDER THE NUMBER 3012236936-2024-0003488 TO CAPTURE THE INTRAOCULAR LENS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21048627 ↗
  • MalfunctionOct 10, 2024Crack; Device Difficult to Setup or Prepare; Crack; Device Difficult to Setup or Prepare

    PROBLEMS WITH THE LAST BATCH OF THE EMERALD CARTRIDGES WERE REPORTED BY THE DOCTOR. THEY SEEM TO HAVE BECOME NARROWER AND THE DOCTOR IS HAVING DIFFICULTY IMPLANTING THE LENSES, THEY ARE USING THE CARTRIDGES WITH THE EMERALD INJECTOR. ADDITIONAL INFORMATION WAS RECEIVED CLARIFYING THE LENS WAS INSERTED, THERE WAS NO INJURY, AND THE SURGERY WAS SUCCESSFULLY COMPLETED. THE NEW BATCH THAT CAME IN WE USED WITHOUT HAVING THIS PROBLEM. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20414070 ↗
  • MalfunctionOct 9, 2024Crack; Device Difficult to Setup or Prepare; Crack; Device Difficult to Setup or Prepare

    PROBLEMS WITH THE LAST BATCH OF THE EMERALD CARTRIDGES WERE REPORTED BY THE DOCTOR. THEY SEEM TO HAVE BECOME NARROWER AND THE DOCTOR IS HAVING DIFFICULTY IMPLANTING THE LENSES, THEY ARE USING THE CARTRIDGES WITH THE EMERALD INJECTOR. ADDITIONAL INFORMATION WAS RECEIVED CLARIFYING THE LENS WAS INSERTED, NO INJURY TO PATIENT, AND THE SURGERY WAS COMPLETED SUCCESSFULLY. THE ONLY PROBLEM WITH THE EMERALD CARTRIDGE, THE LENS IS TOO TIGHT AT THE TIP, REQUIRING A LOT OF FORCE ON THE INJECTOR THREAD. CARTRIDGE BEVEL IS FRAYED. THE FACILITY ALSO REPORTS THERE IS STILL ANOTHER CLOSED BOX OF THE CARTRIDGE MODEL. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20413961 ↗
  • InjurySep 2, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT WHEN USING AN INJECTOR TO IMPLANT A MONOFOCAL INTRAOCULAR LENS (IOL) INTO A PATIENT¿S RIGHT EYE, THE HAPTIC WAS AMPUTATED BY THE INJECTOR. THE IOL WAS REMOVED WITH PACMAN TECHNIQUE. THEY USED ANOTHER LENS OF THE SAME DIOPTER WITH THE SAME CARTRIDGE TO COMPLETE THE SURGERY. A CORNEAL INCISION ENLARGEMENT WAS MADE WITH MANUAL IMPLANTATION OF 3-PIECE SENSAR LOI. EXTERNALIZATION OF HAPTICS VIA NEEDLES MADE, FLANGE CONNECTION WITH BATTERY-OPERATED CAUTERY. SUPERIOR CORNEAL SUTURE (4 STITCHES) MADE. BALANCE SALT SOLUTION (BSS) WAS USED FOR HYDRATION OF THE DEVICE. IT WAS CONFIRMED THAT THE TIP OF THE CARTRIDGE WAS NOT DAMAGED AND THAT THE ISSUE WAS NOT DUE TO MISUSE. IT WAS NOTED THAT THE PATIENT IS WELL. THE LENS AND CARTRIDGE ARE NOT AVAILABLE FOR RETURN AS THEY WERE DISCARDED DURING THE SURGERY BY THE ACCOUNT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Capsular Bag Tear

    MAUDE report 20116290 ↗
  • MalfunctionAug 14, 2024Crack; Crack

    IT WAS REPORTED THAT THE UNKNOWN NON-PRELOADED INTRAOCULAR LENS (IOL) HAD MARKS ON SURFACE AFTER IMPLANTATION. IOL HAD TO BE REMOVED DURING THE SAME PROCEDURE. THE CARTRIDGE WAS SPLITTING ON PARTS WHERE IT FOLDED TOGETHER. THERE WAS INCISION ENLARGEMENT AND SUTURES WERE REQUIRED. NO FURTHER INFORMATION IS PROVIDED. THIS REPORT IS FOR THE CARTRIDGE. A SEPARATE REPORT IS BEING SUBMITTED FOR THE INTRAOCULAR LENS.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 19988287 ↗
  • InjuryJul 15, 2024Mechanical Jam; Ejection Problem; Mechanical Jam; Ejection Problem

    IT WAS REPORTED THE NON PRE-LOADED AR40E MODEL INTRAOCULAR LENS (IOL) WAS STUCK IN THE EMERALDC30 CARTRIDGE. ALTHOUGH THE EXTENT OF PATIENT CONTACT IS UNKNOWN, CONTACT DID OCCUR. AN UNPLANNED VITRECTOMY WAS PERFORMED HOWEVER, THERE WAS NO PATIENT INJURY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING ANOTHER IOL. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Eye Injury; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19743382 ↗
  • InjuryJul 6, 2024Crack; Crack

    IT WAS REPORTED THAT AN UNFOLDER EMERALD CARTRIDGE HAS A BUMP IN THE PLASTIC AND CUTTING THE PATIENT'S OPERATIVE EYE. THE UNEVEN CARTRIDGE TIP WAS CUTTING THE PATIENT, THE TIP IS UNEVEN AND NOT SMOOTH. THERE WAS NO MEDICAL OR SURGICAL INTERVENTION OUTSIDE THE STANDARD OF CARE REQUIRED. THE PRODUCT WAS DISCARDED. NO FURTHER INFORMATION PROVIDED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Eye Injury

    MAUDE report 19685210 ↗
  • MalfunctionJun 21, 2024Crack; Ejection Problem; Crack; Ejection Problem

    IT WAS REPORTED THAT DURING THE INSERTION THE SURGEON NOTICED A BUMP ON THE PLASTIC OF AN EMERALD CARTRIDGE. IT WAS REPORTED IT WAS NOT A SMOOTH INSERTION, DESPITE SURGEON EXTENDING THE INCISION IN THE PATIENT'S OPERATIVE EYE. THE PRODUCT BOX WAS DISCARDED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 19590196 ↗
  • MalfunctionJun 21, 2024Crack; Ejection Problem; Crack; Ejection Problem

    IT WAS REPORTED THAT DURING THE INSERTION THE SURGEON NOTICED A BUMP ON THE PLASTIC OF AN EMERALD CARTRIDGE. IT WAS REPORTED IT WAS NOT A SMOOTH INSERTION, DESPITE SURGEON EXTENDING THE INCISION IN THE PATIENT'S OPERATIVE EYE. THE PRODUCT BOX WAS DISCARDED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 19589474 ↗
  • MalfunctionMay 17, 2024Crack; Crack

    IT WAS REPORTED THAT THE TIP OF THE CARTRIDGE CRACKED DURING THE IMPLANTATION OF THE IOL INTO THE PATIENT'S EYE. THERE WAS NO DELAY IN TREATMENT OR OTHER INTERVENTIONS PERFORMED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19346915 ↗
  • MalfunctionApr 2, 2024Crack; Crack

    IT WAS REPORTED THAT THE PATIENT SUFFERED FROM BLURRED VISION IN THE LEFT EYE FOR A YEAR AND UNDERWENT PHACOEMULSIFICATION AND INTRAOCULAR LENS IMPLANTATION. DURING THE USE OF THE INTRAOCULAR LENS, IT WAS FOUND THAT THE TIP OF THE CARTRIDGE WAS DAMAGED, BUT IT DID NOT CAUSE HARM TO THE PATIENT. THE PROCEDURE COMPLETED SUCCESSFULLY. THERE WAS NO DELAY IN TREATMENT OR OTHER INTERVENTIONS PERFORMED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19024199 ↗
  • MalfunctionMar 14, 2024Contamination; Contamination

    IT WAS REPORTED THAT THE JOHNSON AND JOHNSON(JNJ) PLATINUM AND EMERALD CARTRIDGES ARE LEAVING A FILMY OR STICKY SUBSTANCE ON THE BOTTOM SIDE OF THE LENS WHEN IMPLANTING. TWO EVENTS WERE REPORTED AGAINST THE PLATINUM CARTRIDGE AND ONE EVENT AGAINST THE EMERALD CARTRIDGE. THIS REPORT CAPTURES THE EMERALD CARTRIDGE EVENT LOT# CM14308. THE FILMY OR STICKY SUBSTANCE IS CAUSING THE DOCTOR TO GO IN AND CLEAN OFF THE SUBSTANCE ON THE BOTTOM OF THE LENS EACH TIME IT HAPPENS. THE DOCTOR SPOKE TO HIS COLLEAGUE WHO INFORMED HIM THAT THESE CARTRIDGES HAD THE SAME ISSUE IN THE PAST, YEARS AGO WHERE IT WAS THE COATING THAT IS CALLED GLYCEROL MONOSTEARATE(GMS). HOWEVER, UPON FURTHER REVIEW IT WAS CLARIFIED WITH THE CUSTOMER THAT THE PLATINUM (1MTEC30) CARTRIDGE IS COATED WITH HYALURONIC ACID (HA) AND THE EMERALD CARTRIDGE IS COATED WITH GMS. THE CUSTOMER TRIED TO GET A VIDEO OF WHAT WAS GOING ON BUT IT WAS IMPOSSIBLE WHEN LOOKING THROUGH THE MICROSCOPE AS IT WASN¿T VISIBLE ON THE VIEWING MONITOR. THE CUSTOMER RECEIVED NEW CARTRIDGES IN HOPES THAT IT WOULD FIX THE ISSUE SINCE THE ISSUE AROSE WHILE USING OLDER CARTRIDGES WHICH WERE NEARING EXPIRATION DATE ALTHOUGH NOT EXPIRED. IT APPEAR THAT THE ISS

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18900244 ↗
  • MalfunctionMar 7, 2024Crack; Crack

    IT WAS REPORTED THERE WAS CARTRIDGE TIP DAMAGED. THIS HAPPENED DURING THE IMPLANTATION OF THE IOL. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18862374 ↗
  • MalfunctionJan 29, 2024Crack; Mechanical Jam; Crack; Mechanical Jam

    IT WAS REPORTED THAT THE JOHNSON AND JOHNSON(JNJ) NON PRELOADED CARTRIDGE BROKE PRIOR TO THE LENS INSERTION. THE ISSUE OCCURRED DURING IMPLANTATION. THE TIP OF THE CARTRIDGE TOUCHED THE EYE. CUSTOMER SPECIFIED THE DIRECTIONS FOR USE WERE FOLLOWED AND NO USE ERROR LEAD TO THIS EVENT. A JNJ BACKUP LENS OF A DIFFERENT MODEL DIB00 19.0 DIOPTER WAS SUCCESSFULLY INSERTED. THERE WAS A 15 MINUTE DELAY. NO INCISION ENLARGEMENT, VITRECTOMY, SUTURES, OR MEDICATION OUTSIDE THE STANDARD OF CARE. THERE WAS NO HARM TO THE PATIENT. THE PATIENT¿S DAILY ACTIVITIES HAVE NOT BEEN IMPACTED BY THIS EVENT. PATIENT HAS FULLY RECOVERED. THE SUBSEQUENT PRODUCT INVESTIGATION SHOWS THE CARTRIDGE TIP IS DAMAGED MAKING THIS EVENT REPORTABLE. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18602363 ↗
  • MalfunctionNov 2, 2023Break; Device Difficult to Setup or Prepare; Scratched Material; Break; Device Difficult to Setup or Prepare; Scratched Material

    IT WAS REPORTED THAT THE INTRAOCULAR LENS (IOL), MODEL ZA9003/DIOPTER 19.5, WAS DIFFICULT TO BE TWISTED OUT OF THE CARTRIDGE AND SMALL SCRATCHES APPEARED ON THE EDGE OF THE OPTICS DUE TO THIS ISSUES. THE IOL WAS IMPLANTED. THROUGH FOLLOW-UP WE LEARNED THAT THE PATIENT IS DOING FINE AND THE LENS DID NOT HAVE TO BE EXPLANTED. NO FURTHER DETAILS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18062404 ↗
  • MalfunctionJun 27, 2023Material Fragmentation; Material Split, Cut or Torn; Material Fragmentation; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE CARTRIDGES CAUSED DAMAGE TO SEVERAL AR40E INTRAOCULAR LENSES (IOL'S) DURING THE PROCEDURE ON A SINGLE PATIENT. THIS CAUSED NO LASTING ISSUES FOR THE PATIENT. THROUGH FOLLOW-UP WE LEARNED THAT FOR 2 LENSES, PARTS WERE TORN OFF AFTER ADVANCEMENT TROUGH THE CARTRIDGE AND THEY WERE REMOVED AND REPLACED. TWO (2) FURTHER LENSES WERE SCRATCHED OR CRACKED BY THE CARTRIDGES, WHERE THE BACK HAPTIC GOES INTO THE OPTIC AFTER PUSHING THROUGH THE INSERTER. ONE OF THESE LENSES WAS REMOVED WHILE THE SECOND REMAINS IMPLANTED WITHOUT ISSUES FOR THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR THE SECOND OF TWO LENSES WITH PARTS TORN OFF THE OPTIC, WHICH WAS REMOVED AND REPLACED FROM THE PATIENTS EYE. SEPARATE REPORTS WILL BE SUBMITTED FOR THE OTHER DEVICES INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17213373 ↗
  • MalfunctionJun 27, 2023Material Fragmentation; Material Split, Cut or Torn; Material Fragmentation; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE CARTRIDGES CAUSED DAMAGE TO SEVERAL AR40E INTRAOCULAR LENSES (IOL'S) DURING THE PROCEDURE ON A SINGLE PATIENT. THIS CAUSED NO LASTING ISSUES FOR THE PATIENT. THROUGH FOLLOW-UP WE LEARNED THAT FOR 2 LENSES, PARTS WERE TORN OFF AFTER ADVANCEMENT TROUGH THE CARTRIDGE AND THEY WERE REMOVED AND REPLACED. TWO (2) FURTHER LENSES WERE SCRATCHED OR CRACKED BY THE CARTRIDGES, WHERE THE BACK HAPTIC GOES INTO THE OPTIC AFTER PUSHING THROUGH THE INSERTER. ONE OF THESE LENSES WAS REMOVED WHILE THE SECOND REMAINS IMPLANTED WITHOUT ISSUES FOR THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR THE FIRST OF TWO LENSES WITH PARTS TORN OFF THE OPTIC BY THE CARTRIDGE, WHICH WAS REMOVED AND REPLACED FROM THE PATIENT'S EYE. SEPARATE REPORTS WILL BE SUBMITTED FOR THE OTHER DEVICES INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17213366 ↗
  • MalfunctionJun 27, 2023Break; Break

    IT WAS REPORTED THAT THE CARTRIDGES CAUSED DAMAGE TO SEVERAL AR40E INTRAOCULAR LENSES (IOL'S) DURING THE PROCEDURE ON A SINGLE PATIENT. THIS CAUSED NO LASTING ISSUES FOR THE PATIENT. THROUGH FOLLOW-UP WE LEARNED THAT FOR 2 LENSES, PARTS WERE TORN OFF AFTER ADVANCEMENT TROUGH THE CARTRIDGE AND THEY WERE REMOVED AND REPLACED. TWO (2) FURTHER LENSES WERE SCRATCHED OR CRACKED BY THE CARTRIDGES, WHERE THE BACK HAPTIC GOES INTO THE OPTIC AFTER PUSHING THROUGH THE INSERTER. ONE OF THESE LENSES WAS REMOVED WHILE THE SECOND REMAINS IMPLANTED WITHOUT ISSUES FOR THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR THE LENS WITH SCRATCHES/ CRACKS BY THE CARTRIDGE THAT REMAINS IMPLANTED IN THE PATIENT'S EYE. SEPARATE REPORTS WILL BE SUBMITTED FOR THE OTHER DEVICES INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17213327 ↗

Other LENS, GUIDE, INTRAOCULAR devices

All LENS, GUIDE, INTRAOCULAR devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.