TECNIS CL IOL Z9002 (REF Z900200235)Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · TECNIS CL MONO 13MM 23.5 · Class 3 · Implant
All models of TECNIS CL IOL (42)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This is a Johnson & Johnson Tecnis CL intraocular lens, an implant used during cataract surgery to replace the eye's natural lens. Most of the 27 reports describe the lens breaking, cracking, tearing or having bent haptics during implantation, in some cases requiring the lens to be removed and replaced during the same surgery. The reports include 13 injuries and no deaths, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 27
- Deaths reported
- 0
- Injuries reported
- 13
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 05050474530768), received between Sep 30, 2016 and Oct 31, 2022 — classified as 13 injuries, 14 malfunctions.
The most frequently reported specific device problem is “Break” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break11
- Adverse Event Without Identified Device or Use Problem5
- Material Split, Cut or Torn3
- Device Contamination with Chemical or Other Material2
- Use of Device Problem2
- Contamination1
- Defective Component1
- Device Damaged by Another Device1
Patient problems
- No Code Available12
- No Clinical Signs, Symptoms or Conditions11
- Capsular Bag Tear4
- Vitrectomy3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 31, 2022Break; Break
IT WAS REPORTED THAT INTRAOCULAR LENS HAD BENT HAPTIC. IT IS UNKNOWN IF THE DEVICE HAD PATIENT CONTACT. NO OTHER INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15705248 ↗MalfunctionOct 15, 2022Break; Break
IT WAS REPORTED THAT AN INTRAOCULAR LENS (IOL) WAS FULLY INSERTED IN THE PATIENT'S RIGHT EYE AND WAS REMOVED AND REPLACED DURING THE SAME PROCEDURE DUE TO A BENT HAPTIC. THERE WAS NO VITRECTOMY, INCISION ENLARGEMENT, OR SUTURES REQUIRED. ANOTHER JOHNSON & JOHNSON LENS (SAME MODEL AND DIOPTER) WAS IMPLANTED AS A REPLACEMENT. THE PATIENT IS DOING FINE. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15612688 ↗MalfunctionJul 21, 2022Break; Break
IT WAS REPORTED THAT THE OPTIC OF THE INTRAOCULAR LENS (IOL) CRACKED. THE LENS WAS COMPLETELY INSERTED INTO PATIENT'S LEFT AND WAS REMOVED. ANOTHER LENS OF SAME MODEL AND DIOPTER WAS USED AS THE REPLACEMENT. IT WAS INDICATED THAT THE ISSUE WAS NOT DUE TO USE ERROR. THERE WAS NO PATIENT INJURY AND NO OTHER INTERVENTION WAS REQUIRED. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15080764 ↗MalfunctionMar 29, 2022Break; Break
IT WAS REPORTED THE INTRAOCULAR LENS (IOL) OPTIC RIPPED COMING OUT OF CARTRIDGE AND THERE WAS PATIENT CONTACT BUT NO INJURY. THE DOCTOR SAID THE PLUNGER WAS IN FRONT OF OPTIC AS HE WAS INSERTING. THE IOL IS NOT AVAILABLE FOR RETURN. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13935182 ↗MalfunctionJan 14, 2022Break; Break
IT WAS REPORTED THAT THE INTRAOCULAR LENS (IOL) HAD A BROKEN HAPTIC. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13261415 ↗MalfunctionOct 22, 2021Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT AN INTRAOCULAR LENS (IOL) WAS IMPLANTED AND ONCE IN THE LEFT EYE, IT WAS REALIZED TO BE TORN. IOL THEN WAS CUT OUT AND REPLACED WITH THE SAME MODEL AND DIOPTER IOL. NO INCISION ENLARGEMENT, VITRECTOMY OR SUTURES DONE, NO PATIENT POST-OP INJURY. NO OTHER INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12685835 ↗InjurySep 28, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT AN INTRAOCULAR LENS (IOL) WAS FULLY IMPLANTED AND REMOVED FROM THE PATIENT'S LEFT EYE DURING THE SAME SURGERY DUE TO CAPSULE TEAR. INCISION WAS ENLARGED AND SUTURES WERE DONE. IT WAS INDICATED THAT NO LENS WAS IMPLANTED. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Capsular Bag Tear
MAUDE report 12538834 ↗MalfunctionSep 20, 2021Break; Inaccurate Delivery; Break; Inaccurate Delivery
IT WAS REPORTED THAT A Z9002 MODEL INTRAOCULAR LENS(IOL) WAS PARTIALLY IMPLANTED INTO PATIENT'S RIGHT EYE WHEN PHYSICIAN NOTED THAT THE HAPTIC WAS BROKEN. THE INCISION WAS ENLARGED AND LENS WAS REMOVED AND REPLACED IN THE SAME PROCEDURE WITH ANOTHER LENS OF SAME MODEL AND DIOPTER. THERE WAS NO PATIENT INJURY REPORTED. ACCOUNT VERIFIED THAT THERE WERE NO MEDICAL OR SURGICAL INTERVENTION REQUIRED. PATIENT WAS REPORTEDLY DOING WELL. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12502918 ↗MalfunctionAug 24, 2021Use of Device Problem; Material Split, Cut or Torn; Use of Device Problem; Material Split, Cut or Torn
IT WAS REPORTED THAT A TECNIS MONOFOCAL INTRAOCULAR LENS (IOL) WAS REMOVED AND REPLACED WITH A BACK UP LENS (SAME MODEL/DIOPTER) DURING THE INITIAL PROCEDURE DUE TO THE LENS BEING TORN. THERE WAS NO HARM TO THE PATIENT AND SHE HAS RECOVERED FROM THE SURGERY. NO FURTHER INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12363142 ↗MalfunctionMay 21, 2021Break; Break
IT WAS REPORTED THAT A TECNIS MONOFOCAL INTRAOCULAR LENS (IOL) WAS FULLY INSERTED INTO THE PATIENT'S LEFT EYE, BUT HAD TO BE CUT OUT DURING THE SAME PROCEDURE BECAUSE THE LENS HAD A CRACK IN IT. THERE WAS NO ADDITIONAL SURGICAL INTERVENTION AND THE PATIENT WAS REPORTED TO BE DOING FINE AFTER SURGERY. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11861365 ↗MalfunctionMay 5, 2021Break; Device Damaged by Another Device; Break; Device Damaged by Another Device
IT WAS REPORTED INTRAOCULAR LENS (IOL) WAS OVERRIDDEN BY THE PLUNGER AND THE LENS MAY HAVE BEEN DAMAGED. CARTRIDGE TIP HAD PATIENT CONTACT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING ANOTHER LENS WITH THE SAME MODEL AND DIOPTER SIZE. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11775104 ↗MalfunctionMay 4, 2021Scratched Material; Material Split, Cut or Torn; Scratched Material; Material Split, Cut or Torn
IT WAS REPORTED THAT A Z9002 INTRAOCULAR LENS (IOL) WAS FULLY INSERTED IN THE RIGHT EYE (OD). THERE WAS A TORN LENS. THE LENS WAS ALSO SCRATCHED. IT WAS REMOVED AND REPLACED. THE PROCESS WAS COMPLETED SUCCESSFULLY WITH ANOTHER Z9002 LENS. THE LENS WAS DISCARDED. PATIENT HAS RECOVERED. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11773966 ↗MalfunctionApr 8, 2021Break; Break
IT WAS REPORTED THAT THE INTRAOCULAR LENS (IOL) HAD A BROKEN HAPTIC AND IT WAS NOTICED AFTER INSERTION INTO THE PATIENT'S LEFT EYE. THE LENS WAS REPLACED WITH BACKUP LENS OF SAME MODEL AND DIOPTER TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INJURY AND NO MEDICAL INTERVENTION WAS REQUIRED. THE PATIENT WAS DOING OK POST-OPERATION. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11632728 ↗InjuryJul 24, 2020Difficult to Insert; Defective Component; Difficult to Insert; Defective Component
IT WAS REPORTED THAT AN INTRAOCULAR LENS (IOL) HAD A HAPTIC SO BIG THE SURGEON COULD NOT GET IT POSITIONED. THE CAPSULE TORE AND A VITRECTOMY WAS REQUIRED, HOWEVER, IT WAS NOT PRODUCT RELATED. IT WAS ALSO REPORTED THE PATIENT WILL BE RETURNING APHAKIC. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Capsular Bag Tear; No Code Available; Vitrectomy
MAUDE report 10318331 ↗MalfunctionSep 13, 2019Contamination; Contamination
IT WAS REPORTED THAT DEBRIS WAS FOUND ON HAPTIC OF INTRAOCULAR LENS. LENS WAS IMPLANTED SUCCESSFULLY AFTER THE DEBRIS WAS REMOVED FROM HAPTIC. PATIENT WAS DOING WELL. THIS REPORT CAPTURES INFORMATION FOR LENS. ANOTHER REPORT IS BEING SUBMITTED FOR CARTRIDGE. NO FURTHER INFORMATION PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9041775 ↗InjuryJan 31, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE INTRAOCULAR LENS (IOL) WAS EXPLANTED FROM THE PATIENT'S LEFT EYE DUE TO A RESIDUAL REFRACTIVE ERROR. THERE WERE NO COMPLICATIONS REPORTED. THE LENS WAS REPLACED WITH THE SAME MODEL, Z9002, 20.5 DIOPTER. THE PATIENT WAS DOING GOOD POST OP. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Code Available
MAUDE report 8295462 ↗InjuryJan 31, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A Z9002 23.5 DIOPTER INTRAOCULAR LENS (IOL) WAS IMPLANTED IN THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2018. IT WAS LATER EXPLANTED ON (B)(6) 2018 BECAUSE OF A RESIDUAL REFRACTIVE ERROR. THERE WAS NO INCISION ENLARGEMENT, NO VITRECTOMY, AND NO SUTURES USED. THE REPLACEMENT LENS WAS THE SAME MODEL, BUT DIFFERENT DIOPTER OF 26.5. REPORTEDLY, THE PATIENT IS DOING GOOD. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Code Available
MAUDE report 8295149 ↗InjuryJan 22, 2019Failure to Unfold or Unwrap; Failure to Unfold or Unwrap
IT WAS REPORTED THAT TECNIS MONOFOCAL INTRAOCULAR LENS (MODEL Z9002 23.5 DIOPTER) WAS INSERTED, REMOVED AND REPLACED IN SAME SURGICAL PROCEDURE BECAUSE HAPTIC DID NOT UNFOLD. REPORTEDLY INCISION ENLARGEMENT WAS PERFORMED DURING THE PROCEDURE. ANOTHER LENS WITH SAME MODEL AND SAME DIOPTER WAS USED AS REPLACEMENT FOR THE PATIENT. NO FURTHER INFORMATION PROVIDED.
Patient problems: No Code Available
MAUDE report 8265920 ↗MalfunctionJan 10, 2019Break; Break
IT WAS REPORTED Z9002 23.5 DIOPTER INTRAOCULAR LENS (IOL) THE HAPTIC WAS BENT AND THERE WAS PATIENT CONTACT. NO ADDITIONAL INFORMATION WAS PROVIDED TO JOHNSON & JOHNSON SURGICAL VISION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8233493 ↗InjurySep 7, 2018Positioning Failure; Failure to Align; Positioning Failure; Failure to Align
IT WAS REPORTED THAT SURGEON WAS CONCERNED OF Z9002 MONOFOCAL INTRAOCULAR LENS (IOL) WOULD NOT CENTER IN THE RIGHT EYE (OD) DUE TO PATIENT'S ANATOMY. THE DOCTOR OPTED TO REMOVE AND REPLACE WITH NON-JNJSV (JOHNSON & JOHNSON SURGICAL VISION) PRODUCT. IT WAS NOTED THAT INCISION ENLARGEMENT WAS PERFORMED TO REMOVE THE LENS. HOWEVER, THERE WAS NO PATIENT INJURY AS THE PATIENT IS DOING FINE POST-SURGERY. PRODUCT IS EXPECTED TO BE RETURNED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Code Available
MAUDE report 7856188 ↗InjuryOct 9, 2017Break; Break
IT WAS REPORTED THAT THE HAPTIC OF A Z9002 23.5 DIOPTER INTRAOCULAR LENS (IOL) BROKE OFF WHEN INSERTING INTO THE PATIENT'S OPERATIVE EYE. REPORTEDLY, THE LENS WAS REMOVED AND AN INCISION ENLARGEMENT AND VITRECTOMY WERE PERFORMED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Code Available; Vitrectomy
MAUDE report 6933322 ↗InjuryOct 3, 2017Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
IT WAS REPORTED THAT AN INTRAOCULAR LENS (MODEL Z9002, +23.5 DIOPTER) WAS EXPLANTED FROM THE RIGHT EYE OF A (B)(6) FEMALE PATIENT BECAUSE OF RESIDUAL REFRACTIVE ERRORS. NO INCISION ENLARGEMENT, NO COMPLICATIONS, NO INJURY AND NO ADDITIONAL PROCEDURES WERE REQUIRED. ANOTHER LENS, SAME MODEL HIGHER DIOPTER (+26.0) WAS IMPLANTED AS A REPLACEMENT. THE PATIENT WAS REPORTED AS RECOVERED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Code Available
MAUDE report 6912513 ↗InjurySep 12, 2017Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT THERE IS ONE PATIENT WITH TWO PSEUDOPHAKIC EYES. THE INTRAOCULAR LENSES (IOL'S) WERE IMPLANTED BACK IN 2008. THE PATIENT WAS DIAGNOSED WITH ASTEROID HYALOSIS AND UNDERWENT A YAG PROCEDURE. NOW AN OBSERVATION HAS BEEN MADE THAT THERE ARE DEPOSITS ON BOTH IOL'S (RIGHT AND LEFT EYES). BOTH IOL¿S ARE STILL IMPLANTED. NO ADDITIONAL INFORMATION WAS PROVIDED TO ABBOTT MEDICAL OPTICS. NOTE: 2 MDR'S WILL BE FILED, ONE FOR EACH EYE. THIS REPORT CAPTURES THE LEFT EYE.
Patient problems: No Code Available
MAUDE report 6859898 ↗InjurySep 12, 2017Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT THERE IS ONE PATIENT WITH TWO PSEUDOPHAKIC EYES. THE INTRAOCULAR LENSES (IOL'S) WERE IMPLANTED BACK IN 2008. THE PATIENT WAS DIAGNOSED WITH ASTEROID HYALOSIS AND UNDERWENT A YAG PROCEDURE. NOW AN OBSERVATION HAS BEEN MADE THAT THERE ARE DEPOSITS ON BOTH IOL'S (RIGHT AND LEFT EYES). BOTH IOL¿S ARE STILL IMPLANTED. NO ADDITIONAL INFORMATION WAS PROVIDED TO ABBOTT MEDICAL OPTICS. NOTE: 2 MDR'S WILL BE FILED, ONE FOR EACH EYE. THIS REPORT CAPTURES THE RIGHT EYE.
Patient problems: No Code Available
MAUDE report 6859875 ↗InjuryDec 16, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING IMPLANTATION OF A Z9002 23.5 DIOPTER INTRAOCULAR LENS (IOL) THE PATIENT'S CAPSULE BROKE. THE LENS WAS INSERTED AND THE LENS HAD TO BE CUT IN HALF TO BE REMOVED AS THERE WAS NOT ENOUGH ZONULES TO SUPPORT THE LENS. INCISION ENLARGEMENT AND ONE SUTURE WAS REQUIRED. THROUGH FOLLOW-UP IT WAS CLARIFIED THAT THE PATIENT WENT HOME APHAKIC, WITHOUT A LENS BEING IMPLANTED AND WAS REFERRED TO A RETINAL SPECIALIST. THERE WAS NO ISSUE THE AMO CARTRIDGE OR LENS. THE ISSUE WAS WITH THE PATIENT'S ANATOMY. PATIENTS HAD WEAK ZONULES AND AS A RESULT THE CAPSULE BROKE.
Patient problems: Capsular Bag Tear; No Code Available
MAUDE report 6183465 ↗
Other intraocular lens devices
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens146 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens145 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens144 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens139 reports
- TECNIS SIMPLICITY DIB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens1 recall81 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens129 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens127 reports
- TECNIS IOL PCB00Johnson & Johnson Surgical Vision, Inc. · intraocular lens126 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page