SENSAR IOL AAB00 (REF AAB0000260)Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · SENSAR 1-PC CLEAR MONO 6.0MM 26.0D · Class 3 · Implant

All models of SENSAR IOL (112)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05050474548787), received between Sep 2, 2025 and May 12, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Mechanical Jam1
  • Scratched Material1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Capsular Bag Tear1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 12, 2026Break; Mechanical Jam

    IT WAS REPORTED THAT DURING THE NON-PRELOADED INTRAOCULAR LENS (IOL) IMPLANTATION, WHEN THE IOL WAS ABOUT TO EXIT THE TIP OF THE CARTRIDGE, THE IOL BURST THROUGH THE TIP OF THE CARTRIDGE AND CAME OUT WITH ONE OF THE HAPTICS BROKEN. THE PHYSICIAN¿S IMPRESSION WAS THAT THERE WAS A PROBLEM WITH THE CARTRIDGE. ACCOUNT INDICATED THAT THE IOL WAS IN CONTACT WITH THE PATIENT'S EYE; HOWEVER, IT BECAME LODGED INSIDE THE CARTRIDGE. THE REPLACEMENT IOL WAS FROM A DIFFERENT MANUFACTURER WITH THE SAME DIOPTER. THERE WAS A 15 MINUTE DELAY IN PROCEDURE. THERE WAS NO PATIENT INJURY REPORTED AND NO MEDICAL OR SURGICAL INTERVENTIONS WERE PERFORMED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25156404 ↗
  • InjurySep 2, 2025Scratched Material; Scratched Material

    IT WAS REPORTED THAT AFTER THE IMPLANTATION OF A NON-PRELOADED MONOFOCAL INTRAOCULAR LENS (IOL), A PERMANENT CREASE LINE WAS OBSERVED AT THE OPTIC CENTER. THE LENS WAS REMOVED AND REPLACED DURING THE SAME PROCEDURE, WHICH REQUIRED AN ENLARGED INCISION, SUTURES, AND A VITRECTOMY. THE PROCEDURE WAS DELAYED, AND THE PATIENT REQUIRED MEDICAL ATTENTION AND MEDICATION. THE PATIENT EXPERIENCED TEMPORARY IMPAIRMENT. INSTRUCTIONS FOR USE WERE FOLLOWED. ADDITIONAL INFORMATION RECEIVED ON AUGUST 18 AND 19, 2025, CLARIFIED DETAILS REGARDING THE INTERVENTION MENTIONED IN THE COMPLAINT FORM SUBMITTED BY THE HOSPITAL. THE UNPLANNED VITRECTOMY WAS DUE TO A CAPSULE TEAR DURING THE REMOVAL OF THE DAMAGED LENS, AND THE MEDICATION PRESCRIBED WAS OUTSIDE OF THE STANDARD POST-OPERATIVE CARE. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Capsular Bag Tear