TECNIS IOL ZCT150 (REF ZCT150U165)Recalls & reported problems
Johnson & Johnson Surgical Vision, Inc. · TECNIS TORIC 1-PIECE 6.0MM 16.5D CYL1.50 · Class 3 · Implant
All models of TECNIS IOL (1344)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05050474552074), received between Oct 28, 2016 and Sep 23, 2019 — classified as 4 injuries.
The most frequently reported specific device problem is “Unintended Movement” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Unintended Movement1
Patient problems
- No Code Available4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjurySep 23, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT INTRAOCULAR LENS (IOL) CREATED AN ADDITIONAL CYLINDER, THEREFORE, THE LENS WAS EXPLANTED. THE INCISION WAS ENLARGED AND A 10-0 SUTURE WAS USED TO CLOSE THE WOUND. THERE WERE NO OTHER PATIENT CONSEQUENCES. THE REPLACEMENT LENS WAS A DIFFERENT MODEL AND HIGHER DIOPTER. NO ADDITIONAL INFORMATION WAS PROVIDED TO JOHNSON & JOHNSON SURGICAL VISION.
Patient problems: No Code Available
MAUDE report 9107578 ↗InjuryJul 26, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A MODEL ZCT150 16.5 DIOPTER LENS WAS EXPLANTED DUE TO THE PATIENT EXPERIENCING DOUBLE VISION AND HYPEROPTIC SURPRISE. THE IOL WAS REPLACED WITH A SAME MODEL LENS BUT HIGHER DIOPTER (19.0). REPORTEDLY THE PATIENT IS DOING WELL. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Code Available
MAUDE report 8829594 ↗InjuryJun 7, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE INTRAOCULAR LENS (IOL) WAS EXPLANTED FROM THE LEFT EYE DUE TO UNINTENDED HYPEROPIA AND RESIDUAL ASTIGMATISM. THE EXPLANTED IOL WAS REPLACED WITH A ZCT 225 17.0 IOL. THERE WAS NO ADDITIONAL INTERVENTION RECEIVED. POST OP VISUAL ACUITY: 20/40. POST OP VISUAL ACUITY AFTER LENS REPLACEMENT: 20/20.
Patient problems: No Code Available
MAUDE report 8680612 ↗InjuryOct 28, 2016Unintended Movement; Unintended Movement
IT WAS REPORTED THAT A ZCT150 16.5 DIOPTER INTRAOCULAR LENS WAS EXPLANTED FROM THE PATIENT'S RIGHT EYE DUE TO THE LENS ROTATING 35 DEGREES, EVEN WITH A CTR (CAPSULAR TENSION RING). NO PATIENT POST-OP INJURY WAS REPORTED. THE LENS AND THE CTR WILL NOT BE RETURNED. THE LENS WAS EXPLANTED AND REPLACED WITH A ZCT150 14.0 (SMALLER) DIOPTER INTRAOCULAR LENS IN THE RIGHT EYE. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Code Available
MAUDE report 6064328 ↗
Other LENS, INTRAOCULAR, TORIC OPTICS devices
- TECNIS IOL ZCT150Johnson & Johnson Surgical Vision, Inc. · LENS, INTRAOCULAR, TORIC OPTICS1 recall7 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS43 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS36 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS32 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS30 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS29 reports
- Clareon CNWTT0Alcon Laboratories, Inc. · LENS, INTRAOCULAR, TORIC OPTICS21 reports
- TECNIS IOL DIU150Johnson & Johnson Surgical Vision, Inc. · LENS, INTRAOCULAR, TORIC OPTICS19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page