Skip to content

TECNIS IOL ZCB00 (REF ZCB0000300)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · TECNIS 1-PC CLEAR MONO 6.0MM 30.0D · Class 3 · Implant

All models of TECNIS IOL (1344)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 05050474531918), received between Feb 21, 2017 and Oct 27, 2023 — classified as 5 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Break1
  • Difficult To Position1
  • Difficult to Fold, Unfold or Collapse1
  • Inadequacy of Device Shape and/or Size1
  • Scratched Material1

Patient problems

  • No Code Available4
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Capsular Bag Tear1
  • Injury1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 27, 2023Scratched Material; Scratched Material

    IT WAS REPORTED THAT THE NON-PRELOADED MONOFOCAL INTRAOCULAR LENS (IOL) WAS SCRATCHED AND COULDN'T BE USED. ADDITIONAL INFORMATION RECEIVED STATED THAT THE SCRATCH WAS NOTICED DURING IMPLANTATION OF IOL IN THE PATIENTS RIGHT EYE. THE LENS WAS PARTIALLY INSERTED AND REMOVED AS DOCTOR SAW SCRATCH AND DID NOT WANT TO TAKE CHANCE OF IT BEING SCRATCHED AND NOT WORK IN PATIENT EYE. THERE WAS NO PATIENT INJURY AND NO OTHER MEDICAL OR SURGICAL INTERVENTION REQUIRED. THERE WAS NO USE ERROR. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING ANOTHER BACK UP IOL OF SAME MODEL AND DIOPTER. THE LENS WAS DESTROYED AND NOT AVAILABLE FOR RETURN. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18022618 ↗
  • InjuryOct 3, 2023Difficult to Fold, Unfold or Collapse; Adverse Event Without Identified Device or Use Problem; Difficult to Fold, Unfold or Collapse; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT NON-PRELOADED MONOCULAR INTRAOCULAR LENS (IOL) WITH MODEL ZCB00 DISPLACED INFERIONASALLY. SURGEON HAD TO CUT OUT AND USE BACKUP LENS WITH MODEL ZA9003 IN SAME SURGICAL PROCEDURE. SUTURES WERE REQUIRED, AFTER A NEW LENS, THE PATIENT RECOVERED WITHOUT ISSUE. ADDITIONAL INFORMATION REVEALED THAT THE SURGERY PROCEEDED NORMALLY WITH LENS REMOVAL. THE CAPSULAR BAG WAS RE-INFLATED WITH VISCOELASTIC AND THE CAPSULAR BAG WAS INTACT. THE ZCB00 LENS WAS PLACED IN THE EYE USING A SHOOTER AND ALLOWED TO UNFOLD IN THE CAPSULAR BAG. THE INSERTER WAS NOT ADVANCED INTO THE ANTERIOR CHAMBER. A KUGLEN HOOK WAS USED TO PUSH THE TRAILING HAPTIC FROM THE ANTERIOR CHAMBER INTO THE CAPSULAR BAG, AND A TEAR IN THE POSTERIOR CAPSULE WAS SEEN WITH THE LEADING HAPTIC GOING THROUGH THE TEAR. IOL WAS BROUGHT BACK INTO THE ANTERIOR CHAMBER AND REMOVED FROM THE EYE BY CUTTING EACH LENS INTO THREE PIECES AND REPLACING THEM WITH A THREE-PIECE SULCUS LENS. PATIENT HAS BEEN REFERRED TO A RETINA SPECIALIST FOR EVALUATION BECAUSE THE SULCUS IOL DISLOCATED. APPOINTMENT IS SCHEDULED FOR 9/14/2023.NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Capsular Bag Tear

    MAUDE report 17864964 ↗
  • InjuryAug 28, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A ZCB00 30.0 DIOPTER LENS WAS IMPLANTED IN THE PATIENT'S OPERATIVE EYE ON (B)(6) 2018. IT WAS LATER EXPLANTED ON (B)(6) 2018 BECAUSE THE PATIENT EXPERIENCED ANISOMETROPIA. THERE WAS NO INCISION ENLARGEMENT AND THERE WAS NO PATIENT INJURY. THE REPLACEMENT LENS WAS ANOTHER ZCB00 MODEL. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Code Available

    MAUDE report 7823042 ↗
  • InjuryAug 17, 2018Break; Break

    IT WAS REPORTED THAT THE INTRAOCULAR LENS (IOL) HAPTIC WAS NEARLY BROKEN FROM THE OPTIC. AS A RESULT, LENS WAS REMOVED AND REPLACED WITH AN INCISION ENLARGEMENT. SURGERY WAS SUCCESSFULLY COMPLETED USING THE BACK-UP LENS, SAME MODEL AND DIOPTER. PATIENT WAS DISCHARGED WITH NO COMPLICATIONS. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Code Available

    MAUDE report 7793659 ↗
  • InjurySep 5, 2017Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THAT THE INTRAOCULAR LENS (IOL) WAS EXPLANTED AS THE PATIENT WAS NOT SATISFIED WITH HIS/HER FAR VISION. THE IOL WAS REPLACED WITH AN IOL OF THE SAME MODEL BUT DIFFERENT DIOPTER. NO ADDITIONAL INFORMATION WAS PROVIDED TO ABBOTT MEDICAL OPTICS.

    Patient problems: No Code Available

    MAUDE report 6843612 ↗
  • InjuryFeb 21, 2017Difficult To Position; Difficult To Position

    IT WAS REPORTED THAT AN INTRAOCULAR LENS (IOL) WOULD NOT POSITION CORRECTLY IN THE EYE. REPORTEDLY THE LENS WAS CUT OUT OF THE EYE. AN UNDEFINED PATIENT INTERVENTION/INJURY WAS REPORTED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Injury; No Code Available; No Consequences Or Impact To Patient

    MAUDE report 6347814 ↗

Other intraocular lens devices

All intraocular lens devices →

Sources and updates